Caiming Tang1, Jianhua Tan2, Yujuan Fan1,3, Xianzhi Peng1. 1. State Key Laboratory of Organic Geochemistry, Guangzhou Institute of Geochemistry, Chinese Academy of Sciences, Guangzhou 510640, China. 2. Guangzhou Quality Supervision and Testing Institute, Guangzhou 510110, China. 3. University of Chinese Academy of Sciences, Beijing 100049, China.
Abstract
H-Abstraction reactions occurring during electron impact ionization processes in electron ionization mass spectrometry (EI-MS) are a long-standing and crucial topic in MS research. Yet, some critical relevant mechanisms are controversial and ambiguous, and information about the EI-induced H-abstraction reactions of halogenated organic compounds (HOCs) is completely in the dark. This study provides a systematic investigation of H-abstraction reactions of HOCs taking place in the EI source using 13C6-hexachlorobenzene (13C6-HCB) and 13C6-hexabromobenzene (13C6-HBB) as exemplary compounds by gas chromatography (GC)-high-resolution mass spectrometry (GC-HRMS). The H-abstraction efficiencies were evaluated with the MS signal intensity ratios of ions with H-abstraction relative to the corresponding original ions (without H-abstraction). Ion source temperatures, EI energies, and numbers of heavy isotope atoms (37Cl or 81Br) of isotopologues were investigated in terms of their effects on the H-abstraction efficiencies. The H-abstraction efficiencies of individual isotopologues generally decreased from the first to the last isotopologues of respective ions, and those of individual ions were different from each other, with the highest values of 0.017 and 0.444 for 13C6-HCB and 13C6-HBB, respectively. The overall H-abstraction efficiencies involving all measured ions of 13C6-HCB and 13C6-HBB were 0.004 and 0.128, respectively. With increasing ion source temperatures, the H-abstraction efficiencies first increased to a summit and then began to linearly decrease. EI energies and emission currents could impact the H-abstraction efficiencies but showed no certain tendency. The H-abstraction reactions were inferred to belong to ion-molecule reactions, and the siloxanes bleeding from the GC column might be a hydrogen source. Some strategies were proposed for eliminating or alleviating the interference triggered by the H-abstraction reactions in EI-MS in identification of halogenated organic pollutants (HOPs). Our findings provide a better understanding of the EI-induced H-abstraction reactions of HOCs and may benefit the identification of HOPs in environmental analysis, especially for novel HOPs.
H-Abstraction reactions occurring during electron impact ionization processes in electron ionization mass spectrometry (EI-MS) are a long-standing and crucial topic in MS research. Yet, some critical relevant mechanisms are controversial and ambiguous, and information about the EI-induced H-abstraction reactions of n class="Chemical">halogenated organic compounds (n class="Chemical">HOCs) is completely in the dark. This study provides a systematic investigation of H-abstraction reactions of HOCs taking place in the EI source using 13C6-hexachlorobenzene (13C6-HCB) and 13C6-hexabromobenzene (13C6-HBB) as exemplary compounds by gas chromatography (GC)-high-resolution mass spectrometry (GC-HRMS). The H-abstraction efficiencies were evaluated with the MS signal intensity ratios of ions with H-abstraction relative to the corresponding original ions (without H-abstraction). Ion source temperatures, EI energies, and numbers of heavy isotope atoms (37Cl or 81Br) of isotopologues were investigated in terms of their effects on the H-abstraction efficiencies. The H-abstraction efficiencies of individual isotopologues generally decreased from the first to the last isotopologues of respective ions, and those of individual ions were different from each other, with the highest values of 0.017 and 0.444 for 13C6-HCB and 13C6-HBB, respectively. The overall H-abstraction efficiencies involving all measured ions of 13C6-HCB and 13C6-HBB were 0.004 and 0.128, respectively. With increasing ion source temperatures, the H-abstraction efficiencies first increased to a summit and then began to linearly decrease. EI energies and emission currents could impact the H-abstraction efficiencies but showed no certain tendency. The H-abstraction reactions were inferred to belong to ion-molecule reactions, and the siloxanes bleeding from the GC column might be a hydrogen source. Some strategies were proposed for eliminating or alleviating the interference triggered by the H-abstraction reactions in EI-MS in identification of halogenated organic pollutants (HOPs). Our findings provide a better understanding of the EI-induced H-abstraction reactions of HOCs and may benefit the identification of HOPs in environmental analysis, especially for novel HOPs.
n class="Chemical">Hydrogen
abstractionpan> (H-abstractionpan>) reactionpan>s (or intermolecular
n class="Chemical">hydrogen transfer reactions), which may have important significance
for mass spectrometry (MS) analysis, can commonly take place in various
MS ionization sources, such as electron ionization (EI),[1,2] chemical ionization (CI),[3] electrospray
ionization (ESI),[4−8] atmospheric pressure chemical ionization (APCI),[9,10] matrix-assisted
laser desorption ionization,[11] atmospheric
pressure photoionization (APPI),[12−14] and low-pressure photoionization
(LPPI) sources.[15−18] The mechanisms of H-abstraction reactions on CI, ESI, APCI, and
photoionization (PI) sources have been well studied.[3,5,6,9,13] However, the real mechanisms of H-abstraction
reactions in EI-MS remain far from being completely explicit although
the H-abstraction phenomenon has been observed and investigated for
a very long time (2–5 decades).[1,2,19−24]
It was reported that the generation of molecular ions via
H-abstraction
in the EI source, such as [M + H]+, [M + n class="Chemical">2H]+, and [M + n class="Chemical">3H]+, might be related to the structures of
analytes and the ion source temperatures.[23,24] Some studies concluded that the transferred hydrogen atoms in H-abstraction
reactions in the EI source were derived from H2O in the
ion source and/or samples.[1,2,21,25,26] However, a study revealed that the hydrogen atoms did not stem from
residual water in the ion source, nor residual H2.[23] The author inferred that the hydrogen source
was related to thermal initiation and EI ionization.[23] Most of the previous studies indicated that the H-abstraction
reactions in the EI source belonged to ion–molecule reactions,[1,2,19,22−27] which can be expressed as
This reaction equation is similar to that of some H-abstraction
reactions occurring in the PI source.[9,28−31] The initial ionization process of PI is[9,29,30]and that of EI isIt can be seen from eqs and 3 that
the initial ionization processes
of PI and EI are to some extent similar. As a result, the revealed
mechanisms behind the H-abstraction reactions occurring in the PI
source may provide illuminating insights into the H-abstraction fundamentals
in the EI source. In the PI source, the n class="Chemical">hydrogen source for H-abstractionpan>
reactionpan>s is protic solvents used for injectionpan>, such as n class="Chemical">toluene,[29] water,[32] methanol,[32] ethanol and hexane,[28] whereas aprotic solvents such as tetrachloromethane and aprotic
gases including N2, O2, and CO2 could
compete with analytes to abstract hydrogen atoms, thus impairing the
MS signal intensities of [M + H]+.[28] In addition to solvents, the temperature, ionizer voltage, and reaction
length of the PI source also play critical roles in the H-abstraction
reactions,[28,29,33,34] showing positive correlations with the H-abstraction efficiencies.
Furthermore, the proton affinities of analytes and the reaction enthalpy
changes (ΔH) are also important factors affecting
the H-abstraction reactions in the PI source, with a positive correlation
with the H-abstraction efficiencies.[28,34]
To date,
reported studies concerning H-abstraction reactions in
the EI source have mainly focused on n class="Chemical">quinones.[20,21] However, n class="Chemical">halogenated organic compounds (HOCs), e.g., organochlorines
and organobromines, may also abstract hydrogen atoms in the EI source,
in view of their lone electron pairs of halogen atoms (e.g., Cl and
Br). HOCs are a large category of chemicals produced from both natural
and anthropogenic activities[35−38] and have enormous influences on human beings and
the environment, in either positive or negative ways.[39−43] Many HOCs are persistent, bioaccumulative, and toxic pollutants,
such as polychlorinateddioxins/furans,[44−46] dichlorodiphenyltrichloroethanes
(DDTs),[47−49] polychlorinated biphenyls,[45,50] polybrominated diphenyl ethers,[48,49] polybrominated
biphenyls,[51] chlorinated paraffins,[52] perfluorinated compounds,[53] and etc., and have been posing serious pollution problems
to the environment. In addition to known halogenated organic pollutants
(HOPs), numerous previously unknown, novel, and emerging HOPs have
been screened and identified in various matrices.[54−58] Most HOPs can be identified by gas chromatography
coupled with EI-MS (GC-EI-MS), which is the mainstream approach for
analysis of HOPs.[59] Recently, we successfully
developed a quasi-targeted analysis method for identification of novel
HOPs in environmental matrices using GC-EI-HRMS.[60] Additionally, some non-hyphenated MS techniques such as
EI-FTICR-HRMS have been applied to the identification of HOPs.[61] Nevertheless, anticipatable H-abstraction reactions
in EI-MS may introduce plausible interference peaks and could therefore
compromise the identification of HOPs, especially non-targeted or
quasi-targeted analysis of unknown and novel pollutants. Accordingly,
the H-abstraction reactions of HOCs in EI-MS require systemic investigation
to facilitate the screening and identification of these compounds.
Here, we conducted an in-depth study concerning the H-abstraction
reactions of n class="Chemical">HOCs occurring during the electronpan> impact ionpan>izationpan>
process in EI-MS using GC double-focusing magnetic-sector HRMS (GC-DFS-HRMS),
with two stable isotope-labeled standards, i.e., n class="Chemical">13C6-hexachlorobenzene (13C6-HCB) and 13C6-hexabromobenzene (13C6-HBB), used as model compounds. The H-abstraction efficiencies were
calculated as the ratios of MS signal intensities between ions with
H-abstraction and the corresponding original ions (without H-abstraction),
and the effects of ion source temperatures, EI energies ,and numbers
of heavy isotope atoms (37Cl or 81Br) of isotopologues
on the H-abstraction efficiencies were investigated. This study reveals
some details of the H-abstraction reactions of HOCs in EI-MS, provides
new insights into the possible mechanisms of the EI-induced H-abstraction
reactions, and may be conducive to non-targeted and quasi-targeted
qualitative analyses of HOPs, particularly novel HOPs.
Experimental
Section
Chemicals and Materials
Stock solutions of stable isotope-labeled
standards, n class="Chemical">13C6-HCB (100 μg/mL in n class="Chemical">nonane;
isotopic purity, ≥99%) and 13C6-HBB (100
μg/mL in toluene; isotopic purity, ≥99%), were purchased
from Cambridge Isotope Laboratories Inc. (Andover, MA, USA). Structures
and the CAS numbers of the standards are documented in Table S1. Chromatography-grade isooctane was
purchased from CNW Technologies GmbH (Düsseldorf, Germany).
Acetone and n-hexane used for washing the autosampler
injection syringe were bought from Merck Corp. (Darmstadt, Germany).
Perfluorotributylamine used as a reference standard for HRMS calibration
was purchased from Sigma-Aldrich LLC. (St. Louis, MO, USA). The purchased
stock standards of 13C6-HCB and 13C6-HBB were further diluted with isooctane to prepare
working solutions at concentrations of 5 and 1 μg/mL, respectively.
All the standard solutions were kept in a fridge at −20 °C
prior to use.
GC–HRMS Measurements
The
GC–HRMS system
consisted of dual gas chromatographs (Trace-GC-Ultra) coupled to a
double-focusing magnetic-sector HRMS (BE geometry) and an extra TriPlus
autosampler (GC-DFS-HRMS, Thermo-Fisher Scientific, n class="Chemical">Bremenpan>, Germany).
The prepared working solutionpan>s were directly analyzed by the GC–HRMS.
The chromatographic separationpan> was conpan>ducted with a DB-5MS capillary
column (60 m × 0.25 mm, 0.25 μm film thickness, J&W
Scienpan>tific, Folsom, USA).
The GC temperature-ramping program
is as follows: The temperature was held at 120 °C for 2 min,
ramped to 220 °C at 20 °C/min, held for 16 min; then ramped
to 235 °C at 5 °C/min, held for 7 min; then ramped to 260
°C at 5 °C/min; and finally ramped to 310 °C at 40
°C/min and held for 0.75 min. The total time of a GC run was
40 min. The solvent delay time was 6.5 min. Ultrahigh-purity n class="Chemical">helium
was used as the carrier gas, and a conpan>stant flow rate of 1.0 mL/min
was maintained. The GC injectionpan> port and transfer line were kept
at 260 and 280 °C, respectively. Splitless injectionpan> mode was
applied, and the injectionpan> volume was 1 μL.
The working
settings and parameters of the DFS-HRMS are provided
as follows: The EI source operating in positive mode was applied;
EI energies were set at 35, 45, 55, and 65 eV; ion source temperatures
were set at 160, 190, 220, 250, 280, and 300 °C; filament emission
currents ranged from 0.463 to 0.989 mA with the corresponding EI energies
from 35 to 65 eV; data acquisition was performed in multiple ion detection
(MID) mode; the dwell time was 20 ms for each isotopologue; scan time
segments were set at 8–20 min (for n class="Chemical">13C6-HCB) and 20–36 min (for n class="Chemical">13C6-HBB);
mass resolution (5% peak definition) was ≥10,000 and the MS
detection accuracy was set at ±1 mu. If not specified, all measurements
were performed with an EI energy of 45 eV and ion source temperature
of 250 °C. The HRMS real-time calibration was carried out with
perfluorotributylamine during MID operation.
Chemical structures
of the investigated compounds along with their
dechlorination/den class="Chemical">brominationpan> radicals were depicted with ChemDraw
(Ultra 7.0, Cambridgesoft), and the exact masses of the chlorine/bromineisotopologues were calculated with a mass accuracy of 0.00001 u. Since
the carbon of 13C6-HCB and 13C6-HBB comprises only the 13C isotope, only the chlorine/bromineisotopologues were studied. Hydrogen (1H) was the only
involved hydrogen isotope since the others (deuterium and tritium)
have extremely lower abundances in nature relative to hydrogen. For
a molecule or radical containing n Cl/Br atoms, all
the isotopologues (n + 1) were chosen. The exact
mass-to-charge ratios (m/z) of ions were calculated
by subtracting the exact mass of one electron from that of individual
isotopologues. The m/z values of ions with H-abstraction
were obtained by adding the exact mass of a hydrogen atom to that
of each original ion, and only the H-abstraction reactions transferring
merely one hydrogen atom were taken into consideration. These m/z values were imported into the MID module for detection.
The detailed information including chemical formulas, exact masses,
and exact m/z values of the isotopologues of the
investigated compounds is listed in Table S2, and the representative chromatograms along with high-resolution
mass spectra of the compounds are shown in Figure .
Figure 1
Representative chromatograms and high-resolution
mass spectra of 13C6-hexachlorobenzene (13C6-HCB) and 13C6-hexabromobenzene
(13C6-HBB). Representative chromatograms of (a) 13C6-HCB and (b) 13C6-HBB; representative mass spectra of (c) 13C6-HCB and (d) 13C6-HBB.
P-Cl: dechlorination product ion possessing n Cl atom(s); P-Br: debromination
product ion possessing n Br atom(s); NL: nominal
level; m/z: mass to charge ratio.
Representative chromatograms and high-resolution
mass spectra of n class="Chemical">13C6-hexachlorobenzene (n class="Chemical">13C6-HCB) and 13C6-hexabromobenzene
(13C6-HBB). Representative chromatograms of (a) 13C6-HCB and (b) 13C6-HBB; representative mass spectra of (c) 13C6-HCB and (d) 13C6-HBB.
P-Cl: dechlorination product ion possessing n Cl atom(s); P-Br: debromination
product ion possessing nBr atom(s); NL: nominal
level; m/z: mass to charge ratio.
Data Processing
The H-abstraction efficiency of an
isotopologue is expressed as the MS signal intensity ratio of the
isotopologue with H-abstraction relative to the original isotopologue
(R) and was calculated usingwhere i is
the number of n class="Chemical">37Cl or n class="Chemical">81Br atom(s) in the isotopologue, I′ is the MS signal
intensity of isotopologue i with H-abstraction, I is the MS signal intensity of the original
isotopologue i. The whole H-abstraction efficiency
of an ion (R) was calculated withwhere n is
the number of Cl/Br atoms of the ion. In addition, the overall H-abstraction
efficiency involving all the detected ions of a compound (Rall) was calculated with
Prior to exporting
MS signal intensities, background subtraction was carried out by deducting
the baseline signal intensities close to both ends of individual chromatographic
peaks. If n class="Species">not specified, the number of injectionpan> replicates was six,
and the experimental data derived from these replicated measurements
were applied to calculationpan> of average H-abstractionpan> efficiencies
and standard deviationpan>s (1σ). In order to describe this work
more conpan>cisely and clearly, we use IST-n to refer
to an isotopologue with n – 1 n class="Chemical">37Cl or 81Br atom(s), apply P-Cl to denote a dechlorination product ion having n Cl atom(s), and utilize P-Br to represent
a debromination product ion containing nBr atom(s).
Because IST-1 of P-Cl1 with H-abstraction of 13C6-HCB ([13C6H35Cl]+, m/z: 113.99626) was seriously interfered
by the [C2NF4]+ ion (m/z: 113.99614) of perfluorotributylamine, the H-abstraction efficiency
of IST-1 was not calculated.
Method Performances
As documented in Tables S3– S5,
the precisions (standard deviations)
of the H-abstraction efficiencies of individual n class="Chemical">isotopologues, the
whole H-abstractionpan> efficienpan>cies of individual ionpan>s, and the overall
H-abstractionpan> efficienpan>cies of individual compounds are within the
ranges of 0.00002–0.01608, 0.00011–0.00871, and 0.00015–0.00511,
respectively. These results demonpan>strate that the analysis method could
meet the requiremenpan>ts for investigating the H-abstractionpan> efficienpan>cies
of n class="Chemical">13C6-HCB and 13C6-HBB
in EI-MS.
Statistical Analysis
Statistical analysis was conducted
with SPSS Statistics 19.0 (IBM Inc., Armonk, USA) and Origin 9 (OriginLab
Corp., n class="Chemical">Northamptonpan>, USA). The paired-sample T test
and independent-sample T test (performed with SPSS)
were applied to determine the p values (two-tailed)
with an α of 0.01 as the threshold value for significance. The
differences of the H-abstractionpan> efficiencies and MS signal intensities
amonpan>g n class="Chemical">isotopologues were examined with the paired-sample T test, and those of the whole H-abstraction efficiencies among ions
and the overall H-abstraction efficiencies between the two compounds
were evaluated with the independent-sample T test.
If a p value is less than 0.01, the null hypothesis
(e.g., no difference between two groups of H-abstraction efficiencies)
is rejected, indicating that there is indeed a significant difference.
Linear and non-linear regressions (conducted with Origin) were employed
to reveal the relationships between experimental data and parameters/factors
of interest.
Results and Discussion
H-Abstraction Efficiency
Measurement
Isotope-labeled n class="Chemical">13C6-HCB
and n class="Chemical">13C6-HBB instead
of native HCB and HBB were chosen to perform this study owing to the
fact that 13C6-HCB and 13C6-HBB contain only 13C and Cl/Br atoms and therefore the
interference caused by carbon isotopes is nonexistent. In addition,
as the isotopic purities of the two reference standards were higher
than 99%, potential impurities were thus deduced to contribute negligible
interferences to measured data. Based on the detected MS signal intensities
of the original ions and those with H-abstraction, we calculated three
hierarchical H-abstraction efficiencies, i.e., H-abstraction efficiencies
of individual isotopologues, whole H-abstraction efficiencies of individual
ions, and overall H-abstraction efficiencies of individual compounds.
Thus, H-abstraction efficiencies of the investigated compounds were
evaluated at three levels, namely, isotopologues, ions, and compounds.
H-Abstraction
Efficiencies of Individual Isotopologues
As shown in Figure and Table S3, the first six n class="Chemical">chlorine
n class="Chemical">isotopologues of the molecular ion of 13C6-HCB,
i.e., IST-1 to IST-6, have H-abstraction efficiencies within the range
from 0.00010 ± 0.00009 to 0.0026 ± 0.0003, exhibiting a
gradual decline from the first to the last isotopologues (Figure a). For the first
five chlorine isotopologues of P-Cl5 (IST-1 to IST-5),
the H-abstraction efficiencies were from 0.016 ± 0.001 to 0.0184
± 0.0003, showing a generally decreasing tendency from IST-1
to IST-5 (Figure b).
The H-abstraction efficiencies of individual chlorine isotopologues
of the other four dechlorination product ions (P-Cl1 to
P-Cl4) ranged from 0.0020 ± 0.0001 to 0.010 ±
0.001 (Figure c–f,
respectively), presenting no explicit variation trend along the isotopologue
sequences with statistical significance (Figure c–e). The H-abstraction efficiencies
of IST-1 of the molecular ion and P-Cl5 were significantly
higher than those of other isotopologues of the respective ions (p ≤ 0.001). The H-abstraction efficiency difference
between any two neighboring chlorine isotopologues of the molecular
ion was statistically significant (p ≤ 0.008),
and that between any two neighboring isotopologues of the first three
isotopologues of P-Cl5 was also statistically significant
(p ≤ 0.001).
Figure 2
H-abstraction efficiencies ([M + H]+/[M]·+ or [P + H]+/[P]+ MS signal intensity
ratios) of chlorine/bromine isotopologues of molecular and dechlorination/debromination
product ions of 13C6-HCB and 13C6-HBB during the electron impact ionization process in electron
ionization high-resolution mass spectrometry (EI-HRMS). H-Abstraction
efficiencies of chlorine isotopologues of the (a) molecular ion, (b) P-Cl5, (c) P-Cl4, (d) P-Cl3, (e) P-Cl2, and (f) P-Cl1 of 13C6-HCB; H-abstraction efficiencies of bromine
isotopologues of the (g) molecular ion, (h) P-Br5, (i) P-Br4, (j) P-Br3, (k) P-Br2, and (l) P-Br1 of 13C6-HBB. [M]·+: molecular ion; [M + H]+: the corresponding
molecular ion with hydrogen abstraction; [P]+: dehalogenation
product ion; [P + H]+: the corresponding product ion with
H-abstraction; IST-n: isotopologue of an ion possessing n – 1 37Cl or 81Br atom(s).
Error bars denote the standard deviations (1σ, n = 6).
H-abstraction efficiencies ([M + H]+/[M]·+ or [P + H]+/[P]+ MS signal intensity
ratios) of n class="Chemical">chlorine/n class="Chemical">bromine isotopologues of molecular and dechlorination/debromination
product ions of 13C6-HCB and 13C6-HBB during the electron impact ionization process in electron
ionization high-resolution mass spectrometry (EI-HRMS). H-Abstraction
efficiencies of chlorine isotopologues of the (a) molecular ion, (b) P-Cl5, (c) P-Cl4, (d) P-Cl3, (e) P-Cl2, and (f) P-Cl1 of 13C6-HCB; H-abstraction efficiencies of bromineisotopologues of the (g) molecular ion, (h) P-Br5, (i) P-Br4, (j) P-Br3, (k) P-Br2, and (l) P-Br1 of 13C6-HBB. [M]·+: molecular ion; [M + H]+: the corresponding
molecular ion with hydrogen abstraction; [P]+: dehalogenation
product ion; [P + H]+: the corresponding product ion with
H-abstraction; IST-n: isotopologue of an ion possessing n – 1 37Cl or 81Br atom(s).
Error bars denote the standard deviations (1σ, n = 6).
The H-abstraction efficiencies
of the n class="Chemical">bromine n class="Chemical">isotopologues of
the molecular ion of 13C6-HBB decreased continuously
from IST-1 (0.222 ± 0.009) to IST-7 (0.0005 ± 0.0005) (Figure g), and those of
any neighboring isotopologues showed statistically significant differences
(p ≤ 0.00001). Likewise, the product ions
P-Br5, P-Br4, P-Br3, and P-Br2 of 13C6-HBB presented gradually descending
H-abstraction efficiencies from the first isotopologues to the last
ones, with H-abstraction efficiency ranges of 0.424 ± 0.016 to
0.569 ± 0.006, 0.111 ± 0.007 to 0.23 ± 0.01, 0.173
± 0.003 to 0.195 ± 0.003, and 0.041 ± 0.001 to 0.059
± 0.005, respectively (Figure h–k). For these product ions, the H-abstraction
efficiencies of the bromineisotopologues of each ion (except IST-4,
IST-5, and IST-6 of P-Br5) can be confidently differentiated
from each other (p ≤ 0.002). In addition,
there was a slight decline for the H-abstraction efficiencies from
IST-1 to IST-2 of P-Br1 (p ≤ 0.004)
(Figure l).
Whole
H-Abstraction Efficiencies of Individual Ions
As shown in Figure a, the magnitudes
of measured whole H-abstraction efficiencies of
individual ions of n class="Chemical">13C6-HCB follow the order
n class="Chemical">P-Cl5 > P-Cl1 > P-Cl4 >
P-Cl3 > P-Cl2 > molecular ion, with the
range from 0.0011
± 0.0001 to 0.0172 ± 0.0002 (Table S4). These H-abstraction efficiencies are statistically distinguishable
(p ≤ 0.00004). The scale of the whole H-abstraction
efficiencies of individual ions of 13C6-HBB
showed the following order: P-Br5 > P-Br3 >
P-Br1 > P-Br4 > P-Br2 >
molecular
ion (Figure b), with
the range from 0.037 ± 0.001 to 0.444 ± 0.009 (Table S4). These H-abstraction efficiencies can
also be confidently differentiated (p ≤ 0.0003).
The ion P-Br5 presented an extremely high H-abstraction
efficiency (0.444 ± 0.009), suggesting that around 30.7% of the
amount of the initial ion underwent the H-abstraction reaction. It
is noteworthy that the patterns of the whole H-abstraction efficiencies
of ions of the two compounds are somewhat similar, probably implying
similar mechanisms of H-abstraction reactions for the detected ions
between the two compounds in EI-MS.
Figure 3
Whole H-abstraction efficiencies of individual
ions of (a) 13C6-HCB and (b) 13C6-HBB, and overall H-abstraction
efficiencies of (c) 13C6-HCB and 13C6-HBB in EI-HRMS.
Whole H-abstraction efficiencies of individual
ions of (a) n class="Chemical">13C6-HCB and (b) n class="Chemical">13C6-HBB, and overall H-abstraction
efficiencies of (c) 13C6-HCB and 13C6-HBB in EI-HRMS.
Overall H-Abstraction Efficiencies of Individual Compounds
As shown in Figure c and Table S5, the overall H-abstraction
efficiency of n class="Chemical">13C6-HBB (0.128 ± 0.005)
was more than onpan>e order of magnitude higher than that of n class="Chemical">13C6-HCB (0.0044 ± 0.0001), with the p value <0.00001. This result indicates that 13C6-HBB has a significantly higher H-abstraction capability than 13C6-HCB in EI-MS.
Distribution Patterns of
Measured MS Signal Intensities of Molecular
Isotopologues
As shown in in Figure and Table S6,
it is very surprising that the distribution patterns of the measured
MS signal intensities of n class="Chemical">isotopologues between the molecular ionpan> ([M]·+) and the molecular ionpan> with H-abstractionpan>
([M + H]+) of each compound are apparently different. For
the [M]·+ of n class="Chemical">13C6-HCB, the measured MS signal intensities were in the following order:
IST-2 > IST-3 > IST-1 > IST-4 > IST-5 > IST-6 >
IST-7 (Figure a),
which is identical to the
theoretical relative abundance order. Meanwhile, the detected isotopologue
abundances of the [M + H]+ of 13C6-HCB presented the order IST-1 > IST-2 > IST-3 > IST-4 >
IST-5 >
IST-6 > IST-7 (Figure b). The detected abundance of IST-1 of [M + H]+ was significantly
higher than those of IST-2 and IST-3 (p ≤
0.00005), which contradicts the scenario of [M]·+ and therefore the order of theoretical relative abundances
of isotopologues.
Figure 4
Distribution patterns of MS signal intensities of chlorine/bromine
isotopologues of the molecular ions ([M]·+) and the molecular ions with hydrogen abstraction ([M + H]+) of 13C6-HCB and 13C6-HBB on GC-EI-HRMS. Intensities of chlorine isotopologues of (a) [M]·+ and (b) [M + H]+ of 13C6-HCB; intensities of bromine isotopologues of (c) [M]·+ and (d) [M + H]+ of 13C6-HBB.
Distribution patterns of MS signal intensities of n class="Chemical">chlorine/n class="Chemical">bromine
isotopologues of the molecular ions ([M]·+) and the molecular ions with hydrogen abstraction ([M + H]+) of 13C6-HCB and 13C6-HBB on GC-EI-HRMS. Intensities of chlorine isotopologues of (a) [M]·+ and (b) [M + H]+ of 13C6-HCB; intensities of bromineisotopologues of (c) [M]·+ and (d) [M + H]+ of 13C6-HBB.
In addition to n class="Chemical">13C6-HCB, n class="Chemical">13C6-HBB also showed different
measured isotopologue abundance
orders between [M]•+ and [M + H]+. For
[M]•+, the order of the measured isotopologue abundances
was IST-4 > IST-3 > IST-5 > IST-2 > IST-6 > IST-1 >
IST-7 (Figure c),
which is well
consistent with the normal distribution and the theoretical relative
abundance order. However, the measured isotopologue abundances of
[M + H]+ exhibited the following order: IST-3 > IST-2
>
IST-4 > IST-1 > IST-5 > IST-6 > IST-7 (Figure d), which is completely different
from that
of [M]•+ and accordingly the theoretical relative
abundance order. These findings also point to the significant differences
of H-abstraction efficiencies among different chlorine/bromineisotopologues
in EI-MS.
Parameters and Factors Affecting H-Abstraction Efficiency Variation
To investigate whether some parameters can influence the H-abstraction
efficiencies in EI-MS, we measured the H-abstraction efficiencies
of the molecular ion of n class="Chemical">13C6-HCB with varied
ionpan> source temperatures and EI energies. Some previous studies have
proved that the ionpan> source temperature could dramatically impact the
extents of H-abstractionpan> reactionpan>s in EI-MS, positively correlating
with the reactionpan> extents.[23,24] This observationpan> could
be attributable to the fact that the average translationpan>al energies
of the n class="Chemical">analyte ions and hydrogen donor molecules increase with rising
temperature and therefore enhance the collision rate, accordingly
raising the H-abstraction reaction extents. In the present study,
however, we observed a different variation mode of H-abstraction efficiencies
with ion source temperatures than those found in previous studies.
As shown in Figure a and Table S7, the measured H-abstraction
efficiencies first increased from 0.0071 ± 0.0008 to 0.0088 ±
0.0003 with the ion source temperatures changing from 160 to 190 °C
and then almost linearly decreased to 0.0008 ± 0.0001 with the
temperatures varying from 190 to 300 °C. As Ahmed et al. reported
that the relationship between the H-abstraction reaction extents and
the APPI source temperatures follows an Arrhenius-type curve,[29] we accordingly plotted the measured H-abstraction
efficiencies versus the reciprocal Kelvin temperatures. However, we
did not get the Arrhenius-type temperature-dependent curve describing
the relationship between the measured H-abstraction efficiencies and
the ion source temperatures. Instead, the H-abstraction efficiencies
first linearly increased from the lowest value to a summit with the
1/K values varying from 0.00174 to 0.00216 K–1 and
then began to decrease as the 1/K increased to 0.00231 K–1 (Figure b). This
observation also demonstrates that potential isotopic impurities (such
as HCB) in the reference standard of 13C6-HCB
could not interfere with the measured H-abstraction efficiencies.
Figure 5
Plots
of the whole H-abstraction efficiencies of the molecular
ion ([M]·+) of 13C6-HCB versus (a) the ion source temperatures, (b) the reciprocal Kelvin temperatures of ion
source, (c) the EI energies, and (d) the emission currents of EI filament, as well as
the correlations of (e) the H-abstraction
efficiencies versus the varied MS signal intensities caused by different
ion source temperatures and (f) the H-abstraction
efficiencies versus the varied MS signal intensities trigged by different
EI energies. The number of injection replicates was three.
Plots
of the whole H-abstraction efficiencies of the molecular
ion ([M]·+) of n class="Chemical">13C6-HCB versus (a) the ionpan> source temperatures, (b) the reciprocal Kelvin temperatures of ionpan>
source, (c) the EI enpan>ergies, and (d) the emissionpan> currenpan>ts of EI filamenpan>t, as well as
the correlationpan>s of (e) the H-abstractionpan>
efficienpan>cies versus the varied MS signal intenpan>sities caused by differenpan>t
ionpan> source temperatures and (f) the H-abstractionpan>
efficienpan>cies versus the varied MS signal intenpan>sities trigged by differenpan>t
EI enpan>ergies. The number of injectionpan> replicates was three.
As indicated in Figure ,d, the H-abstraction efficiencies can be significantly
influenced
by EI energies and emission currents of the ion source filament. At
an EI energy of 45 eV, the H-abstraction efficiency reached the highest
(0.0044 ± 0.0007), which was significantly higher than the rest
(p ≤ 0.003). Meanwhile, the H-abstraction
efficiency at an EI energy of 35 eV was the lowest (0.0007 ±
0.0002), which was significantly lower than those at other EI energies
(p ≤ 0.01). The H-abstraction efficiencies
at EI energies of 55 and 65 eV were relatively comparable (0.0014
± 0.0002 and 0.00168 ± 0.00001) and cannot be differentiated
with statistical significance (p = 0.2). We deduce
that internal energies of n class="Chemical">analyte ionpan>s increase as EI energies increase
and thus the rate of collisionpan> betweenn class="Chemical">analyte ions and hydrogen donor
molecules is enhanced, thus improving H-abstraction efficiencies.
Meanwhile, the increasing internal energies of [M]·+ ions further enhance internal energies of [M + H]+ ions, accordingly impairing the stability of the [M + H]+ ions. As a result, the H-abstraction efficiencies are inferred
to increase with EI energies at a lower energy region and decrease
in a higher-energy region, reaching a summit at medium EI energy,
which is consistent with the observation in this study (Figure c). Since the emission currents
generally increased with the increase of EI energies, the order of
H-abstraction efficiencies observed at different emission currents
were the same as that of the H-abstraction efficiencies at the corresponding
EI energies. However, the pattern of H-abstraction efficiencies plotted
versus EI energies is apparently different from that of H-abstraction
efficiencies versus emission currents (Figure c,d). This observation might be ascribed
to the fact that the emission currents varied with the variation of
EI energies more drastically in a lower-energy region than in a higher-energy
region.
As shown in Figure e, when the MS signal intensity variation was caused
by the variation
of ion source temperatures, a strong positive linear correlation between
H-abstraction efficiencies and MS signal intensities could be observed
(R2 = 0.969). n class="Chemical">Nevertheless, when the MS
signal intensity variationpan> was triggered by the variationpan> of EI energies,
onpan>ly a weak positive linear correlationpan> between H-abstractionpan> efficiencies
and MS signal intensities was found (R2 = 0.274, Figure f). It is noteworthy that the highest MS signal intensity and the
highest H-abstractionpan> efficiency were all obtained at an EI energy
of 45 eV, whereas at other EI energies (35, 55, and 65 eV), the MS
signal intensities and the H-abstractionpan> efficiencies were relatively
close (Table S7). These results demonpan>strate
that the MS signal intensity may be a parameter affecting the H-abstractionpan>
efficiencies of n class="Chemical">organochlorines on EI-MS.
As illustrated in Figure , the H-abstraction
efficiencies of n class="Chemical">isotopologues generally
decreased with the increase of heavy n class="Disease">isotope atoms (37Cl
or 81Br) of the isotopologues. The relationships between
the H-abstraction efficiencies and the numbers of 37Cl
or 81Br atom(s) were well fitted with exponential curves
for the molecular ions and the P-Cl5/P-Br5 of 13C6-HCB and 13C6-HBB, with
the R2 values ranging from 0.975 to 0.998. In addition
to the ions involved in Figure , the ions P-Br4, P-Br3 and P-Br2 also exhibited exponentially descending H-abstraction efficiencies
of bromineisotopologues with the increase of 81Br atoms
of the isotopologues, with R2 ≥ 0.982 (Figure S1). These findings suggest that chlorine/bromine
isotopic effects may play a crucial role in the variation of H-abstraction
efficiencies among chlorine/bromineisotopologues of HOCs on EI-MS.
Figure 6
Correlations
between H-abstraction efficiencies and the 37Cl or 81Br atom numbers of chlorine/bromine isotopologues.
(a) [M + H]+/[M]·+ intensity
ratios of the chlorine isotopologues of the molecular ion of 13C6-HCB; (b) [P + H]+/[P]+ intensity ratios of the chlorine isotopologues
of the P-Cl5 of 13C6-HCB; (c) [M + H]+/[M]·+ intensity ratios of the bromine isotopologues of the molecular ion
of 13C6-HBB; (d) [P
+ H]+/[P]+ intensity ratios of the bromine isotopologues
of the P-Br5 of 13C6-HBB. Solid curves
refer to exponential regressions and shaded areas represent the corresponding
95% confidence intervals.
Correlations
between H-abstraction efficiencies and the n class="Chemical">37Cl or n class="Chemical">81Br atom numbers of chlorine/bromineisotopologues.
(a) [M + H]+/[M]·+ intensity
ratios of the chlorine isotopologues of the molecular ion of 13C6-HCB; (b) [P + H]+/[P]+ intensity ratios of the chlorine isotopologues
of the P-Cl5 of 13C6-HCB; (c) [M + H]+/[M]·+ intensity ratios of the bromineisotopologues of the molecular ion
of 13C6-HBB; (d) [P
+ H]+/[P]+ intensity ratios of the bromineisotopologues
of the P-Br5 of 13C6-HBB. Solid curves
refer to exponential regressions and shaded areas represent the corresponding
95% confidence intervals.
Tentative Mechanistic Interpretation
Ion–Molecule Reactions
and Hydrogen Source
Previous
studies have concluded that the H-abstraction reactions in the EI
source are ion–molecule reactions,[1,2,19,22−27] which can be expressed as eq . Prior to the H-abstraction reactions, the molecules of n class="Chemical">analytes
are ionpan>ized as shown in eq , giving rise to molecular and fragmenpan>tal ionpan>s, which are
electronpan>-deficienpan>t radical ionpan>s. Inpan> light of the similar ionpan>izationpan>
processes of PI and EI as well as the reported outcomes,[28,29] we infer the most predominant ionpan>–molecule reactionpan>s are
that as shown by eq , instead of others such as
On
the other hand, eq is
energetically unfavorable
compared to eq , while eqs and 9 are low-probability events relative to eq due to the production of H· and H+ is not energetically favorable. In addition, we
found that not only the molecular ions but also the den class="Chemical">halogenationpan>
fragmental ionpan>s participated in the H-abstractionpan> reactionpan>s in the
EI source, and the fragmental ionpan>s exhibited evidently higher H-abstractionpan>
efficiencies than the molecular ionpan>s (Figure ). This result manifests that the “molecules”
in the ionpan>–molecule reactionpan>s were not the molecules of analytes,
and the “ions” in the reactions were the molecular and
fragmental ions of the analytes, which is consistent with eq . Then, the subsequent
questions are what are the real “molecules” in the ion–molecule
reactions and where do the hydrogen atoms stem from.
We inferred
that the reaction molecules might be n class="Chemical">siloxanes that
ran off from the packing materials (95% n class="Chemical">polydimethylsiloxane and 5%
phenyl group) of the GC column and/or emigrated from the silicone
inlet septum of the GC.[62] Therefore, the
H-abstraction reactions can be described aswhere R
represents a replaceable
siloxane group. It has been reported that the methyl of toluene prefers
to contribute hydrogen atoms for H-abstraction reactions in a PI source
in contrast with the phenyl group.[13] Accordingly,
it can be anticipated that the methyl of siloxanes is an excellent
hydrogen donor for the H-abstraction reactions in the EI source. In
conclusion, the siloxanes bleeding from the GC column and/or injection
port pad might participate in the H-abstraction reactions in the EI
source and could be a potential hydrogen source.
Another evidence
for the above conclusion is the difference of
overall H-abstraction efficiencies betweenn class="Chemical">13C6-HCB and n class="Chemical">13C6-HBB; i.e., 13C6-HCB presented significantly lower H-abstraction efficiency
compared to 13C6-HBB (Figure c). The electronegativity of chlorine is
higher than that of bromine; thus, the radical ions of 13C6-HCB are more competent to abstract hydrogen atoms from
hydrogen donors. However, in the present study, we found that the
overall H-abstraction efficiency of 13C6-HCB
is significantly lower than that of 13C6-HBB.
This implies that the densities of gaseous molecules of the hydrogen
donors during the H-abstraction reactions were different for 13C6-HCB and 13C6-HBB. If
not in this case, the H-abstraction efficiency of 13C6-HCB ought to be higher than that of 13C6-HBB. It can be inferred that the densities of gaseous molecules
of the hydrogen donors for H-abstraction reactions of 13C6-HCB were lower than those for H-abstraction reactions
of 13C6-HBB. This inference is in line with
the reality that 13C6-HCB eluted from the GC
column at a relatively lower temperature than 13C6-HBB and the higher oven temperature could cause more siloxanes to
bleed from the column.
Temperature-Dependent Reactions
In previous studies,
the H-abstraction efficiencies in the EI source were found to exponentially
correlate with the ion source temperatures,[23,24] which was also observed in another study concerning the H-abstraction
reactions in the PI source.[29] n class="Chemical">Nevertheless,
in our study, the H-abstractionpan> efficiencies rose first and then fell
with the increasing temperatures of the ionpan> source (Figure a). From 160 to 190 °C,
the H-abstractionpan> efficiencies rose to a summit, which is conpan>sistent
with the findings of the previous reports.[23,24,29] Meanwhile, the H-abstractionpan> efficiencies
gradually decreased to the lowest value from 190 to 300 °C, which
conpan>tradicts the outcomes of previous studies.[23,24,29] We deduce that the decline of H-abstractionpan>
efficiencies with the increase of ionpan> source temperature might be
attributable to the thermostability of [M + H]+. The pyrolysis
rate of [M + H]+ rises as the temperature increases. On
the other hand, the rate of H-abstractionpan> reactionpan>s generating [M
+ H]+ may also increase alonpan>g with the increase of ionpan>
source temperature. When the pyrolysis rate and generationpan> rate of
[M + H]+ are equivalent, the H-abstractionpan> efficiency reaches
the highest.
Chlorine and Bromine Isotopic Effects in
H-Abstraction Reactions
The ion–molecule H-abstraction
reactions of n class="Chemical">13C6-HCB and n class="Chemical">13C6-HBB in the EI source
may be analogous to the H-abstraction reactions of chlorine and bromine
radicals. We hypothesize a diatomic hydride HA (A is an imaginary
element), and the H-abstraction reaction between HA and Cl· in the gas phase is
The transition-state
complex is A--H--Cl. Meanwhile, taking the n class="Chemical">chlorine isotopes into
account, the transitionpan>-state complexes become n class="Chemical">A--H--35Cl and A--H--37Cl. Due to zero-point energies (ZPEs),
the bond H–35Cl is slightly weaker than the bond
H–37Cl. As a consequence, the energy barrier of
A--H--35Cl is slightly lower than that of A--H--37Cl. Then, the formation of A--H--35Cl is a little faster
than that of A--H--37Cl in the reaction as indicated in eq , thereby generating
more H35Cl in comparison with H37Cl. Extending
this inference to the ion–molecule H-abstraction reactions
of 13C6-HCB in the EI source, we conclude that
the production of [M + H]+ or [P + H]+ of the
compound was slightly higher for the lighter chlorine isotopologues
than for the heavier ones. Analogously, it can be deduced that the
yields of ions with H-abstraction in the EI source of 13C6-HBB were a little higher for the lighter bromineisotopologues
than for the heavier ones. These theoretical inferences are in agreement
with the experimental outcomes of this study (Figure and Figure S1).
In addition to the aforementioned activation energy differences
derived from the ZPE differences among n class="Chemical">chlorine/n class="Chemical">bromine isotopologues,
some other factors might also contribute to the H-abstraction efficiency
differences among isotopologues, including different average velocities
of isotopologues and on-column chlorine/bromine isotope fractionation.
In the mixture outflowing from the GC column, lighter isotopologues
have higher average velocities than heavier isotopologues since all
the isotopologues are at the same temperature condition. After electron
impact at the same EI energy, the lighter isotopologues are anticipated
to get more translational energy in contrast to the heavier isotopologues,
thus having higher average velocities. Additionally, after traversing
the electric field with the same accelerating voltage, the lighter
isotopologues get higher average velocities than the heavier ones.
The higher average velocities can lead to higher collisional rate
constants during the collision reactions between analyte ions and
siloxanes, resulting in higher H-abstraction efficiencies of the lighter
isotopologues relative to the heavier. Furthermore, as revealed in
our previous study, HCB and HBB presented large inverse chlorine/bromine
isotope fractionation on GC columns; viz., the lighter chlorine/bromineisotopologues outflowed from GC columns later than the heavier isotopologues.[63,64] On the other hand, the density of gas-phase siloxanes in the ion
source might be increasing from the beginning to the end of a GC run,
leading to more collision chances for the lighter isotopologues than
the heavier ones. As a consequence, the H-abstraction efficiencies
of chlorine/bromineisotopologues in the EI source decrease with the
increase of molecular weights of the isotopologues. In summary, all
the above factors can cause higher H-abstraction efficiencies for
the lighter chlorine/bromineisotopologues than the heavier ones,
which coincides with the observation in the current study (Figure and Figure S1). We deduce that these factors collaboratively
resulted in the significant discrepancies of H-abstraction efficiencies
among chlorine/bromineisotopologues.
It is noteworthy that
the correlations between the H-abstraction
efficiencies and the n class="Chemical">37Cl or n class="Chemical">81Br atom numbers
of chlorine/bromineisotopologues exhibit well-fitted Arrhenius-type
curves (Figure and Figure S1). This observation implies that the
H-abstraction reactions in the EI source were activation energy-dependent
(kinetically controlled) and the chlorine and bromine isotope effects
played a part in the reactions.
Implications for EI-MS
Analysis of HOPs
n class="Chemical">Nowadays, nonpan>-targeted
and quasi-targeted analyses as well as suspect screening have become
feasible identificationpan> methods for known and unknown HOPs.[60] The characteristic isotopologue distributionpan>s
are always applied to screening HOPs of interest.[60,65−67] The comparisonpan> between the theoretical isotopologue
distributionpan>s and the detected mass spectra of n class="Chemical">isotopologues can facilitate
the identification of HOPs, particularly novel HOPs, thus promoting
the reliability of identification results. Even so, it remains unclear
whether the H-abstraction reactions in the EI source can affect the
identification reliability of HOPs when isotopologue distribution
comparison is applied. For instance, the ion P-Cl5 with
H-abstraction (P-Cl5 + H) of 13C6-HCB has an identical chemical formula to that of the molecular ion
of 13C6-pentachlorobenzene, and thus they cannot
be differentiated by HRMS even though isotopologue distribution comparison
is employed (Table S8). Likewise, the ion
P-Br5 + H of 13C6-HBB cannot be distinguished
from the molecular ion of 13C6-pentabromobenzene
by HRMS in combination with isotopologue distribution comparison.
As a result, the dechlorination/debromination ions with H-abstraction
of the HOCs containing more chlorine/bromine atoms could trigger interference
in the analysis of those containing less chlorine/bromine atoms when
HRMS and isotopologue distribution comparison are employed, possibly
leading to false-positive identification results. Since the H-abstraction
efficiencies of some dechlorination/debromination ions are relatively
high (e.g., the P-Br5 of 13C6-HBB),
the interference may be considerable. Therefore, other separation
techniques such as GC and ion-mobility MS are necessary to couple
with HRMS for identification of HOCs, thus eliminating the interferences
and improving the validity of analysis results. However, in some scenarios,
the GC retention time ranges of HOC congeners containing different
Cl/Br atoms may partially overlap. Therefore, more data such as chlorine/bromine
isotopologue distribution patterns and quantitative structure–retention
relationships are warranted for getting rid of the possible interference
caused by H-abstraction reactions in EI-MS.[60] In addition, when using EI-MS data to evaluate the relative abundances
of carbonisotopologues and carbon isotope ratios of HOCs, analysts
should take into account the contribution of signals from the ions
with H-abstraction to the carbonisotopologues containing one 13C atom.
Conclusions
In this work, we systematically
investigated the H-abstraction
reactions of n class="Chemical">13C6-HCB and n class="Chemical">13C6-HBB in the EI source using GC–HRMS. The H-abstraction
efficiencies were assessed with the MS signal intensity ratios of
the ions with H-abstraction relative to the corresponding original
ions. Three levels of H-abstraction efficiencies, that is, H-abstraction
efficiency of each isotopologue, whole H-abstraction efficiency of
each ion, and overall H-abstraction efficiency of each compound, were
calculated. For most ions, the H-abstraction efficiencies of individual
isotopologues declined from the first to the last isotopologues. The
whole H-abstraction efficiencies of individual ions can be distinguished
from each other for individual compounds, and the overall H-abstraction
efficiency of 13C6-HCB was significantly lower
than that of 13C6-HBB. In addition, the isotopologue
distributions of measured MS signal intensities between [M]·+ and [M + H]+ for each compound were evidently
different, and those of the latter contradicted the distributions
of theoretical relative abundances of isotopologues. The impacts of
several instrumental parameters and factors including temperatures
of the ion source, EI energies, and numbers of 37Cl or 81Br atoms of isotopologues on the H-abstraction efficiencies
were revealed. The increase of ion source temperatures first enhanced
the H-abstraction efficiencies to a maximum and then began to linearly
reduce the efficiencies. The H-abstraction efficiencies were significantly
affected by the EI energies and emission currents and reached the
highest at an EI energy of 45 eV and emission current of 0.9 mA. The
mechanisms underlying the H-abstraction reactions of HOCs in the EI
source are tentatively proposed in light of the previous studies and
evidence found in the current study. The H-abstraction reactions might
pertain to ion–molecule reactions, and the siloxanes running
off from the GC column possibly participated in the reactions as a
hydrogen source. Furthermore, the H-abstraction reactions might be
kinetically controlled, and chlorine and bromine isotopic effects
were observed in the reactions. This study sheds light on the potential
mechanisms and magnitudes of the H-abstraction reactions of HOCs in
the EI source and will facilitate the qualitative analysis of HOPs
in environmental research, particularly for identification of novel
HOPs. In the future, it will be worthwhile to apply computational
quantum chemistry in unveiling the in-depth mechanisms behind the
EI-induced H-abstraction reactions of HOCs. The details of the possible
hydrogen donor and siloxanes bleeding from the GC column also merit
further investigation.
Authors: Sujan Fernando; Karl J Jobst; Vince Y Taguchi; Paul A Helm; Eric J Reiner; Brian E McCarry Journal: Environ Sci Technol Date: 2014-08-25 Impact factor: 9.028
Authors: Miren Pena-Abaurrea; Karl J Jobst; Ralph Ruffolo; Li Shen; Robert McCrindle; Paul A Helm; Eric J Reiner Journal: Environ Sci Technol Date: 2014-07-18 Impact factor: 9.028