Dose reduction using digital fluoroscopy versus digital subtraction angiography in endovascular aneurysm repair: A prospective randomized trial.

Kenneth S Zurcher1, Sailendra G Naidu2, Samuel R Money3, William M Stone3, Richard J Fowl4, Grace Knuttinen1, Rahmi Oklu1, Lisa A Rotellini Coltvet1, Daniel Crawford5, Matthew R Buras6, William Pavlicek7.   

Abstract

OBJECTIVE: Endovascular aneurysm repair (EVAR) can result in high radiation dose to patients and operators. This prospective randomized study aimed to assess whether patient radiation dose sustained during EVAR could be decreased by predominantly using digital fluoroscopy (DF) vs the standard technique using digital subtraction angiography (DSA).
METHODS: Between February 2011 and June 2017, patients with EVAR of infrarenal abdominal aortic aneurysms were prospectively enrolled and randomly assigned to a standard treatment DSA cohort or a DF cohort in which two or fewer DSA acquisitions were allowed for confirmatory imaging. Primary end points included dose-area product (DAP) and cumulative air kerma. Secondary end points included technical success and conversion to DSA standard treatment (if DF was inadequate for visualization).
RESULTS: For all 43 patients enrolled (26 in the DF cohort, 17 in the DSA cohort), technical success was 100%. Of the 26 DF patients, 5 (19%) required conversion to the DSA cohort. In an intention-to-treat analysis, mean DAP was significantly lower in the DF cohort than in the DSA cohort (132 vs 174 Gy·cm2; P = .04). When patients were separated by number of DSA acquisitions (two or fewer vs three or more), mean DAP decreased 41% (109 vs 185 Gy·cm2; P = .005) and cumulative air kerma decreased 40% (578 vs 964 mGy; P = .004).
CONCLUSIONS: In most patients (81%), DF or limited DSA was adequate for visualization during EVAR. In both intention-to-treat DF and limited-DSA cohorts, mean DAP was significantly decreased. If image quality allows, a DF-only or limited-DSA approach to EVAR decreases radiation dose.
Copyright © 2020 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Digital subtraction angiography; Dose reduction; Endovascular aneurysm repair; Fluoroscopy; Radiation dose; Stent graft

Mesh:

Year:  2020        PMID: 32276019     DOI: 10.1016/j.jvs.2020.02.050

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  2 in total

1.  [Comparison of radiation exposure in common hepatic interventions : A retrospective analysis of DeGIR registry data].

Authors:  Jonathan Nadjiri; Tobias Geith; Tobias Waggershauser; Lothar Heuser; Dominik Morhard; Arno Bücker; Philipp M Paprottka
Journal:  Radiologe       Date:  2021-01       Impact factor: 0.635

2.  Occupational and Patient Radiation Dose and Quality Implications of Femoral Access Imaging During Coronary Angiography.

Authors:  Kelly Wilson-Stewart; Davide Fontanarosa; Eva Malacova; Steven Gett; Allan Kruger; Jamie V Trapp
Journal:  J Multidiscip Healthc       Date:  2021-07-12
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.