E E Nazarova1, G V Tereshchenko2, D A Kupriyanov2,3, N S Smetanina2,4, G A Novichkova2.
Abstract
BACKGROUND: Timely diagnosis of cardiac iron overload is important for children with transfusion-dependent anaemias and requires modern measure methods. Nowadays, myocardial iron quantification is performed by magnetic resonance (MR) breath-hold techniques, sensitive to respiratory motion and unfeasible in patients who are unable to hold their breath. Free-breathing T2* mapping sequences would allow to scan children who cannot hold their breath for a specified duration. Our aim was to test a free-breathing T2* mapping sequence, based on motion correction by multiple signal accumulation technique.Entities:
Keywords: Artefacts; Heart; Iron overload; Magnetic resonance imaging; Myocardium
Mesh:
Year: 2020 PMID: 32303909 PMCID: PMC7165216 DOI: 10.1186/s41747-020-00156-3
Source DB: PubMed Journal: Eur Radiol Exp ISSN: 2509-9280
Imaging parameters for breath-holding (BH-mTFE) and free-breathing (FB-mTFE) sequences
| BH-mTFE | FB-mTFE | |
|---|---|---|
| Number of slices | 1 | 1 |
| Acquisition matrix | 124 × 119 | 124 × 122 |
| Resolution (mm) | 2.4 × 2.5 × 8 | 2.4 × 2.4 × 8 |
| Field of view (mm) | 297 × 297 | 297 × 292 |
| Slice thickness (mm) | 8 | 8 |
| k-space sampling | Cartesian | Cartesian |
| Flip angle (degree) | 25 | 25 |
| Echo time (ms) | 1.3 | 1.3 |
| ∆ Echo time (ms) | 1.2 | 1.2 |
| Repetition time (ms) | 21 | 22 |
| Number of excitations | 1 | 8 |
| Electrocardiography-gating | Cardiac triggered | Cardiac triggered |
| Respiration | Breath hold ≈ 15 sec | Respiratory trigger |
| Duration | 12 s | 2 min, 11 s |
Fig. 1T2* map with a region of interest (ROI) placed in the interventricular septum
Fig. 2Boxplots of the signal-to-noise ratio (SNR) for breath-holding (BH-mTFE) and free-breathing (FB-mTFE) sequences. SNR of BH-mTFE = 112.6, interquartile range 71.1−334.1. SNR of FB-MTFE = 212.8; interquartile range 148.5−566.5. • Outliers
Fig. 3Example of data excluded from analysis regarding a 5-year old boy with Diamond–Blackfan anaemia who underwent the cardiac magnetic resonance without anaesthetic support. a Original image obtained using the breath-holding technique. b Original image obtained using the free-breathing technique. c T2* map obtained using the breath-holding technique. d T2* map obtained using the free-breathing technique
Fig. 4Bland–Altman plots showing the good agreement for T2* values in the interventricular septum between the first reader (a) and the second reader (b)
Fig. 5Multi-phase turbo field-echo images and T2* maps of a 9-year old boy affected with a β-thalassemia. a Original image obtained using the breath-holding technique. b Original image obtained using the free-breathing technique. c T2* map obtained using the breath-holding technique. d T2* map obtained using the free-breathing technique
Fig. 6Scatterplot of T2* values obtained using free-breathing (FB) and breath-holding (BH) sequences (n = 32; r = 0.99; R2 = 0.98; p < 0.001). Continuous thin lines show the 95% confidence interval