| Literature DB >> 32341418 |
Patryck Lloyd-Donald1, Leonid Churilov2, Brandon Cheong1, Rinaldo Bellomo3, Peter R McCall1, Johan Mårtensson4, Neil Glassford3, Laurence Weinberg5,6.
Abstract
The TEG6S is a novel haemostasis analyser utilising resonance technology. It offers potentially greater coagulation information and ease of use, however has not been independently validated in a clinical setting. We aimed to determine if the TEG6S is reliable between devices and across time points. We performed a prospective observational study with ethical approval. For interdevice reliability, we performed simultaneous analysis on two TEG6S devices on 25 adult ICU patients. For time point reliability, we performed repeated sampling across five different time points on 15 adult participants. Blood was collected with informed consent, or as standard care, before four-channel citrated kaolin analysis. We observed almost perfect interdevice reliability across all TEG parameters. The Lin's concordance correlation coefficients (95% CI, major axis regression slope, intercept) were R-time: 0.96 (0.92-0.99, 0.88, 0.57); K-time: 0.93 (0.87-0.98, 1.07, 0.00); Alpha Angle: 0.87 (0.78-0.96, 1.20, -14.10); Maximum Amplitude: 0.99 (0.98-0.99, 1.02, -1.38); Clot Lysis: 0.89 (0.82-0.97, 1.20, 0.07). Additionally, we observed moderate-to-high reliability across time points. Demonstrating almost perfect agreement across different devices and moderate-to-high reliability across multiple time points, suggests the TEG6S platform can be used with haemostatic accuracy and generalisability. This has potentially significant implications for clinical practice and multi-site research programs.Entities:
Mesh:
Substances:
Year: 2020 PMID: 32341418 PMCID: PMC7184600 DOI: 10.1038/s41598-020-63964-y
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Conventional coagulation test results for patients used for interdevice testing (n = 25).
| Coagulation Test | Median (IQR) Normal reference range |
|---|---|
| Prothrombin time (sec) | 13 (11–15) 11–15 |
| International normalised ratio | 1.2 (1.1–1.4) |
| Activated partial thromboplastin time (sec) | 30 (25–38) 22–41 |
| Fibrinogen (g/L) | 2.7 (2.1–4.4) 2.0–4.0 |
| Platelets (x109 cell/L) | 129 (62–186) 150–400 |
Interdevice reliability of thromboelastograph parameters between two TEG6S devices.
| Parameter | Lin’s Concordance Coefficient (95% CI) | Reduced Major Axis Intercept | Slope |
|---|---|---|---|
| CK R time | 0.96 (0.93–0.99) | 0.57 | 0.88 |
| CK K time | 0.93 (0.87–0.98) | 0.01 | 1.07 |
| CK alpha | 0.87 (0.78–0.96) | −14.10 | 1.20 |
| CK MA | 0.99 (0.98–1.00) | −1.38 | 1.03 |
| CK LY30 | 0.89 (0.82–0.97) | 0.07 | 1.20 |
| HK R time | 0.95 (0.92–0.99) | 0.33 | 0.94 |
| HK K time | 0.99 (0.98–1.00) | 0.04 | 0.99 |
| HK alpha | 0.95 (0.91–0.99) | −9.90 | 1.13 |
| HK MA | 1.00 (0.99–1.00) | 0.19 | 1.00 |
| RT R time | 0.94 (0.89–0.99) | −0.03 | 0.95 |
| RT K time | 1.00 (1.00–1.00) | 0.03 | 0.95 |
| RT alpha angle | 0.99 (0.99–1.00) | 2.00 | 0.98 |
| RT MA | 1.00 (1.00–1.00) | 0.32 | 1.00 |
| RT LY30 | 0.95 (0.92–0.98) | 0.02 | 0.79 |
| RT ACT | 0.96 (0.93–0.99) | −2.96 | 0.94 |
| FF MA | 0.99 (0.98–1.00) | −0.55 | 1.06 |
| FF FLEV | 0.99 (0.98–1.00) | −13.60 | 1.07 |
Abbreviations: CK: standard kaolin; HK: standard kaolin with heparinase; RT: rapid thromboelastograph; R: reaction; K: kinetic; MA: maximum amplitude, LY: amplitude at 30 minutes; ACT: activated clotting time; FF: functional fibrinogen; FLEV: functional fibrinogen level. Reduced major axis slope different to 1 is indicative of the presence of proportional bias; the combination of the slope close to 1 and intercept different from 0 is indicative of the presence of fixed bias.
Time point (test re-test) reliability across thromboelastograph parameters.
| Parameter | Intra-class correlation coefficient (95% CI) |
|---|---|
| CK R time | 0.81 (0.66–0.91) |
| CK K time | 0.69 (0.48–0.85) |
| CK alpha | 0.61 (0.39–0.80) |
| CK MA | 0.96 (0.91–0.98) |
| CK LY30 | 0.80 (0.61–0.91) |
| HK R time | 0.70 (0.50–0.86) |
| HK K time | 0.83 (0.68–0.92) |
| HK alpha | 0.56 (0.34–0.77) |
| HK MA | 0.96 (0.93–0.98) |
| RT R time | 0.73 (0.54–0.87) |
| RT K time | 0.96 (0.91–0.98) |
| RT alpha | 0.96 (0.92–0.98) |
| RT MA | 0.97 (0.94–0.99) |
| RT LY30 | 0.96 (0.91–0.98) |
| RT ACT | 0.73 (0.54–0.87) |
| FF MA | 0.97 (0.94–0.99) |
| FF FLEV | 0.97 (0.94–0.99) |
Abbreviations: CK: standard kaolin; HK: standard kaolin with heparinase; RT: rapid thromboelastograph; R: reaction; K: kinetic; MA: maximum amplitude, LY: amplitude at 30 minutes; ACT: activated clotting time; FF: functional fibrinogen; FLEV: functional fibrinogen level.
Figure 1Citrated kaolin (CK): Interdevice agreement of reaction time, kinetics time, alpha angle, maximal amplitude and lysis.