| Literature DB >> 32326588 |
Se Yong Jung1, Jae Il Shin2, Jae Young Choi1, Su-Jin Park3, Nam Kyun Kim1,4.
Abstract
The functional change of the right ventricle (RV) after atrial septal defect (ASD) via transcatheter closure is well known. We assessed the immediate RV functional change using velocity vector imaging (VVI) with intracardiac echocardiography (ICE). Seventy-four patients who underwent transcatheter closure of an ASD were enrolled. VVI in the "home view" of ICE showing the RV was obtained before and after the procedure. Velocity, strain, strain rate (SR), and longitudinal displacement were analyzed from VVI data, and the changes of these parameters before and after the procedure were compared. The velocity of the RV decreased after ASD transcatheter closure (3.97 ± 1.48 to 3.56 ± 1.4, p = 0.024), especially in the RV inlet and outlet. The average strain decreased (-19.21 ± 5.79 to -16.87 ± 5.03, p = 0.002), as did the average SR (-2.28 ± 0.64 to -2.03 ± 0.61, p = 0.006). The average longitudinal displacement did not differ. With the VVI technique, we could clearly observe RV functional change immediately after transcatheter closure of the ASD. RV functional change with regional difference may reflect the heterogeneity of volume reduction and suggest subclinical RV dysfunction. These findings can enhance our understanding of the physiologic changes in the RV during reverse remodeling.Entities:
Keywords: atrial septal defect; intracardiac echocardiography; right ventricle function; transcatheter closure; velocity vector imaging
Year: 2020 PMID: 32326588 PMCID: PMC7230484 DOI: 10.3390/jcm9041132
Source DB: PubMed Journal: J Clin Med ISSN: 2077-0383 Impact factor: 4.241
Figure 1“Home view” of intracardiac echocardiography. RA: right atrium, RV: right ventricle. “Home view” was obtained to analyze the vector velocity image of the right ventricle, before and after the transcatheter closure.
Figure 2Vector velocity image of the right ventricle in “home view”. RA: right atrium, RV: right ventricle. The right ventricle was divided into six segments, and the velocity, strain, strain rate, and longitudinal displacement of each part was analyzed from the vector velocity image obtained in “home view”.
Baseline characteristics and hemodynamic data.
| Total Patients | N = 74 |
|---|---|
| Age (years) | 21.7 (range, 3–59) |
| Male/Female | 20/54 |
| Weight (Kg) | 35.9 ± 23.3 |
| Height (cm) | 125.9 ± 34.4 |
| BSA (m2) | 1.1 ± 0.5 |
| Systolic pulmonary artery pressure (mmHg) | 37± 11 |
| LVEF (%) | 67.0 ± 5.2 |
| Qp/Qs | 2.2 ± 0.7 |
| Device Size (mm) | 20 (10–36) |
| Major complication | none |
| Residual shunt | none |
BSA: body surface area, LVEF: left ventricular ejection fraction, Qp/Qs: pulmonary blood flow to systemic blood flow ratios.
Changes of velocity in each RV segment on VVI (cm/s) after ASD device closure.
| Location | Preprocedure | Postprocedure | Mixed-Model * | |
|---|---|---|---|---|
| RV Inlet | 5.05 ± 2.53 | 4.16 ± 2.34 | 0.0077 | 0.01 |
| Midinlet | 4.11 ± 2.15 | 4.03 ± 2.69 | 0.8358 | 0.8 |
| Apex-inlet | 3.18 ± 2.03 | 3.33 ± 2.27 | 0.6264 | 0.618 |
| Apex-outlet | 1.89 ± 1.31 | 1.70 ± 1.53 | 0.2402 | 0.222 |
| Midoutlet | 4.12 ± 2.00 | 3.49 ± 1.63 | 0.1545 | 0.04 |
| RV Outlet | 5.42 ± 2.09 | 4.67 ± 1.69 | 0.0015 | 0.001 |
| Average | 3.97 ± 1.48 | 3.56 ± 1.40 | 0.0248 | 0.022 |
RV: right ventricle, VVI: vector velocity imaging. *: Adjusted by age, sex, body surface area (BSA), RVP, Qp/Qs.
Change of strain in each RV segment on VVI (%) after ASD device closure.
| Location | Preprocedure | Postprocedure | Mixed-Model * | |
|---|---|---|---|---|
| RV Inlet | −15.73 ± 10.11 | −15.33 ± 12.11 | 0.7687 | 0.817 |
| Midinlet | −15.26 ± 9.26 | −12.05 ± 9.40 | 0.015 | 0.02 |
| Apex-inlet | −16.72 ± 10.84 | −14.85 ± 10.21 | 0.2733 | 0.22 |
| Apex-outlet | −24.13 ± 9.56 | −22.61 ± 11.24 | 0.3526 | 0.323 |
| Midoutlet | −25.02 ± 10.41 | −21.89 ± 10.59 | 0.0333 | 0.028 |
| RV Outlet | −18.42 ± 12.04 | −14.51 ± 10.21 | 0.0312 | 0.027 |
| Average | −19.21 ± 5.79 | −16.87 ± 5.03 | 0.0021 | 0.002 |
ASD: atrial septal defect, RV: right ventricle, VVI: vector velocity imaging. *: Adjusted by age, sex, body surface area (BSA), RVP, Qp/Qs.
Change of SR in each RV segment on VVI (1/s) after ASD device closure.
| Location | Preprocedure | Postprocedure | Mixed-Model * | |
|---|---|---|---|---|
| RV Inlet | −2.21 ± 1.54 | −1.90 ± 1.47 | 0.1388 | 0.197 |
| Midinlet | −2.02 ± 1.14 | −1.78 ± 1.06 | 0.0819 | 0.139 |
| Apex-inlet | −2.07 ± 1.03 | −1.93 ± 1.10 | 0.605 | 0.458 |
| Apex-outlet | −2.78 ± 1.14 | −2.46 ± 1.37 | 0.0651 | 0.062 |
| Midoutlet | −2.63 ± 1.31 | −2.47 ± 1.01 | 0.3489 | 0.338 |
| RV Outlet | −1.96 ± 1.15 | −1.79 ± 1.51 | 0.3187 | 0.438 |
| Average | −2.28 ± 0.64 | −2.03 ± 0.61 | 0.006 | 0.008 |
RV: right ventricle, VVI, vector velocity imaging. *: Adjusted by age, sex, body surface area (BSA), RVP, Qp/Qs.
Change of longitudinal displacement in each RV segment on VVI (mm) after ASD device closure.
| Location | Preprocedure | Postprocedure | Mixed-Model * | |
|---|---|---|---|---|
| RV Inlet | 5.37 ± 3.45 | 4.22 ± 3.11 | 0.0045 | 0.005 |
| Midinlet | 4.25 ± 2.81 | 3.75 ± 2.88 | 0.221 | 0.951 |
| Apex-inlet | 3.34 ± 2.59 | 3.25 ± 2.91 | 0.822 | 0.816 |
| Apex-outlet | 2.77 ± 2.43 | 2.33 ± 2.13 | 0.1499 | 0.155 |
| Midoutlet | 5.50 ± 3.19 | 4.91 ± 2.62 | 0.1261 | 0.104 |
| RV Outlet | 7.73 ± 3.44 | 6.87 ± 3.07 | 0.025 | 0.02 |
| Average | 4.83 ± 2.34 | 4.22 ± 1.99 | 0.3161 | 0.321 |
ASD: atrial septal defect, RV: right ventricle, VVI: vector velocity imaging. *: Adjusted by age, sex, body surface area (BSA), RVP, Qp/Qs.