Literature DB >> 32340804

Intermediate-term outcomes of aortic valve replacement using a bioprosthesis with a novel tissue.

Douglas R Johnston1, Bartley P Griffith2, John D Puskas3, Joseph E Bavaria4, Lars G Svensson1.   

Abstract

OBJECTIVES: The COMMENCE trial was conducted to evaluate the safety and effectiveness of an aortic bioprosthesis with novel RESILIA tissue (Edwards Lifesciences, Irvine, Calif). Reports of early noncalcific valve failure resulting from thrombosis or leaflet tears in other valves warrant careful evaluation of early valve performance.
METHODS: Patients underwent clinically indicated surgical aortic valve replacement with the Edwards Pericardial Aortic Bioprosthesis, Model 11000A (Edwards Lifesciences) in a prospective, multinational, multicenter (n = 27), single-arm, Food and Drug Administration Investigational Device Exemption trial. Events were adjudicated by an independent clinical events committee; echocardiograms were analyzed by an independent core laboratory.
RESULTS: Between January 2013 and March 2016, 689 patients received the study valve. Mean age was 67.0 ± 11.6 years. Mean Society of Thoracic Surgeons predicted risk of mortality was 2.0% ± 1.8%. Follow-up duration was 3.7 ± 1.2 years, with a total of 2533 patient years of follow-up and a median follow-up of 4 years. Early all-cause mortality was 1.2%, thromboembolism 2.3%, all bleeding 1.0%, and major paravalvular leak 0.1%. One- and 4-year actuarial freedom from all-cause mortality was 97.7% (95% confidence interval, 96.5%-98.8%) and 91.9% (95% confidence interval, 89.7%-94.1%), respectively. At 4 years, New York Heart Association functional class improved compared with baseline in 63.0%, effective orifice area was 1.5 ± 0.5 cm2, and mean gradient was 11.0 ± 5.6 mm Hg. Freedom from moderate or greater transvalvular insufficiency was 99.7%. There were no events of structural valve deterioration.
CONCLUSIONS: Safety and hemodynamic performance of this aortic bioprosthesis with RESILIA tissue at 4 years are favorable. This novel tissue does not appear to result in unexpected early thrombosis events or noncalcific structural valve deterioration.
Copyright © 2020. Published by Elsevier Inc.

Entities:  

Keywords:  aortic valve replacement, bioprosthetic valve, RESILIA, tissue

Year:  2020        PMID: 32340804     DOI: 10.1016/j.jtcvs.2020.01.095

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  2 in total

1.  Surgical aortic valve replacement with a stented pericardial bioprosthesis: 5-year outcomes.

Authors:  Robert J M Klautz; François Dagenais; Michael J Reardon; Rüdiger Lange; Michael G Moront; Louis Labrousse; Neil J Weissman; Vivek Rao; Himanshu J Patel; Fang Liu; Joseph F Sabik
Journal:  Eur J Cardiothorac Surg       Date:  2022-08-03       Impact factor: 4.534

2.  Durability of bioprosthetic aortic valves in patients under the age of 60 years - rationale and design of the international INDURE registry.

Authors:  Bart Meuris; Michael A Borger; Thierry Bourguignon; Matthias Siepe; Martin Grabenwöger; Günther Laufer; Konrad Binder; Gianluca Polvani; Pierluigi Stefano; Enrico Coscioni; Wouter van Leeuwen; Philippe Demers; Francois Dagenais; Sergio Canovas; Alexis Theron; Thierry Langanay; Jean-Christian Roussel; Olaf Wendler; Giovanni Mariscalco; Renzo Pessotto; Beate Botta; Peter Bramlage; Ruggero de Paulis
Journal:  J Cardiothorac Surg       Date:  2020-05-27       Impact factor: 1.637