Evaluation of diaphragm thickening by diaphragm ultrasonography: a reproducibility and a repeatability study.

Iacopo Cappellini1, Fabio Picciafuochi2, Maurizio Bartolucci2, Simona Matteini3, Gianni Virgili3,4, Chiara Adembri3,2.   

Abstract

PURPOSE: We have focused on the two-dimensional (B-mode) and the time-motion (M-mode) analysis of the zone of apposition to determine the reliability of diaphragm ultrasonography in the clinical environment.
METHODS: Ten healthy volunteers were enrolled and studied by three operators with different skills in ultrasonography. For every volunteer, each operator acquired three images of the diaphragm for each side, both in B-mode and in M-mode. Then a fourth operator calculated the thickening fraction (TF), by means of the formula TF = (TEI - TEE)/TEE (TEI is the thickness at end inspiration and TEE the thickness at end expiration). Afterwards, intraclass correlation coefficients (ICCs) were computed on TF to establish reproducibility and repeatability both in the B- and M-modes. A Coefficient of Repeatability or repeatability (CR) ≤ 0.3 was considered acceptable.
RESULTS: Both B-mode (CRs 0.16-0.26) and M-mode (CRs 0.10-0.15) were sufficiently repeatable to assess TF, except for the less experienced operator (CRs B-Mode 0.20-0.32). Reproducibility was moderate to good between operators with CRs much narrower for the M-Mode (0.13-0.14).
CONCLUSIONS: The results of our study have shown that diaphragm ultrasound is repeatable and reproducible when carried out by a radiologist or an intensivist with a basic curriculum in ultrasonography. The method is more accurate when using the M-mode for less experienced operators, and in this case, repeatability and reproducibility are not sufficient to make clinical decisions. No TF value lower than 36% was obtained using both techniques. This suggests the existence of a cut-off value that could be used as an initial tool to discriminate healthy subjects from those affected by diaphragmatic dysfunction. CLINICAL TRIAL REGISTRATION: EUDRACT 2015-004635-12.

Entities:  

Keywords:  Diaphragm ultrasonography; Repeatability

Year:  2020        PMID: 32358646     DOI: 10.1007/s40477-020-00462-x

Source DB:  PubMed          Journal:  J Ultrasound        ISSN: 1876-7931


  4 in total

1.  Ultrasound variations of diaphragm activity between prone position versus supine position in ventilated patients: a cross-sectional comparative study.

Authors:  Ahlem Trifi; Sami Abdellatif; Cyrine Abdennebi; Foued Daly; Yosr Touil; Salah Ben Lakhal
Journal:  J Ultrasound       Date:  2020-09-01

2.  Diaphragm ultrasound to stratify COVID-19 patients in the emergency department?

Authors:  Abdallah Fayssoil; Sebastien Beaune; Benjamin Davido; Nicolas Mansencal
Journal:  J Clin Ultrasound       Date:  2021-10-11       Impact factor: 0.869

3.  EXpert consensus On Diaphragm UltraSonography in the critically ill (EXODUS): a Delphi consensus statement on the measurement of diaphragm ultrasound-derived parameters in a critical care setting.

Authors:  Mark E Haaksma; Jasper M Smit; Alain Boussuges; Alexandre Demoule; Martin Dres; Giovanni Ferrari; Paolo Formenti; Ewan C Goligher; Leo Heunks; Endry H T Lim; Lidwine B Mokkink; Eleni Soilemezi; Zhonghua Shi; Michele Umbrello; Luigi Vetrugno; Emmanuel Vivier; Lei Xu; Massimo Zambon; Pieter R Tuinman
Journal:  Crit Care       Date:  2022-04-08       Impact factor: 9.097

4.  Intra-Rater Reliability of Shear Wave Elastography for the Quantification of Respiratory Muscles in Adolescent Athletes.

Authors:  Małgorzata Pałac; Paweł Linek
Journal:  Sensors (Basel)       Date:  2022-09-01       Impact factor: 3.847

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.