Edoardo Monaco1, Giorgio Bruni1, Sara Lo Torto1, Alessandro Carrozzo2, Matthew Daggett3, Alessandro Annibaldi1, Adnan Saithna3, Andrea Ferretti1. 1. Orthopaedic Unit and Kirk Kilgour Sports Injury Centre, S. Andrea Hospital, University of Rome La Sapienza, Rome, Italy. 2. Orthopaedic Unit and Kirk Kilgour Sports Injury Centre, S. Andrea Hospital, University of Rome La Sapienza, Rome, Italy. alessandrocarrozzo27@gmail.com. 3. Kansas City University of Medicine and Biosciences, Kansas City, MO, USA.
Abstract
PURPOSE: The aim of this study was to determine the inter- and intra-observer reliability of knee laxity assessment using a non-invasive navigation system in a population of healthy young athletes. It was hypothesized that knee laxity parameters recorded using non-invasive computer navigation would demonstrate good inter- and intra-observer reliability. METHODS: Healthy volunteers aged between 18 to 30 years were recruited to the study. Static and dynamic knee laxity parameters including anterior tibial translation and tibial rotation during the pivot shift test were recorded on awake patients using non-invasive computer navigation by two independent observers: at the first visit each athlete was evaluated by the consultant and resident surgeons independently; 6 weeks after the first visit all the participants were re-tested only by the resident surgeon. Inter- and intra-observer reliability was calculated and then interpreted according to Cicchetti's criteria. RESULTS: One hundred healthy volunteers were recruited to the study, of these 38 were women (38%), and the average age was 25.5 ± 2.4 years. According to Cicchetti's criteria the intra- and inter-observer reliability for static measurements were fair for anterior tibial translation (0.572 and 0.529, respectively) and excellent for total passive tibial rotation (0.859 and 0.883, respectively). For the dynamic measurements of translation and rotation during the pivot shift maneuver both measurements demonstrated good to excellent reliability with intra and inter observer reliability ranging from 0.684 to 0.936. CONCLUSION: Non-invasive navigation for the assessment of knee laxity is associated with fair to excellent inter- and intra-observer reliability in a population of healthy volunteers.
PURPOSE: The aim of this study was to determine the inter- and intra-observer reliability of knee laxity assessment using a non-invasive navigation system in a population of healthy young athletes. It was hypothesized that knee laxity parameters recorded using non-invasive computer navigation would demonstrate good inter- and intra-observer reliability. METHODS: Healthy volunteers aged between 18 to 30 years were recruited to the study. Static and dynamic knee laxity parameters including anterior tibial translation and tibial rotation during the pivot shift test were recorded on awake patients using non-invasive computer navigation by two independent observers: at the first visit each athlete was evaluated by the consultant and resident surgeons independently; 6 weeks after the first visit all the participants were re-tested only by the resident surgeon. Inter- and intra-observer reliability was calculated and then interpreted according to Cicchetti's criteria. RESULTS: One hundred healthy volunteers were recruited to the study, of these 38 were women (38%), and the average age was 25.5 ± 2.4 years. According to Cicchetti's criteria the intra- and inter-observer reliability for static measurements were fair for anterior tibial translation (0.572 and 0.529, respectively) and excellent for total passive tibial rotation (0.859 and 0.883, respectively). For the dynamic measurements of translation and rotation during the pivot shift maneuver both measurements demonstrated good to excellent reliability with intra and inter observer reliability ranging from 0.684 to 0.936. CONCLUSION: Non-invasive navigation for the assessment of knee laxity is associated with fair to excellent inter- and intra-observer reliability in a population of healthy volunteers.