Else A Aalbersberg1, Lotte van Andel2,3, Martine M Geluk-Jonker2, Jos H Beijnen3, Marcel P M Stokkel2, Jeroen J M A Hendrikx2,3.
Abstract
BACKGROUND: Lymph node dissection is a therapeutic option for prostate cancer patients with a high risk of- or proven lymph node metastases. Radioguided surgery after intravenous injection of [99mTc]Tc-PSMA could improve the selectivity of lymph node dissection. The aim of this project was to develop an automated synthesis method for [99mTc]Tc-PSMA, using the disposables and chemicals used at our institute for [68Ga]Ga-PSMA labeling. Furthermore, quality control procedures and validation results of the automated production of [99mTc]Tc-PSMA conform cGMP and cGRPP are presented.Entities:
Keywords: 99mTc; Automated synthesis; PSMA; PSMA-I&S; Prostate cancer; Quality control; Radioguided surgery; [99mTc]Tc-PSMA
Year: 2020 PMID: 32358637 PMCID: PMC7195502 DOI: 10.1186/s41181-020-00095-9
Source DB: PubMed Journal: EJNMMI Radiopharm Chem ISSN: 2365-421X
Fig. 1Set-up of the automated Scintomics system for radiolabeling of 99mTc-PSMA
Fig. 2Structure of [99mTc]Tc-PSMA-I&S
Fig. 3Instant thin layer chromatography of both [99mTc]Tc-nanocoll and [99mTc]Tc-PSMA. The retention factor of 99mTc-colloids is less than 0.2 and that of [99mTc]Tc-PSMA greater than 0.7
Results of three validation batches of [99mTc]Tc-PSMA
| Item | Time (hours) | Batch 1 | Batch 2 | Batch 3 | |||
|---|---|---|---|---|---|---|---|
| 75.05% | 77.94% | 74.82% | |||||
Colorless No particles | Colorless No particles | Colorless No particles | |||||
Colorless No particles | Colorless No particles | Colorless No particles | Colorless No particles | Colorless No particles | Colorless No particles | ||
Colorless No particles | Colorless No particles | Colorless No particles | Colorless No particles | Colorless No particles | Colorless No particles | ||
Colorless No particles | Colorless No particles | ||||||
| 7.5 | 7.5 | 7.5 | |||||
| 7.5 | 7.5 | 7.5 | 7.5 | 7.5 | 7.5 | ||
| 7.5 | 7.5 | 7.5 | 7.5 | 7.5 | 7.5 | ||
| 7.5 | 7.5 | ||||||
| TSB clear | TSB clear | TSB clear | |||||
| < 50 IU/2 ml | < 50 IU/2 ml | < 50 IU/2 ml | |||||
| Main peak at 141 keV | Main peak at 141 keV | Main peak at 141 keV | |||||
| 1,46% | 1,92% | 2,82% | |||||
| 6,56% | 7,02% | 6,83% | 7,39% | 7,81% | 6,64% | ||
| 6,65% | 7,18% | 7,60% | 6,47% | 8,51% | 8,40% | ||
| 10,96% * | 9,44% | ||||||
( | 0,34% | 0,48% | 0,84% | ||||
| 0,68% | 1,06% | 0,28% | 0,78% | 1,22% | 0,90% | ||
| 0,58% | 1,02% | 0,82% | 1,10% | 1,10% | 1,56% | ||
| 1,46% | 1,34% | ||||||
| 98,20% | 97,60% | 96,34% | |||||
| 92,76% | 91,92% | 92,89% | 91,83% | 90,97% | 92,46% | ||
| 92,77% | 91,80% | 91,58% | 92,43% | 90,39% | 90,04% | ||
| 87.58% * | 89,22% * | ||||||
(Criteria: < 200 μg/patient dosage) | < 200 μg/patient dose | < 200 μg/patient dose | < 200 μg/patient dose | ||||
The results of the three validation batches (batch 1, batch 2, batch 3) are shown at different time points (t = 0, t = 2, t = 6, and t = 24 h). At t = 0 one sample was taken, whilst at the other time points a samples was measured from either the final vial or a prepared patient syringe kept at room temperature. The accepted limit/criteria is noted in the first column. * indicates that the value does not meet the criteria
Fig. 4High performance liquid chromatography (HPLC) chromatogram of the radiodetector. The retention time of unbound [99mTc]pertechnetate is 2.5 min. The retention time of [99mTc]Tc-PSMA is between 10 and 15 min, with different peaks accounting for different protonation states
Comparison between the automated method described in this article and the methods described in literature
| Aalbersberg et al. (current method) | Robu et al. | Robu et al. | Kratzik et al. | |
|---|---|---|---|---|
| Automated | Manual | Manual | Manual | |
| HEPES | Ammonium acetate & disodium tartrate | Sodium hydrogen phosphate & Sodium tartrate | Unknown | |
| 40 μg PSMA-I&S (=28.7 nmol) | 20–30 nmol PSMA-I&S | 25 nmol PSMA-I&S | PSMA-I&S (unknown quantity) | |
| 100 °C 20 min | 90 °C 20 min | 90 °C 20 min | 100 °C 10 min | |
| Yes | Yes | No | No |