Fiberoptic bronchoscopy precaution strategies.
| Procedures | Notes |
|---|---|
| 1. Enter unit, close door and verify patient's information. | |
| 2. Adjust ventilator oxygen concentration to 100%. | Monitor patient's vital signs |
| 3. Physician prepares bronchoscope; nurse prepares patient and needed supplies; nursing assistant connects power and suction device. | Patient should be deeply sedated (RASS ≤ −4) to avoid chocking; make sure balloon pressure is appropriate. |
| 4. Assistant suspends ventilation prior to bronchoscope insertion into the airway. | Minimize aerosol dispersion |
| 5. Nurse assists physician to collect secretion specimens using closed collection method (see | Minimize scope withdrawal from the airway to reduce aerosol exposure |
| 6. Similarly, assistant suspends ventilation prior to bronchoscope withdrawal from the airway. | Minimize aerosol dispersion |
Fig. 1Method to connect bronchoscope and specimen container.
Disinfection procedure, revisions and rationales.
| Procedures | Traditional methods | Revised methods | Rationales |
|---|---|---|---|
| Preprocess | Preprocess at bedside | Packed in disposable biocontainment bag and preprocess occurs in designated room | Reduce the spread of pathologic organisms |
| Manual cleaning | Bronchoscope immersed in enzyme solution for 5–10 min | Bronchoscope immersed in enzyme solution for 15–20 min | Promote enhanced breakdown of microorganism protein |
| Brush inserted to clean bronchoscope | Large syringe filled with enzyme solution is used to repeatedly flush the bronchoscope and tubing | Avoid spills during cleaning | |
| Drying | Inflated with air for 30 s to dry | ||
| Machine cleaning | Machine specially used for fiberoptic bronchoscope cleaning and disinfection | ||
| Flushing | Sterile purified water used to flush bronchoscope | ||
| Drying | 75% alcohol used to flush bronchoscope followed by inflation with air for 30 s | Bronchoscope flushed with 75% alcohol then immersed in alcohol for 10 minutes followed by inflation with air for 30 s | To prolong disinfection time |
| Storage | Placed in cabinet | Exposure to ultraviolet lamp for one hour prior to storage in cabinet | Ultraviolet rays inactivate the virus |