V Ram Peddi1, Parul S Patel1, Courtney Schieve2, Stan Rose2, M Roy First2.
Abstract
BACKGROUND TruGraf is a blood-based biomarker test that measures differential expression of a collection of genes that have been shown to correlate with surveillance biopsy results. However, in the majority of U.S. transplant centers, surveillance biopsies are not performed. The objectives of this study were to evaluate the clinical validity of TruGraf in stable kidney transplant recipients and to demonstrate the potential clinical utility of serial TruGraf testing in a center not utilizing surveillance biopsies. MATERIAL AND METHODS Serum creatinine levels, TruGraf testing at multiple time points, and subsequent clinical follow-up were obtained for 28 patients. RESULTS Overall concordance of TruGraf results, when compared with independent clinical assessment of testing, was 77% (54/70) for all tests; 79% (22/28) for test 1, 75% (21/28) for test 2, and 79% (11/14) for test 3. The negative predictive value (NPV) was 98.0%. Analysis of clinical utility indicated that 77% of TruGraf results would have been useful in patient management. CONCLUSIONS Our results indicate the value of serial TruGraf testing in those transplant centers that do not perform surveillance biopsies as part of their standard of care. The high negative predictive value indicates the ability of TruGraf to confirm immune quiescence with a high degree of probability in patients with a Transplant eXcellence (TX) result, without the need to perform a surveillance biopsy.Entities:
Mesh:
Substances:
Year: 2020 PMID: 32341330 PMCID: PMC7204430 DOI: 10.12659/AOT.920839
Source DB: PubMed Journal: Ann Transplant ISSN: 1425-9524 Impact factor: 1.530
Clinical characteristics of testing cohorts.
| Characteristic | Cohort A | Cohort B |
|---|---|---|
| Total Subjects | 14 | 14 |
| Male | 6 | 5 |
| Female | 8 | 9 |
| Average age at time of enrollment, (median) | 56 (63) | 63 (65) |
| Serum creatinine level (mg/dL) at time of TruGraf testing (mean±SD) | ||
| 1st test | 1.09±0.208 | 1.01±0.367 |
| 2nd test | 1.03±0.233 | 0.95±0.338 |
| 3rd test | NA | 0.89±0.309 |
| Testing Interval | ||
| Time between Test 1 and Test 2 (days) | 156 | 93 |
| Time between Test 2 and Test 3 (days) | NA | 200 |
NA – not applicable
Figure 1Serum Creatinine levels (mg/dL) and TruGraf results in patients with 2 sequential tests Cohort A (A) and 3 sequential tests Cohort B (B). Green – True Negative; Blue – True Positive; Red – False Negative; Grey – False Positive.
Sequential results of the TruGraf blood test and comparison with clinical phenotype in 28 kidney transplant recipients with stable renal function.
| Clinical phenotype TX | Clinical phenotype not-TX | |
|---|---|---|
| TruGraf Blood Test TX | 50 | 1 |
| TruGraf Blood Test not-TX | 15 | 4 |
Accuracy=54/70 (77%); Accuracy of TruGraf TX result=50/51 (98%); NPV=98.0; PPV=21.1; Sensitivity=80.0; Specificity=76.9.