GEC-ESTRO/ACROP recommendations for quality assurance of ultrasound imaging in brachytherapy.

Frank-André Siebert1, Christian Kirisits2, Taran Paulsen Hellebust3, Dimos Baltas4, Frank Verhaegen5, Saskia Camps6, Bradley Pieters7, György Kovács8, Bruce Thomadsen9.   

Abstract

Ultrasound (US) is an important imaging modality in brachytherapy (BT). In particular for low-dose-rate (LDR) and high-dose-rate (HDR) prostate implants transrectal ultrasound (TRUS) is widespread. Besides the common use of US for prostate implants, US can also be applied in gynecological and anal cancer therapies as examples amongst others. The BRAPHYQS (BRAchytherapy PHYsics Quality assurance System) and UroGEC (urology) working groups of GEC-ESTRO (GEC: Groupe Européen de Curiethérapie, committee of ESTRO: European SocieTy for Radiotherapy & Oncology) elaborated upon guidelines describing quality assurance (QA) methods for US in BT. The total quality management (QM) for the unit includes acceptance testing, commissioning and periodic image testing. In 2008, the AAPM (American Association of Physicists in Medicine) published the TG (Task group) 128 report. Whereas the TG 128 focuses on US systems and prostate BT, the current recommendations also cover tests for stepping devices and include other interstitial or intracavitary treatment sites in BT, such as anal implants and gynecological BT. The recommendations presented herein do not replace regular maintenance for the US devices performed by the vendor. They are the QA of US in BT but are not sufficient for the whole maintenance of medical US devices. Moreover, national regulations and recommendations should also be followed. For the tests presented in this report tolerances or action limits are given. These recommendations explain practical test procedures of US devices in BT. They will help the clinics to perform a high level of quality in the use of US for BT in Europe.
Copyright © 2020 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Brachytherapy; Quality assurance; Stepping device; Ultrasound

Mesh:

Year:  2020        PMID: 32335363     DOI: 10.1016/j.radonc.2020.02.024

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  2 in total

1.  High-Dose-Rate Interstitial Brachytherapy for Deeply Situated Gynecologic Tumors Guided by Combination of Transrectal and Transabdominal Ultrasonography: A Technical Note.

Authors:  Yuri Shimizu; Naoya Murakami; Takahito Chiba; Tomoya Kaneda; Hiroyuki Okamoto; Satoshi Nakamura; Ayaka Takahashi; Tairo Kashihara; Kana Takahashi; Koji Inaba; Kae Okuma; Yuko Nakayama; Jun Itami; Hiroshi Igaki
Journal:  Front Oncol       Date:  2022-01-26       Impact factor: 6.244

2.  A new technique for performing interstitial implants for gynecologic malignancies using transvaginal ultrasound guidance.

Authors:  Xiao-Jing Yan; Yi Yang; Xi Chen; Shi-Guang Wang; Shu-Huai Niu; Hui-Xian Niu; Hong Liu
Journal:  Front Oncol       Date:  2022-08-11       Impact factor: 5.738

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.