Qing Huang1, Yuxia Liu2, Cuizhong Zhang1, Zhenfa Zhang1, Fengping Liu1, Jinyun Peng1. 1. College of Chemistry and Chemical Engineering, Guangxi Normal University for Nationalities, Chongzuo 532200, China. 2. College of Physics and Electronic Engineering, Guangxi Normal University for Nationalities, Chongzuo 532200, China.
Abstract
Owing to the importance of epinephrine as a neurotransmitter and hormone, sensitive methods are required for its detection. We have developed a sensitive photoelectrochemical (PEC) microsensor based on gold quantum dots (Au QDs) decorated on a nickel tetraminophthalocyanine-graphene oxide (NiTAPc-Gr) composite. NiTAPc was covalently attached to the surface of graphene oxide to prepare NiTAPc-Gr, which exhibits remarkable stability and PEC performance. In situ growth of Au QDs on the NiTAPc-Gr surface was achieved using chemical reduction at room temperature. The synthesized materials were characterized by Fourier transform infrared spectroscopy, ultraviolet-visible spectroscopy, X-ray photoelectron spectroscopy, scanning electron microscopy, transmission electron microscopy, and electrochemical impedance spectroscopy. Au QDs@NiTAPc-Gr provided a much greater photocurrent than NiTAPc-Gr, making it suitable for the ultrasensitive PEC detection of epinephrine. The proposed PEC strategy exhibited a wide linear range of 0.12-243.9 nM with a low detection limit of 17.9 pM (S/N = 3). Additionally, the fabricated PEC sensor showed excellent sensitivity, remarkable stability, and good selectivity. This simple, fast, and low-cost strategy was successfully applied to the analysis of human serum samples, indicating the potential of this method for clinical detection applications.
Owing to the importance of epinephrine as a neurotransmitter and hormone, sensitive methods are required for its detection. We have developed a sensitive photoelectrochemical (PEC) microsensor based on gold quantum dots (Au QDs) decorated on a nickel tetraminophthalocyanine-graphene oxide (NiTAPc-Gr) composite. NiTAPc was covalently attached to the surface of graphene oxide to prepare NiTAPc-Gr, which exhibits remarkable stability and PEC performance. In situ growth of Au QDs on the NiTAPc-Gr surface was achieved using chemical reduction at room temperature. The synthesized materials were characterized by Fourier transform infrared spectroscopy, ultraviolet-visible spectroscopy, X-ray photoelectron spectroscopy, scanning electron microscopy, transmission electron microscopy, and electrochemical impedance spectroscopy. Au QDs@NiTAPc-Gr provided a much greater photocurrent than NiTAPc-Gr, making it suitable for the ultrasensitive PEC detection of epinephrine. The proposed PEC strategy exhibited a wide linear range of 0.12-243.9 nM with a low detection limit of 17.9 pM (S/N = 3). Additionally, the fabricated PEC sensor showed excellent sensitivity, remarkable stability, and good selectivity. This simple, fast, and low-cost strategy was successfully applied to the analysis of human serum samples, indicating the potential of this method for clinical detection applications.
Epinephrine (EP), an important neurotransmitter
and hormone, can
improve the survival rate of cardiac arrestpatients by increasing
the force and rate of heart contractions.[1,2] However,
excess EP or subcutaneous injection of EP into a vein can be deadly,
as it can cause a sudden rise in blood pressure, cerebral hemorrhage,
or even ventricular fibrillation.[3,4] For this reason,
the use of EP in sports is banned by the World Anti-Doping Agency.
Therefore, it is very important to realize the ultrasensitive sensing
of EP. Some conventional detection techniques, such as colorimetry,[5] fluorescence spectrophotometry,[6] and high-performance liquid chromatography (HPLC),[7] have been successfully used for the determination
of EP. Recently, electrochemical analysis,[8] capillary electrophoresis,[9] and chemiluminescence[10] have been shown to be applicable to the detection
of EP. Furthermore, photoelectrochemical (PEC) analysis is a rapidly
developing method that can provide high precision, remarkable sensitivity,
and easy integration using simple equipment based on appropriate photoactive
materials.[11−15] Various organic semiconductors[16,17] and inorganic semiconductors[18,19] have been used owing
to their unique functions and photochemical activity.Phthalocyaninemetal derivatives have been widely used in applications in the spin-dyeing
industry, medicine, and electrocatalytic analysis owing to their attractive
features and functions.[20−22] Nickel tetraminophthalocyanine (NiTAPc) is a phthalocyanine metal
derivative well known for its excellent optical properties, chemical
stability, and low cost.[23,24] Moreover, high electrocatalytic
activity can be achieved owing to the large number of amino groups
in NiTAPc.[25] Significantly, NiTAPc shows
strong absorption in the region of 600–800 nm and high molar
extinction coefficients in the near-infrared region, which has been
exploited for PEC analysis.[26,27] Recently, graphene oxide
(GO) has been exploited in the field of PEC analysis.[28] GO, which is usually prepared from graphite by oxidation
using a strong acid, consists of sheets bearing carboxyl, hydroxyl,
and epoxy groups and exhibits photoelectric activity.[29] GO has good water solubility and excellent mechanical stability,
which make it suitable for use as a template to form composites with
various nanoparticles (NPs) or amino polymers for chemical analysis.[30−32] Previous studies have shown that
NiTAPc, which is rich in amino groups, can be covalently attached
to the surface of GO via chemical reaction with carboxyl groups to
form nickel tetraminophthalocyanine-functionalized graphene oxide
(NiTAPc-Gr).[33] NiTAPc-Gr exhibits the advantages
of both its constituent materials, showing not only excellent PEC
properties and high mechanical stability but also a large specific
surface area. This structure with an enhanced adsorption capacity
has been successfully used for supercapacitors and micromolecule detection.[34,35]As unique functional materials, metallic NPs have been widely
researched in various fields.[36−38] To date, metallic NPs have been extensively applied in photodetectors,[39] energy applications,[40] organic compound analysis,[41,42]and medical diagnosis
and therapy.[43] Notably, localized surface
plasmon resonance in Au quantum dots (QDs) can be excited, which facilities
the absorption of visible and even near-infrared photons, effectively
boosts the rate of electron–hole formation, and promotes the
separation of photogenerated charge carriers near the semiconductor,
which can be transformed into a strong and stable electrical signal.[44,45] Owing to their high specific surface area, excellent photocatalytic
activity, and good biocompatibility, Au QDs have been applied as efficient
light-harvesting enhancers in PEC analysis; for instance, TiO2-MoS2-Au NPs[46] and CdS-Au
QDs.[47] Au NPs integrated with ZnAgInS QDs[48] have been used for specific purposes. The development
of novel photoelectrodes is inevitable because of shortcomings during
the inception phase.In this work, we report the PEC analysis
of EP using a heterostructure composite consisting of Au QDs decorated
on NiTAPc-Gr (Au QDs@NiTAPc-Gr) as a signal indicator. This PEC sensor
has significant advantages: (1) The Au QDs@NiTAPc-Gr composite material
with specific structural features and a high specific surface area
was synthesized by coupling the photoactive template material NiTAPc-Gr
and a signal enhancer (Au QDs), which not only provided stability
and increased the photoelectric catalytic activity but also reduced
the self-aggregation of Au QDs, leading to an enhanced electrical
signal output. (2) The sensitivity of the PEC sensor is comparable
to that of enzymes; however, unlike enzyme sensors, the PEC sensor
does not suffer from inactivation. (3) The Au QDs@NiTAPc-Gr heterostructure
is ultrasensitive to visible and even near-infrared light, suggesting
that the PEC sensor has broad applicability for the clinical detection
of small molecules owing to the strong penetrability of the near-infrared
light into cell tissues. However, such sensors have general disadvantages
such as relatively short lifetimes, usually requiring replacement
after 1–3 years. Moreover, the electrolytic solution should
be carefully maintained and replenished regularly. Nevertheless, the
PEC sensor was successfully applied to ultrasensitive EP detection,
exhibiting rapid response, high stability, wide linear detection range,
and selectivity. These findings provide insights into the development
of heterostructures for PEC analysis and new methods for EP detection.
Results and Discussion
Characterization of PEC Materials
X-ray photoelectron
spectroscopy (XPS) was used for the elemental analysis of the Au QDs@NiTAPc-Gr
composite material (Figure A). The high-resolution Ni 2p XPS spectrum showed two peaks
at 855.7 and 870.0 eV (Figure B), assigned to Ni 2p3/2 and Ni 2p1/2 of NiTAPc-Gr, respectively.[49] In Figure C, the O 1s peaks
at 531.7 and 533.2 eV were attributed to C–O and C=O
bonds, respectively.[50] The N 1s XPS spectra
(Figure D) of the
Au QDs@NiTAPc-Gr composite material indicated that the peaks of nitrogen
functionalities appeared at 399.2 eV (the N in N–H bonds),
400.5 eV (the N in N–H bonds), and 401.6 eV (the N in C–N
bonds).[51] The C1s peaks at 284.8, 286.0,
and 287.9 eV corresponded to C–C, C–O, and C=O,
respectively (Figure E).[52] Furthermore, the peaks located at
85.2 and 88.8 eV were attributable to Au 4f (Figure H). These findings imply the successful preparation
of the Au QDs@NiTAPc-Gr composite material.
Figure 1
XPS analysis of the Au QDs@NiTAPc-Gr composite: (A) survey scans
and high-resolution spectra of (B) Ni 2p, (C) O 1s, (D) N 1s, (E)
C 1s, and (H) Au 4f. (F) Fourier transform infrared (FTIR) and (G)
UV–vis spectra of (a) GO, (b) NiTAPc, (c) NiTAPc-Gr, and (d)
Au QDs@NiTAPc-Gr.
XPS analysis of the Au QDs@NiTAPc-Gr composite: (A) survey scans
and high-resolution spectra of (B) Ni 2p, (C) O 1s, (D) N 1s, (E)
C 1s, and (H) Au 4f. (F) Fourier transform infrared (FTIR) and (G)
UV–vis spectra of (a) GO, (b) NiTAPc, (c) NiTAPc-Gr, and (d)
Au QDs@NiTAPc-Gr.Furthermore, the
PEC materials GO, NiTAPc, NiTAPc-Gr, and Au QDs@NiTAPc-Gr were characterized
using Fourier transform infrared (FTIR) spectroscopy, as shown in Figure F. GO exhibited absorption
peaks at 1724 and 1618 cm–1 (curve a), which are
related to the symmetric and asymmetric telescopic vibrations of −COO–.[53] NiTAPc showed a bending
vibration at 1609 cm–1 (curve b) related to the
presence of −NH2. In NiTAPc-Gr, NiTAPc, which is
rich in amino groups, was covalently bound to the surface of GO via
reactions with carboxyl groups. As a result, NiTAPc-Gr exhibited a
strong absorption peak corresponding to C=O in amido linkages
at 1694 cm–1 (curve c). Furthermore, the symmetric
telescopic vibration of GO at 1724 cm–1 disappeared,
which could be attributed to the p−π-conjugated effect
of amido linkages, resulting in a shift of the C=O absorption
frequency toward lower wavenumbers.[54] In
the case of Au QDs@NiTAPc-Gr, Au QDs were in situ grown on the surface
of NiTAPc-Gr, and characteristic peaks of amido linkages and l-cysteine were observed at 1687, 1644, and 1613 cm–1.The PEC materials in dimethylformamide (DMF) were characterized
using ultraviolet–visible (UV–vis) spectroscopy (Figure G). The absorption
spectrum of NiTAPc showed two intense Q bands at ∼637 and 715
nm (curve b), whereas the Q bands of NiTAPc-Gr appeared at ∼629
and 683 nm (curve c). After further modification with Au QDs, which
can absorb a range of visible light, the PEC material exhibited an
increased absorbance at 623 nm, indicating the successful modification
of Au QDs on NiTAPc-Gr.[55]The morphologies
of synthesized PEC materials were characterized by scanning electron
microscopy (SEM) and transmission electron microscopy (TEM). As shown
in Figure A, after
reacting with GO, Hovenia dulcis thunb-like
structures were clearly observed on the surfaces of the GO sheets
(Figure B,C). Furthermore,
TEM images revealed that NiTAPc-Gr was modified with Au QDs (Figure D) and the Au QDs
on the NiTAPc-Gr surface possessed dispersed lattice planes with an
average size of approximately 4 nm. These results indicated that the
Au QDs@NiTAPc-Gr composite was successfully prepared.
Figure 2
SEM images of (A) NiTAPc (left), enlarged SEM (top right),
and a
picture of hovenia dulcis thunb (bottom right) and (B) and (C) NiTAPc-Gr.
(C) and (D) TEM images of Au QDs@NiTAPc-Gr.
SEM images of (A) NiTAPc (left), enlarged SEM (top right),
and a
picture of hovenia dulcis thunb (bottom right) and (B) and (C) NiTAPc-Gr.
(C) and (D) TEM images of Au QDs@NiTAPc-Gr.
PEC Characterization of Modified Electrodes
Electrochemical impedance spectroscopy (EIS) measurements were carried
out at a potential of 0.2 V to characterize the fabricated PEC sensor
in a solution of 3 mM [Fe(CN)6]3–/4– containing 0.1 M KCl. As shown in Figure A, the charge-transfer resistance (Ret) of
bare indium tin oxide (ITO) was approximately 49.9 Ω (curve
a). A decrease in Ret was observed after modification with NiTAPc
(∼41.8 Ω, curve b), which was attributed to the excellent
conductivity of this material. An increased Ret was observed for the
GO-modified ITO electrode (∼82.4 Ω, curve c). Additionally,
the Ret of Au QDs@NiTAPc-Gr (∼279.3 Ω, curve e) was much
larger than that of NiTAPc-Gr (∼233.6 Ω, curve d) owing
to additional scattering at the surface of NiTAPc-Gr when the electron
mean free path became comparable to the thickness of the metal film.[56] These results indicated that Au ions were reduced
in the HAuCl4 aqueous solution during the preparation of
Au QDs@NiTAPc-Gr, resulting in the successful modification of the
NiTAPc-Gr surface with Au QDs. In addition, cyclic voltammetry (CV)
tests of various electrodes were conducted in the same aqueous solution
between −0.2 and 0.8 V at a scan rate of 100 mV/s. As shown
in Figure C, the anodic
peak current for bare ITO was approximately 820.8 μA. When the
electrode surface was modified with NiTAPc, the anodic peak current
increased to 908.2 μA because of the good conductivity of this
material. The GO-, NiTAPc-Gr-, and Au QDs@NiTAPc-Gr-modified electrodes
exhibited lower anodic peak currents (782.1, 601.7, and 537.9 μA,
respectively) owing to steric hindrance and electronic repulsion.
Furthermore, Nyquist plots showing the effect of dark conditions and
light illumination on the charge-transfer behavior in Au QDs@NiTAPc-Gr/ITO
are displayed in Figure B. The corresponding EIS measurements were performed in 0.1 M pH
8.0 phosphate-buffered saline (PBS) buffer with or without EP (150
nM) at an applied potential of 0.2 V. The Ret for the PEC sensor under
light illumination was lower than that in dark conditions, indicating
that the separation efficiency of electron–hole (e–/h+) pairs was enhanced under light irradiation.
Figure 3
(A) EIS spectra and (C)
CV curves of (a) bare
ITO electrode, (b) NiTAPc/ITO, (c) GO/ITO, (d) NiTAPc-Gr/ITO, and
(e) Au QDs@NiTAPc-Gr/ITO in 3 mM [Fe(CN)6]3–/4– containing 0.1 M KCl solution. (B) Nyquist plots for Au QDs@NiTAPc-Gr/ITO
in 0.1 M PBS in dark and under light conditions with or without EP
(150 nM) at an applied potential of 0.2 V. (D) Photocurrent responses
of various electrodes for EP detection.
(A) EIS spectra and (C)
CV curves of (a) bare
ITO electrode, (b) NiTAPc/ITO, (c) GO/ITO, (d) NiTAPc-Gr/ITO, and
(e) Au QDs@NiTAPc-Gr/ITO in 3 mM [Fe(CN)6]3–/4– containing 0.1 M KCl solution. (B) Nyquist plots for Au QDs@NiTAPc-Gr/ITO
in 0.1 M PBS in dark and under light conditions with or without EP
(150 nM) at an applied potential of 0.2 V. (D) Photocurrent responses
of various electrodes for EP detection.Subsequently, the photocurrent responses of various PEC sensors were
characterized as shown in Figure D. Photocurrent vs time plots were plotted for the
PEC sensors at a potential of −350 mV in the presence of 150
nM EP in 0.1 M pH 8.0 PBS. The photocurrent responses of NiTAPc/ITO,
GO/ITO, and NiTAPc-Gr/ITO were 0.8, 0.3, and 1.0 μA, respectively,
within 20 s. In the case of Au QDs@NiTAPc-Gr/ITO, the photocurrent
increased to 4.5 μA. The photocurrent of Au QDs@NiTAPc-Gr was
4.5 times higher than that of NiTAPc-Gr, which could be attributed
to the effects of Au QDs toward enhancing the response of NiTAPc-Gr.
These results indicated that the Au QDs@NiTAPc-Gr composite can be
selected as a photocatalytic material for fabricating PEC sensors.
Optimization of Experimental Conditions
To obtain high sensitivity for the determination of EP, two relevant
experimental conditions, namely, the pH of the supporting electrolyte
and the applied potential (V), were examined. In
the presence of 150 nM EP, the photocurrent response increased with
the increase in pH from 4.0 to 8.0, with the maximum value obtained
at pH 8.0 (Figure B). Thus, pH 8.0 PBS buffer was chosen as the optimum condition.
Additionally, the effect of V on the PEC sensor was
examined, as shown in Figure C. The photocurrent response increased linearly with the increase
in V in range from 0 to −350 mV, which can
be described by the linear equation I(μA) =
−0.0087V (mV) + 1.26, R2 = 0.9924 (Figure D). At V higher than −350 mV, the
photocurrent began to decrease. Therefore, an applied potential of
−350 mV was used in subsequent experiments.
Figure 4
(A) Schematic diagram
of the PEC detection on Au QDs@NiTAPc-Gr/ITO.
(B) Effect of PBS solution with different pH values, (C) applied potential
on photocurrent in responses of the PEC sensor, and (D) corresponding
data.
(A) Schematic diagram
of the PEC detection on Au QDs@NiTAPc-Gr/ITO.
(B) Effect of PBS solution with different pH values, (C) applied potential
on photocurrent in responses of the PEC sensor, and (D) corresponding
data.
Detection of EP
Under optimal sensing conditions,
the photocurrent responses of the PEC sensor to different concentrations
of EP (0.12–243.9 nM) were recorded (Figure A), with each test repeated four times. The
photocurrent increased with the increase in EP concentration and a
good linear relationship was observed, which can be expressed as I(μA) = 0.0195C (nM) + 1.72, R2 = 0.9992 (Figure B), with a limit of detection (LOD) of 17.9
pM.
Figure 5
(A) Photoelectrochemical
responses of the PEC sensor in the presence of EP in different concentrations
of 0.12–243.9 nM and (B) corresponding data. (C) Response of
Au QDs@NiTAPc-Gr/ITO during continuous detection of EP under periodic
light irradiation for 650 s. (D) Influence of different interferences
on EP detection.
(A) Photoelectrochemical
responses of the PEC sensor in the presence of EP in different concentrations
of 0.12–243.9 nM and (B) corresponding data. (C) Response of
Au QDs@NiTAPc-Gr/ITO during continuous detection of EP under periodic
light irradiation for 650 s. (D) Influence of different interferences
on EP detection.In the Au QDs@NiTAPc-Gr structure, the NiTAPc-Gr network
and Au QD sensitizer possess different absorption bands owing to their
different energy gaps, allowing adequate utilization of the energy
of the excitation light.[57] Au QDs and NiTAPc-Gr
exhibit cascade band-edge levels that can promote the ultrafast transfer
of charge and effectively inhibit the recombination of negatively
charged electrons (e–) and positively charged holes
(h+) when red excitation light is transmitted through the
photosensitive Au QDs@NiTAPc-Gr material. Therefore, the photocurrent
response is obviously enhanced. Under light irradiation, the photogenerated
electrons in the valence band (VB) are injected in the conduction
band (CB) through a cascade starting from the Au NP to form the electron–hole
pairs. EP as an electron donor can block the recombination of the
photogenerated holes and facilitate the transfer of electrons from
the conduction band of Au QDs to NiTAPc-Gr and then to the surface
of the ITO electrode, resulting in the oxidation of EP to EP+ in the electrolyte and the generation of a strong current response.[27,58,59] The presence of EP enhances the
electron transfer between the photosensitive materials, resulting
in an increase of photocurrent. Thus, different concentrations of
EP affect the magnitude of the photocurrent. A schematic diagram of
the PEC detection on the sensor is shown in Figure A.Additionally, using several analytical
parameters, the performance of the PEC sensor was compared with that
of some of the previously reported strategies for EP detection. As
shown in Table , the
current PEC strategy provides a much lower LOD and a somewhat wider
linear range, which can be attributed to the large specific surface
area and excellent stability and photoelectric conversion capacity
of Au QDs@NiTAPc-Gr under red-light illumination. Thus, the proposed
PEC sensor has great potential for the determination of EP.
Table 1
Comparison
of the Performance of Various Methods for Epinephrine Detection
material/method
linear range (μM)
LOD (μM)
RSD (%)
R
refs
Au-MWCNT-PANI-RuO2/EC
7.69 × 10–2–4.9
0.18
0.9760
(60)
MIPs/MWNTs/EC
0.30–1 × 103
0.03
1.30
0.9980
(61)
Chit-fCNT bio-nanocomposite/EC
0.05–10
0.03
4.50
(62)
GCPE(centri-voltammetry)/EC
0.20–20
1.30 × 10–2
2.53
0.9975
(63)
50–500
4.30 × 10–2
0.9980
GQD-CS/EC
0.36–380
3.00 × 10–4
2.80
0.9983
(64)
caffeic acid/EC
2–80
0.20
0.9987
(65)
β-NiS@rGO/Au
nanocomposites/EC
2–1 × 106
0.54
2.04
0.9924
(66)
UPLC-MS/MS
2 × 10–4–14
1.00 × 10–4
0.9999
(67)
EP–MIP/CL
5 × 10–3–10
3.00 × 10–3
0.9980
(68)
AuNPs@NiTAPc-Gr/PEC
1.20 × 10–4–0.2439
1.79 × 10–5
1.36
0.9996
this work
Stability, Reproducibility, and Selectivity
of the PEC Sensor
To investigate the stability of the proposed
PEC sensor, the photocurrent response during continuous detection
of EP was recorded under periodic light irradiation for 650 s. Good
short-term stability was observed with a relative standard deviation
(RSD) of 1.09%. Subsequently, the PEC sensor was stored in a refrigerator
at 4 °C and monitored occasionally. After 1 month, 90.8% of the
initial photocurrent value was recorded, revealing that the PEC sensor
has long-term stability. Furthermore, the photocurrents of five newly
modified working electrodes tested in the same experiment gave an
RSD of 1.62%, reflecting the good reproducibility of PEC sensor. The
photocurrent response of the PEC sensor was also investigated in the
presence of several possible interferents at a 100-fold higher concentration
than EP (75.0 vs 7.5 × 103 nM). As shown in Figure D, the initial photocurrent
response increased rapidly after EP was added without interferents,
and the subsequent addition of interferents, including Cu2+, K+, Mg2+, Ca2+, Fe2+, glucose, noradrenaline (NA), uric acid (UA), l-Cys, dopamine
(DA), and tyrosine, did not cause a significant change in the photocurrent.
These results demonstrated the potential applicability of the PEC
sensor for complex sample analysis.
Practical Applications
To verify the practical applicability
of the proposed PEC sensor, different amounts of EP were added to
human serum samples (final concentrations of 5, 100, and 200 nM).
These human serum samples were analyzed by the standard recovery method
using the PEC sensor. As shown by the analysis results in Table , the recovery ranged
from 98.80 to 99.44%. The parallel determination was performed five
times (n = 5), and the RSD was less than 4%. These
results indicated that the PEC sensor could be applied for EP detection
in human serum samples.
Table 2
Results of the Detection
of Epinephrine in Human Serum Samples (n = 5)
added (nM)
found (nM)
RSD (%)
recovery (%)
human serum
5
4.94
3.45
98.80
100
99.17
1.44
99.17
200
198.89
1.52
99.44
Conclusions
In this work, to construct a PEC sensor for EP, a novel Au QDs@NiTAPc-Gr
composite was devised by uniformly growing Au QDs on the surface of
NiTAPc-Gr via a chemical reduction method. The integration of Au QDs
and NiTAPc-Gr produced synergetic effects that enhanced the photoelectric
conversion capacity and absorption efficiency, thus increasing the
photocurrent signal. The synthesized composite showed low self-aggregation
of Au QDs, a large specific surface area, and excellent biocompatibility.
This proposed PEC sensor based on Au QDs@NiTAPc-Gr exhibited a wide
linear range (0.12–243.9 nM), a low LOD (17.9 pM), high stability,
good reproducibility, and good selectivity for ultrasensitive EP detection.
Finally, this PEC strategy was successfully applied to the biological
analysis and detection of EP in human serum samples, with recoveries
ranging from 98.80 to 99.44%. In view of these results, the PEC strategy
has great potential for real-time monitoring of real samples.
Experimental Section
Materials and Reagents
Adrenaline hydrochloride, ascorbic
acid (AA), l-cysteine, thionyl chloride, ammonium molybdate,
and HAuCl4 were purchased from Aladdin Industrial Corporation
(Shanghai, China). Graphite was purchased from Xilong Chemical Industry
Co., Ltd. (Shenzhen, China). 4-Nitrophthalimide was obtained from
Meryer Chemical Technology Co., Ltd. (Shanghai, China). Sodium sulfide
nonahydrate (Na2S·9H2O) and carbamide were
purchased from Sinopharm Chemical Reagent Co.,Ltd. (Shanghai, China).
Phosphate-buffered saline (PBS; 0.1 M, pH 7.0) buffer was prepared
using KH2PO4, Na2HPO4·12H2O, and KCl. The human serum sample was obtained from Chongzuo
People’s Hospital (Guangxi, China).
Apparatus
SEM and TEM images were recorded using an
EVO-18 microscope (ZEISS, Oberkochen, Germany) and an FEI Tecnai-G2
F30 microscope (FEI Co., Hillsboro, OR) spectrometer, respectively.
XPS spectra were obtained using a K-α spectrometer (Thermo Fisher
Scientific Co., Waltham, MA). The FTIR spectra were collected with
a Spectrum 65 FTIR spectrophotometer (PerkinElmer Co., Ltd., Waltham,
MA). The UV–vis absorption spectra were obtained using a Shimadzu
UV-6100 UV–vis-NIR spectrophotometer (Shanghai Mapada Instruments
Co., Ltd., Shanghai, China). Red excitation light was provided by
a PEAC 200A system (Ada Hengsheng Technology Development Co., Ltd.,
Tianjin, China). The distance between the illumination source and
the sample cell was maintained at 10 cm. PEC measurements were performed
using an electrochemical workstation (CHI760e, Chenhua Instrument
Co., Ltd., Shanghai, China). ITO slices (≤6 Ω, South
China Xiangcheng Technology Co., Ltd., Shenzhen, China) with an active
surface area of 0.25 cm2 were used as the working electrode
vs Ag/AgCl as the reference electrode.
Synthesis of NiTAPc
NiTAPc was prepared by the reduction
of the corresponding nitro-substituted intermediate.[69] Briefly, 4-nitrophthalimide (1.75 g), carbamide (10 g),
ammonium molybdate (0.025 g), and NiCl2 (1.14 g) were mixed
and then fused by heating to a temperature of 160 °C, providing
the crucial precursor nickeltetranitrophthalocyanine. Then, the synthesized
blue-purple solid and Na2S·9H2O (3.73 g)
were dissolved in DMF (50 mL) and heated at 60 °C for 1.5 h.
Finally, NiTAPc was collected by filtration, washed with water and
ultrapure absolute ethyl alcohol several times, and dried in a vacuum
drying chamber at 70 °C for 2 h.
Preparation of the Au QDs@NiTAPc-Gr Composite
GO was synthesized
from graphite by a modified Hummers’ method.[70] NiTAPc-Gr was synthesized via the following steps. GO (1
mg) was added to a mixed solution of thionyl chloride (10 mL) and
DMF (10 mL), stirred, and then heated at 70 °C for 24 h. NiTAPc
(200 mg) was added to the above mixture after thionyl chloride was
removed by vacuum distillation and heated for a further 96 h. The
product was washed with ultrapure water and absolute ethyl alcohol
several times and then dried in a vacuum drying chamber at 70 °C
for 6 h. Au QDs@NiTAPc-Gr was prepared via the in situ growth of Au
QDs on the surface of NiTAPc-Gr, as described in a previous report
from the Yuan group.[71] NiTAPc-Gr was easily
dispersed in DMF, and then 1 mL of NiTAPc-Gr suspension (0.1 wt %)
was mixed with 5 mL of l-cysteine aqueous solution (1 mM)
using a high-speed shaker for 1 h. Subsequently, 5 mL of HAuCl4 aqueous solution (0.3 mM) and 10 mL of AA aqueous solution
(5 mM) were added to the mixture, which was stirred rapidly at room
temperature for 3 h to give NiTAPc-Gr decorated with Au QDs. Finally,
the product was washed with ultrapure water several times and dried
in a vacuum drying chamber at 70 °C for 6 h. The synthesis mechanism
is shown in Scheme .
Scheme 1
Schematic Illustration of the Synthetic Mechanism
of Au QDs@NiTAPc-Gr
Electrode Fabrication
The ITO slices were cleaned with acetone, ethanol, and ultrapure
water and then dried under an infrared lamp. The ITO substrate was
then coated with 5 μL of a dispersion of Au QDs@NiTAPc-Gr (1
mg) in DMF (1 mL), which was allowed to dry naturally in the air.
In addition, ITO substrates were coated with NiTAPc, GO, and NiTAPc-Gr
using dispersions with the same concentration and volume as mentioned
above. Furthermore, a bare ITO electrode was prepared by coating an
ITO substrate with 5 μL DMF. The coated area of the modified
electrodes was 0.25 cm2.
Authors: Daniela C Marcano; Dmitry V Kosynkin; Jacob M Berlin; Alexander Sinitskii; Zhengzong Sun; Alexander Slesarev; Lawrence B Alemany; Wei Lu; James M Tour Journal: ACS Nano Date: 2010-08-24 Impact factor: 15.881
Authors: Gavin D Perkins; Chen Ji; Charles D Deakin; Tom Quinn; Jerry P Nolan; Charlotte Scomparin; Scott Regan; John Long; Anne Slowther; Helen Pocock; John J M Black; Fionna Moore; Rachael T Fothergill; Nigel Rees; Lyndsey O'Shea; Mark Docherty; Imogen Gunson; Kyee Han; Karl Charlton; Judith Finn; Stavros Petrou; Nigel Stallard; Simon Gates; Ranjit Lall Journal: N Engl J Med Date: 2018-07-18 Impact factor: 91.245