| Literature DB >> 32292283 |
Anamarija Rade1, Anamarija Đuras1, Irena Kocijan1, Patricija Banković Radovanović2, Ana Turčić3.
Abstract
INTRODUCTION: Serum samples of haemodialysed patients collected through vascular access devices, e.g. central venous catheter (CVC) can contain residual heparin, which can cause incomplete clotting and consequently fibrinogen interference in serum protein electrophoresis (SPE). We hypothesized that this problem may be overcome by addition of thrombin and aimed to find a simple thrombin-based method for fibrinogen interference removal.Entities:
Keywords: central venous catheter; electrophoresis; fibrinogen; interference; thrombin
Mesh:
Substances:
Year: 2020 PMID: 32292283 PMCID: PMC7138005 DOI: 10.11613/BM.2020.020705
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Figure 1Comparison of serum protein electrophoresis of haemodialysed patient’s sample drawn in two different tubes regarding the monoclonal band in the beta fraction (marked with arrows) due to the fibrinogen interference. CAT - clot activator tube. RST - rapid serum tube.
Presence of monoclonal band in the beta fraction (fibrinogen interference) of SPE in samples of haemodialysed patients collected through CVC in CAT and RST tubes
| Band absent | 6 | 0 | 0.031 |
| Band present | 6 | 39 | |
| CAT - clot activator tube. CVC - central venous catheter. RST - rapid serum tube. SPE - serum protein electrophoresis. P < 0.05 was considered statistically significant. | |||
Figure 2Comparison of serum protein electrophoresis of haemodialysed patient’s sample drawn through a central venous catheter into a CAT tube (reference) and treated with two thrombin-based methods regarding the efficiency in removing the fibrinogen interference (marked with arrows). 1 - Reference serum collected in clot activator tube (CAT). 2 - CAT serum transferred in rapid serum tube vacutainer. 3 - CAT serum treated with Multifibren U reagent.
Efficiency of two thrombin-based methods for fibrinogen interference removal
| CAT tube (native serum) (1) | 6 | 45 | < 0.001 | (2)/(1)(3) |
| 500 µL CAT serum in RST tube (2) | 45 | 6 | ||
| 100 µL CAT Serum treated with 10 µL of Multifibren U (3) | 6 | 45 | ||
| CAT - clot activator tube. RST - rapid serum tube. SPE - serum protein electrophoresis. P < 0.05 was considered statistically significant. | ||||
Comparison of protein concentration of all SPE fractions of reference sample (CAT serum) and two thrombin-based methods for removing fibrinogen interference
| Albumin, g/L | 39.2 | 42.1 | 40.1 | < 0.001 | (2)/(1)(3) |
| alfa 1, g/L | 1.8 | 1.9 | 1.9 | < 0.001 | (2)/(1)(3) |
| alfa 2, g/L | 6.6 | 7.1 | 6.6 | < 0.001 | (2)/(1)(3) |
| beta, g/L | 11,2 | 6.9 | 11.3 | < 0.001 | (2)/(1)(3) |
| gamma, g/L | 7.5 | 9.4 | 7.2 | < 0.001 | (2)/(1)(3) |
| CAT - clot activator tube. RST - rapid serum tube. SPE- serum protein electrophoresis. SPE fractions were presented as median (interquartile range). P < 0.05 was considered statistically significant. | |||||