Edwin Seet1, Jinbin Zhang2, Joselo Macachor1, Chandra M Kumar3.
Abstract
General anaesthesia is sometimes favoured over regional anaesthesia in ophthalmic surgery. The use of supraglottic airway (SGA) or laryngeal mask airway (LMA) as the primary airway device is increasing due to numerous advantages over tracheal intubation. Compared with 1st generation SGAs, 2nd generation SGAs have an added benefit of isolating the airway from the alimentary tract. However, the vertical profile of SGAs may encroach into the surgical field and hence interfere with surgery. We investigated the vertical projections of 1st generation SGAs (LMA Classic, Ambu AuraFlex) and commonly used 2nd generation SGAs in our institution (LMA ProSeal, LMA Supreme, LMA Protector, Ambu AuraGain and I-gel) in a manikin model. Each device was connected to a corrugated catheter mount or angled connector following insertion as per usual clinical practice in our institutions. Vertical projections of all devices were measured from the chin using a centimetre ruler. Securing of airway device to the chin with an adhesive tape was possible for the LMA Classic and Ambu AuraFlex with straight corrugated connector, whereas the stiffer 2nd generations SGAs required the addition of an angled connector or straight corrugated tubing to direct the airway tube caudally, away from the surgical field. The LMA ProSeal had the lowest vertical projection amongst the 2nd generation SGAs and may be the suitable choice for ophthalmic surgery. We also describe a novel technique of utilising a 1st generation SGA with placement of an orogastric tube, although with some reservations. This study has several limitations and transferability of our findings into clinical practice is questionable as the use of a manikin may not fully imitate the real condition of the patient. Our study is the first study comparing vertical projected height of different SGAs in manikin, but future studies should investigate the use of SGA in the clinical setting during ophthalmic surgery.Entities:
Keywords: General anaesthesia; Laryngeal mask airway; Ophthalmic anaesthesia; SAD; SGA; Supraglottic airway
Year: 2020 PMID: 32274646 PMCID: PMC7223643 DOI: 10.1007/s10877-020-00507-w
Source DB: PubMed Journal: J Clin Monit Comput ISSN: 1387-1307 Impact factor: 2.502
Fig. 1a Ambu AuraFlex with corrugated connector; b LMA Classic with straight corrugated connector; c1 LMA ProSeal with angle connector; c2 LMA ProSeal with corrugated connector; d1 LMA Supreme with angle connector; d2 LMA Supreme with corrugated connector; e1 Ambu AuraGain with angle connector; e2 Ambu AuraGain with corrugated connector; f1 LMA Protector with angle connector; f2 LMA Protector with corrugated connector; g1 I-gel with angled connector; g2 I-gel with straight corrugated connector
Type of SGA, manufacturers, material of SGA, difficulties encountered in securing to the chin and vertical projected heights
| Supraglottic airway | Manufacturer | Figure | Bite block | Material of device | Was it possible to secure the device to the chin? | Vertical projection (cm) |
|---|---|---|---|---|---|---|
| Ambu AuraFlex with corrugated connector | Ambu A/S, Ballerup, United Kingdom | A | No | Polyvinyl chloride | Yes | 3.4 |
| LMA Classic with straight corrugated connector | Teleflex Medical, Westmeath, Ireland | B | No | Silicone | Yes | 3.8 |
| LMA Proseal with angle connector | Teleflex Medical, Westmeath, Ireland | C1 | Yes | Silicone | Possible with difficulty | 4.5 |
| LMA ProSeal with corrugated connector | Teleflex Medical, Westmeath, Ireland | C2 | Yes | Silicone | Possible with difficulty | 5.9 |
| LMA Supreme with angle connector | Teleflex Medical, Westmeath, Ireland | D1 | Yes | Polyvinyl chloride | No | 9.5 |
| LMA Supreme with corrugated connector | Teleflex Medical, Westmeath, Ireland | D2 | Yes | Polyvinyl chloride | No | 9.8 |
| Ambu AuraGain with angle connector | Ambu A/S, Glen Burnie, Maryland, USA | E1 | Yes | Polyvinyl chloride | No | 9.9 |
| Ambu AuraGain with corrugated connector | Ambu A/S, Glen Burnie, Maryland, USA | E2 | Yes | Polyvinyl chloride | No | 11.6 |
| LMA Protector with angle connector | Teleflex Medical, Westmeath, Ireland | F1 | Yes | Silicone | No | 11.0 |
| LMA Protector with corrugated connector | Teleflex Medical, Westmeath, Ireland | F2 | Yes | Silicone | No | 12.0 |
| I-gel with angled connector | Intersurgical Ltd, Berkshire, United Kingdom | G1 | Yes | Styrene ethylene butadiene styrene | No | 11.8 |
| I-gel with straight corrugated connector | Intersurgical Ltd, Berkshire, United Kingdom | G2 | Yes | Styrene ethylene butadiene styrene | No | 13.1 |
Fig. 2Legend: LMA Classic with an orogastric tube