Maria Infantino1, Valentina Grossi1, Barbara Lari1, Riccardo Bambi1, Alessandro Perri1, Matteo Manneschi1, Giovanni Terenzi1, Irene Liotti1, Giovanni Ciotta1, Cristina Taddei1, Maurizio Benucci2, Patrizia Casprini3, Francesca Veneziani4, Sergio Fabbri1,4, Adolfo Pompetti5, Mariangela Manfredi1.
Abstract
A pandemic of coronavirus disease 2019 (COVID-19) caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has been spreading throughout the world. Though molecular diagnostic tests are the gold standard for COVID-19, serological testing is emerging as a potential surveillance tool, in addition to its complementary role in COVID-19 diagnostics. Indubitably quantitative serological testing provides greater advantages than qualitative tests but today there is still little known about serological diagnostics and what the most appropriate role quantitative tests might play. Sixty-one COVID-19 patients and 64 patients from a control group were tested by iFlash1800 CLIA analyzer for anti-SARS CoV-2 antibodies IgM and IgG. All COVID-19 patients were hospitalized in San Giovanni di Dio Hospital (Florence, Italy) and had a positive oro/nasopharyngeal swab reverse-transcription polymerase chain reaction result. The highest sensitivity with a very good specificity performance was reached at a cutoff value of 10.0 AU/mL for IgM and of 7.1 for IgG antibodies, hence near to the manufacturer's cutoff values of 10 AU/mL for both isotypes. The receiver operating characteristic curves showed area under the curve values of 0.918 and 0.980 for anti-SARS CoV-2 antibodies IgM and IgG, respectively. iFlash1800 CLIA analyzer has shown highly accurate results for the anti-SARS-CoV-2 antibodies profile and can be considered an excellent tool for COVID-19 diagnostics.Entities:
Keywords: SARS coronavirus; coronavirus; humoral immunity; immune responses; virus classification
Year: 2020 PMID: 32330291 PMCID: PMC7264663 DOI: 10.1002/jmv.25932
Source DB: PubMed Journal: J Med Virol ISSN: 0146-6615 Impact factor: 2.327
Figure 1Reciever operating characterstic analysis for anti‐SARS‐CoV‐2 antibodies detection
Figure 2Distribution of anti‐SARS‐CoV‐2 IgM and IgG antibodies levels in COVID‐19 patients and in the control group at the manufacturer's cutoff
Performance characteristics (with 95% confidence intervals) of anti‐SARS‐CoV‐2 antibodies IgM and IgG at different cutoff values as determined by CLIA method
| Anti‐SARS‐CoV‐2 IgM antibodies | |||||||
|---|---|---|---|---|---|---|---|
| Cutoff value | 6.7 AU/mL | 7.5 AU/mL | 9.4 AU/mL | 10.0 AU/mL | 11.3 AU/mL | 12.2 AU/mL | 13.4 AU/mL |
| Sensitivity | 76.7% (59.7‐89.2) | 73.3% (56.0‐86.8) | 73.3% (56.0‐86.8) | 73.3% (56.0‐86.8) | 70.0% (52.4 ‐ 84,3) | 66.7% (48.9‐81.7) | 66.7% (48.9‐81.7) |
| Specificity | 90.6% (81.9‐96.2) | 90.6% (81.9‐96.2) | 92.2% (84.0‐97.1) | 92.2% (84.0‐97.1) | 92.2% (84.0 ‐ 97,1) | 92.2% (84.0‐97.1) | 93.7% (86.1‐98.0) |
| PPV | 79.3% (62.5‐91.2) | 78.6% (61.3‐90.9) | 81.5% (64.3‐92.9) | 81.5% (64.3‐92.9) | 80.8% (63.1‐92.6) | 80.0% (61.8‐92.3) | 83.3% (65.4‐94.5) |
| NPV | 89.2% (80.2‐95.2) | 87.9% (78.6‐94.3) | 88.1% (78.9‐94.4) | 88.1% (78.9‐94.4) | 86.8% (77.4‐93.4) | 85.5% (76.0‐92.5) | 85.7% (76.3‐92.6) |
| LR+ | 8.18 | 7.82 | 9.39 | 9.39 | 8.96 | 8.53 | 10.7 |
| LR− | 0.26 | 0.29 | 0.29 | 0.29 | 0.33 | 0.36 | 0.36 |
| OR | 31.8 | 26.6 | 32.5 | 32.5 | 27.5 | 23.6 | 30.0 |
Abbreviations: CLIA, chemiluminescence immunoassay; LR + , positive likelihood ratio; LR−, negative likelihood ratio; NPV, negative predictive value; OR, odds ratio; PPV, positive predictive value.