Accurate neonatal heart rate monitoring using a new wireless, cap mounted device.

Caroline Henry1, Lara Shipley1, Carole Ward1, Siavash Mirahmadi2, Chong Liu2, Steve Morgan2, John Crowe2, James Carpenter3, Barrie Hayes-Gill2, Don Sharkey1.   

Abstract

AIM: A device for newborn heart rate (HR) monitoring at birth that is compatible with delayed cord clamping and minimises hypothermia risk could have advantages over current approaches. We evaluated a wireless, cap mounted device (fhPPG) for monitoring neonatal HR.
METHODS: A total of 52 infants on the neonatal intensive care unit (NICU) and immediately following birth by elective caesarean section (ECS) were recruited. HR was monitored by electrocardiogram (ECG), pulse oximetry (PO) and the fhPPG device. Success rate, accuracy and time to output HR were compared with ECG as the gold standard. Standardised simulated data assessed the fhPPG algorithm accuracy.
RESULTS: Compared to ECG HR, the median bias (and 95% limits of agreement) for the NICU was fhPPG -0.6 (-5.6, 4.9) vs PO -0.3 (-6.3, 6.2) bpm, and ECS phase fhPPG -0.5 (-8.7, 7.7) vs PO -0.1 (-7.6, 7.1) bpm. In both settings, fhPPG and PO correlated with paired ECG HRs (both R2  = 0.89). The fhPPG HR algorithm during simulations demonstrated a near-linear correlation (n = 1266, R2  = 0.99).
CONCLUSION: Monitoring infants in the NICU and following ECS using a wireless, cap mounted device provides accurate HR measurements. This alternative approach could confer advantages compared with current methods of HR assessment and warrants further evaluation at birth.
© 2021 The Authors. Acta Paediatrica published by John Wiley & Sons Ltd on behalf of Foundation Acta Paediatrica.

Entities:  

Keywords:  heart rate; infant; newborn; photoplethysmography; resuscitation; technologies; wireless

Mesh:

Year:  2020        PMID: 32281685     DOI: 10.1111/apa.15303

Source DB:  PubMed          Journal:  Acta Paediatr        ISSN: 0803-5253            Impact factor:   2.299


  4 in total

Review 1.  Toward greater nuance in delayed cord clamping.

Authors:  Laura Marrs; Susan Niermeyer
Journal:  Curr Opin Pediatr       Date:  2022-04-01       Impact factor: 2.856

2.  Feasibility of a Novel ECG Electrode Placement Method in Newborn Infants.

Authors:  Caroline Henry; Lara Shipley; Stephen Morgan; John A Crowe; James Carpenter; Barrie Hayes-Gill; Don Sharkey
Journal:  Neonatology       Date:  2022-02-07       Impact factor: 5.106

3.  Multicentre paired non-inferiority study of the cardiorespiratory monitoring performance of the wireless and non-adhesive Bambi® belt measuring diaphragm activity in neonates: study protocol.

Authors:  Anouk W J Scholten; Zhuozhao Zhan; Hendrik J Niemarkt; Marieke Vervoorn; Ruud W van Leuteren; Frans H de Jongh; Anton H van Kaam; Edwin R van den Heuvel; G Jeroen Hutten
Journal:  BMJ Paediatr Open       Date:  2022-06

4.  A Textile Sleeve for Monitoring Oxygen Saturation Using Multichannel Optical Fibre Photoplethysmography.

Authors:  Hattan K Ballaji; Ricardo Correia; Serhiy Korposh; Barrie R Hayes-Gill; Francisco U Hernandez; Byron Salisbury; Stephen P Morgan
Journal:  Sensors (Basel)       Date:  2020-11-17       Impact factor: 3.576

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.