RETeval Portable Electroretinogram Parameters in Different Severity Stages of Glaucoma.

Yoshiyuki Kita1, Gábor Holló2, Tsunehiro Saito1, Yousuke Momota1, Ritsuko Kita1, Kazushige Tsunoda3, Akito Hirakata1.   

Abstract

PRéCIS:: Four parameters of the noninvasive, portable RETeval electroretinogram (ERG) system were found to correlate with visual field mean deviation and optical coherence tomography (OCT) thickness parameters, and may therefore be suitable for glaucoma detection.
PURPOSE: To investigate the RETeval full-field ERG parameters for accuracy of separating glaucoma and normal eyes, and correlation with glaucoma severity. PATIENTS AND METHODS: Sixty-two eyes of 62 primary open-angle glaucoma patients [visual field mean deviation (MD) range: -0.44 to -31.15 dB] and 39 eyes of 39 healthy controls underwent one RETeval test (photopic negative response protocol), OCT imaging, and Humphrey 30-2 visual field testing. The glaucoma patients were divided into early (MD≥-6dB, n=33) and moderate-to-advanced (MD<-6 dB, n=29) groups.
RESULTS: Significant correlations were found between the best-performing 4 RETeval ERG parameters and the glaucoma severity measures (MD and OCT thickness parameters) for all eyes, all glaucoma eyes and the moderate-to-advanced glaucoma eyes [photopic negative response amplitude at 72 ms (PhNR 72) and MD: r=-0.333, -0.414, and -0.485, respectively, P≤0.008; PhNR 72 and average circumpapillary retinal nerve fiber layer thickness; r=-0.429, -0.450, and -0.542, respectively, P≤0.002]. Except for P-ratio, there was no significant difference between the area under the receiver-operating characteristic (AUROC) values of the OCT thickness parameters (range: 0.927 to 0.938) and the 4 best-performing RETeval ERG parameters (range: 0.839 to 0.905) in the early glaucoma versus control separation. For differentiating the control and the moderate-to-advanced glaucoma eyes, the AUROC values of the 4 best-performing RETeval ERG parameters ranged between 0.924 and 0.958, and no significant difference was found between them and those of the OCT parameters.
CONCLUSIONS: The noninvasive, portable RETeval full-field ERG device may be useful to detect glaucoma in moderate-to-advanced stages.

Entities:  

Mesh:

Year:  2020        PMID: 32287150     DOI: 10.1097/IJG.0000000000001509

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  3 in total

1.  Asymmetric Functional Impairment of ON and OFF Retinal Pathways in Glaucoma.

Authors:  Alan W Kong; Marcus L Turner; Hoover Chan; Robert L Stamper; Benjamin F Arnold; Luca Della Santina; Yvonne Ou
Journal:  Ophthalmol Sci       Date:  2021-05-10

2.  Prediction of glaucoma severity using parameters from the electroretinogram.

Authors:  Marc Sarossy; Jonathan Crowston; Dinesh Kumar; Anne Weymouth; Zhichao Wu
Journal:  Sci Rep       Date:  2021-12-13       Impact factor: 4.379

3.  Comparison of Machine Learning Approaches to Improve Diagnosis of Optic Neuropathy Using Photopic Negative Response Measured Using a Handheld Device.

Authors:  Tina Diao; Fareshta Kushzad; Megh D Patel; Megha P Bindiganavale; Munam Wasi; Mykel J Kochenderfer; Heather E Moss
Journal:  Front Med (Lausanne)       Date:  2021-12-03