Literature DB >> 32269934

Effect of acquisition depth and precompression from probe and couplant on shear wave elastography in soft tissue: an in vitro and in vivo study.

Xiuming Wang1, Yue Hu1, Jia'an Zhu1, Junxue Gao1, Si Chen1, Fang Liu1, Wenxue Li1, Yiqun Liu1, Bilig Ariun1.   

Abstract

BACKGROUND: Shear wave elastography is a promising method to diagnose early musculoskeletal lesions. We aimed to explore the feasible depth and acceptable precompression applied by probe and couplant for soft tissues in the present system.
METHODS: Ex-vivo muscles were evaluated at depths of 0.5-6 cm by 3 operators, using 1-5 mm couplant thickness and 0-3.0 kPa probe pressure. We compared the shear wave speed (SWS) and used intraclass correlation coefficients to assess reproducibility. In vivo skin and subcutaneous superficial fascia from volunteers were tested at depths ranging from 0.1-0.5 cm with 1-20 mm couplant thickness.
RESULTS: The SWS of ex-vivo muscles varied and increased with depth, and could not be acquired at 6 cm because the shear wave failed to be detected. Furthermore, while the SWS of ex-vivo muscles were not affected by the couplant thickness, it was affected by probe pressure. Most cases demonstrated a satisfactory agreement degree of the intraoperator reproducibility (ICC, 0.81-0.95) and a substantial interoperator reproducibility (ICC >0.60). Inter- and intra-operator reproducibility was better at a depth of 0.5-4 cm than at 5 cm. In the in vivo study, when tissues within a 0.2 cm depth were evaluated, the SWS that was acquired using a couplant thickness of >10 mm was different from that acquired using other thicknesses.
CONCLUSIONS: The SWS acquired at a depth of ≤3 cm with a suitable amount of couplant is recommended. 2020 Quantitative Imaging in Medicine and Surgery. All rights reserved.

Entities:  

Keywords:  Acquisition depth; pressure; shear wave elastography; soft tissue

Year:  2020        PMID: 32269934      PMCID: PMC7136734          DOI: 10.21037/qims.2020.01.15

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  29 in total

1.  Influence of measurement depth on the stiffness assessment of healthy liver with real-time shear wave elastography.

Authors:  Cong-Zhi Wang; Jian Zheng; Ze-Ping Huang; Yang Xiao; Dan Song; Jie Zeng; Hai-Rong Zheng; Rong-Qin Zheng
Journal:  Ultrasound Med Biol       Date:  2013-12-19       Impact factor: 2.998

2.  Ultrasonographic tissue quantification of the breast using acoustic radiation force impulse technology: phantom study and clinical application.

Authors:  Mitsuhiro Tozaki; Masahiro Saito; Chanwoong Joo; Miki Yamaguchi; Sachiko Isobe; Yukari Ogawa; Kanako Homma; Eisuke Fukuma
Journal:  Jpn J Radiol       Date:  2011-09-17       Impact factor: 2.374

3.  Ultrasound shear wave velocity in skeletal muscle: A reproducibility study.

Authors:  C Dorado Cortez; L Hermitte; A Ramain; C Mesmann; T Lefort; J B Pialat
Journal:  Diagn Interv Imaging       Date:  2015-06-25       Impact factor: 4.026

Review 4.  WFUMB guidelines and recommendations for clinical use of ultrasound elastography: Part 2: breast.

Authors:  Richard G Barr; Kazutaka Nakashima; Dominique Amy; David Cosgrove; Andre Farrokh; Fritz Schafer; Jeffrey C Bamber; Laurent Castera; Byung Ihn Choi; Yi-Hong Chou; Christoph F Dietrich; Hong Ding; Giovanna Ferraioli; Carlo Filice; Mireen Friedrich-Rust; Timothy J Hall; Kathryn R Nightingale; Mark L Palmeri; Tsuyoshi Shiina; Shinichi Suzuki; Ioan Sporea; Stephanie Wilson; Masatoshi Kudo
Journal:  Ultrasound Med Biol       Date:  2015-03-18       Impact factor: 2.998

Review 5.  Shear-Wave Elastography: Basic Physics and Musculoskeletal Applications.

Authors:  Mihra S Taljanovic; Lana H Gimber; Giles W Becker; L Daniel Latt; Andrea S Klauser; David M Melville; Liang Gao; Russell S Witte
Journal:  Radiographics       Date:  2017 May-Jun       Impact factor: 5.333

6.  The influence of precompression on elasticity of thyroid nodules estimated by ultrasound shear wave elastography.

Authors:  A C L Lam; S W A Pang; A T Ahuja; K S S Bhatia
Journal:  Eur Radiol       Date:  2015-11-25       Impact factor: 5.315

7.  Shear wave elastography evaluation of the median and tibial nerve in diabetic peripheral neuropathy.

Authors:  Ying He; Xi Xiang; Bi-Hui Zhu; Li Qiu
Journal:  Quant Imaging Med Surg       Date:  2019-02

8.  Elastic modulus of muscle and tendon with shear wave ultrasound elastography: variations with different technical settings.

Authors:  Brian Chin Wing Kot; Zhi Jie Zhang; Arthur Wai Chun Lee; Vivian Yee Fong Leung; Siu Ngor Fu
Journal:  PLoS One       Date:  2012-08-31       Impact factor: 3.240

9.  Principles and clinical application of ultrasound elastography for diffuse liver disease.

Authors:  Woo Kyoung Jeong; Hyo K Lim; Hyoung-Ki Lee; Jae Moon Jo; Yongsoo Kim
Journal:  Ultrasonography       Date:  2014-03-24

10.  Influencing Factors of 2D Shear Wave Elastography of the Muscle - An Ex Vivo Animal Study.

Authors:  Marga B Rominger; Pascal Kälin; Monika Mastalerz; Katharina Martini; Volker Klingmüller; Sergio Sanabria; Thomas Frauenfelder
Journal:  Ultrasound Int Open       Date:  2018-07-05
View more
  1 in total

1.  Assessment of ultrasound shear wave elastography within muscles using different region of interest sizes, manufacturers, probes and acquisition angles: an ex vivo study.

Authors:  Xiuming Wang; Jiaan Zhu; Junxue Gao; Yue Hu; Yiqun Liu; Wenxue Li; Si Chen; Feifei Liu
Journal:  Quant Imaging Med Surg       Date:  2022-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.