Patient-specific instrument for unicompartmental knee arthroplasty does not reduce the outliers in alignment or improve postoperative function: a meta-analysis and systematic review.

Mingyang Li1, Yi Zeng1, Yuangang Wu1, Yuan Liu1, Wenxing Wei1, Limin Wu1, Bo-Qiang Peng2, Jiayi Li3, Bin Shen4.   

Abstract

BACKGROUND: Unsatisfactory alignment in unicompartmental knee arthroplasty (UKA) is one potential cause of postoperative failure. Patient-specific instruments (PSIs) are designed to improve the alignment of the prostheses, but the effect of PSIs on the alignment or clinical outcome is controversial and lacks validated evidence. We conducted a meta-analysis and systematic review to determine the effect of PSIs on UKA outcomes for the first time.
MATERIALS AND METHODS: A systematic literature search in MEDLINE, EMBASE, CNKI (Chinese database) and Cochrane Central Register of Controlled Trials (up to June 2019) was performed to collect studies that compared PSIs with conventional instruments. Two reviewers independently screened all the records on the basis of inclusion and exclusion criteria. Quality assessments with Cochrane's quality assessment tool or Newcastle-Ottawa scale (NOS) were conducted, the data were extracted, and statistical analyses were completed.
RESULTS: Ten studies with 444 knees were included. The meta-analysis confirmed that PSIs contributed to reduced errors in the alignment of the femoral compartment in the sagittal plane (mean difference = - 2.53, CI [- 3.14, - 1.99], P < 0.01) and the tibial compartment in both the coronal (mean difference = - 0.97, CI [- 1.44, - 0.49], P < 0.01) and the sagittal plane (mean difference = - 1.29, CI [- 1.81, - 0.76], P < 0.01). One study supported that PSIs reduced outliers in inexperienced surgeons; however, all studies investigating PSIs among experienced surgeons suggested that PSIs cannot reduce the percentage of outliers. There was no significant difference in the postoperative score (mean difference = - 0.06, CI [- 0.36, 0.23], P = 0.68) or rate of complications (RR = 1.02, CI [0.15, 6.79], P = 0.99) between PSIs and conventional instruments.
CONCLUSION: The findings of this study suggest PSIs could not reduce the percentage of outliers in UKA patients for experts, and postoperative scores and complication rates are not improved by PSIs, compared with conventional instruments. Based on this meta-analysis and systematic review, no practical benefit to UKAs in experts was detected in PSIs. The findings of this study also suggest that PSIs improved alignment of UKA and might be beneficial to inexperienced surgeons, but it is still unclear whether this improvement is clinically significant and the evidence of inexperienced surgeons is limited. Therefore, more high-quality RCTs are need to be carried out in the future.

Entities:  

Keywords:  Alignment; PSI; Patient-specific instrument; Total knee arthroplasty; Unicompartmental knee arthroplasty

Year:  2020        PMID: 32306092     DOI: 10.1007/s00402-020-03429-z

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  3 in total

Review 1.  [Focal femoral resurfacing and unicompartmental knee replacement : Between osteotomy and total knee replacement].

Authors:  Philipp Henle; Matthias J Feucht; Christian Stärke
Journal:  Orthopade       Date:  2021-04-13       Impact factor: 1.087

2.  Patient-specific instrumentation (PSI) in total ankle arthroplasty: a systematic review.

Authors:  Qiuyuan Wang; Nianfei Zhang; Wanshou Guo; Weiguo Wang; Qidong Zhang
Journal:  Int Orthop       Date:  2021-08-04       Impact factor: 3.075

3.  Comparison of joint awareness after medial unicompartmental knee arthroplasty and high tibial osteotomy: a retrospective multicenter study.

Authors:  Shotaro Watanabe; Ryuichiro Akagi; Taishi Ninomiya; Takeshi Yamashita; Masamichi Tahara; Seiji Kimura; Yoshimasa Ono; Satoshi Yamaguchi; Seiji Ohtori; Takahisa Sasho
Journal:  Arch Orthop Trauma Surg       Date:  2021-07-16       Impact factor: 3.067

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.