Screening of obstructive sleep apnea in patients who snore using a patch-type device with electrocardiogram and 3-axis accelerometer.

Ying-Shuo Hsu1,2, Tien-Yu Chen3,4, Dean Wu5,6,7, Chia-Mo Lin8,9,10, Jer-Nan Juang11, Wen-Te Liu5,11,12,13,14,15.   

Abstract

STUDY
OBJECTIVES: People with obstructive sleep apnea (OSA) remain undiagnosed because of the lack of easy and comfortable screening tools. Through this study, we aimed to compare the diagnostic accuracy of chest wall motion and cyclic variation of heart rate (CVHR) in detecting OSA by using a single-lead electrocardiogram (ECG) patch with a 3-axis accelerometer.
METHODS: In total, 119 patients who snore simultaneously underwent polysomnography with a single-lead ECG patch. Signals of chest wall motion and CVHR from the single-lead ECG patch were collected. The chest effort index (CEI) was calculated using the chest wall motion recorded by a 3-axis accelerometer in the device. The ability of CEI and CVHR indices in diagnosing moderate-to-severe OSA (apnea-hypopnea index ≥ 15) was compared using the area under the curve (AUC) by using the DeLong test.
RESULTS: CVHR detected moderate-to-severe OSA with 52.9% sensitivity and 94.1% specificity (AUC: 0.76, 95% confidence interval: 0.67-0.84, optimal cutoff: 21.2 events/h). By contrast, CEI identified moderate-to-severe OSA with 80% sensitivity and 79.4% specificity (AUC: 0.87, 95% confidence interval: 0.80-0.94, optimal cutoff: 7.1 events/h). CEI significantly outperformed CVHR regarding the discrimination ability for moderate-to-severe OSA (ΔAUC: 0.11, 95% confidence interval: 0.009-0.21, P = .032). For determining severe OSA, the performance of discrimination ability was greater (AUC = 0.90, 95% confidence interval: 0.85-0.95) when combining these two signals.
CONCLUSIONS: Both CEI and CVHR recorded from a patch-type device with ECG and a 3-axis accelerometer can be used to detect moderate-to-severe OSA. Thus, incorporation of CEI is helpful in the detection of sleep apnea by using a single-lead ECG with a 3-axis accelerometer.
© 2020 American Academy of Sleep Medicine.

Entities:  

Keywords:  apnea-hypopnea index; chest wall motion; cyclic variation of heart rate; electrocardiogram; obstructive sleep apnea; sleep apnea; sleep-disordered breathing

Mesh:

Year:  2020        PMID: 32267228      PMCID: PMC7954076          DOI: 10.5664/jcsm.8462

Source DB:  PubMed          Journal:  J Clin Sleep Med        ISSN: 1550-9389            Impact factor:   4.062


  36 in total

1.  The significance of the sympathetic nervous system in the pathophysiology of periodic leg movements in sleep.

Authors:  Adrian G Guggisberg; Christian W Hess; Johannes Mathis
Journal:  Sleep       Date:  2007-06       Impact factor: 5.849

Review 2.  Estimation of the global prevalence and burden of obstructive sleep apnoea: a literature-based analysis.

Authors:  Adam V Benjafield; Najib T Ayas; Peter R Eastwood; Raphael Heinzer; Mary S M Ip; Mary J Morrell; Carlos M Nunez; Sanjay R Patel; Thomas Penzel; Jean-Louis Pépin; Paul E Peppard; Sanjeev Sinha; Sergio Tufik; Kate Valentine; Atul Malhotra
Journal:  Lancet Respir Med       Date:  2019-07-09       Impact factor: 30.700

3.  Pathogenesis of upper airway occlusion during sleep.

Authors:  J E Remmers; W J deGroot; E K Sauerland; A M Anch
Journal:  J Appl Physiol Respir Environ Exerc Physiol       Date:  1978-06

4.  Readmission after Adult Uvulopalatopharyngoplasty: A Population-Based Inpatient Cohort Study in Taiwan.

Authors:  Ying-Shuo Hsu; Wei-Chung Hsu; Jenq-Yuh Ko; Te-Huei Yeh; Chia-Hsuan Lee; Kun-Tai Kang
Journal:  Otolaryngol Head Neck Surg       Date:  2018-11-27       Impact factor: 3.497

5.  Cardiovascular variability during periodic leg movements: a spectral analysis approach.

Authors:  Emilia Sforza; Vincent Pichot; Jean Claude Barthelemy; José Haba-Rubio; Frederic Roche
Journal:  Clin Neurophysiol       Date:  2005-05       Impact factor: 3.708

6.  Clinical Practice Guideline for Diagnostic Testing for Adult Obstructive Sleep Apnea: An American Academy of Sleep Medicine Clinical Practice Guideline.

Authors:  Vishesh K Kapur; Dennis H Auckley; Susmita Chowdhuri; David C Kuhlmann; Reena Mehra; Kannan Ramar; Christopher G Harrod
Journal:  J Clin Sleep Med       Date:  2017-03-15       Impact factor: 4.062

7.  The National Healthy Sleep Awareness Project Sleep Health Surveillance Questionnaire as an Obstructive Sleep Apnea Surveillance Tool.

Authors:  Youngsin Jung; Mithri R Junna; Jayawant N Mandrekar; Timothy I Morgenthaler
Journal:  J Clin Sleep Med       Date:  2017-09-15       Impact factor: 4.062

8.  Cyclical variation of the heart rate in sleep apnoea syndrome. Mechanisms, and usefulness of 24 h electrocardiography as a screening technique.

Authors:  C Guilleminault; S Connolly; R Winkle; K Melvin; A Tilkian
Journal:  Lancet       Date:  1984-01-21       Impact factor: 79.321

9.  Support vector machines for automated recognition of obstructive sleep apnea syndrome from ECG recordings.

Authors:  Ahsan H Khandoker; Marimuthu Palaniswami; Chandan K Karmakar
Journal:  IEEE Trans Inf Technol Biomed       Date:  2009-01

10.  Ambulatory screening tool for sleep apnea: analyzing a single-lead electrocardiogram signal (ECG).

Authors:  Solveig Magnusdottir; Hugi Hilmisson
Journal:  Sleep Breath       Date:  2017-09-07       Impact factor: 2.816

View more
  4 in total

Review 1.  Assessment of autonomic function by long-term heart rate variability: beyond the classical framework of LF and HF measurements.

Authors:  Junichiro Hayano; Emi Yuda
Journal:  J Physiol Anthropol       Date:  2021-11-30       Impact factor: 2.867

2.  At-home wireless monitoring of acute hemodynamic disturbances to detect sleep apnea and sleep stages via a soft sternal patch.

Authors:  Nathan Zavanelli; Hojoong Kim; Jongsu Kim; Robert Herbert; Musa Mahmood; Yun-Soung Kim; Shinjae Kwon; Nicholas B Bolus; F Brennan Torstrick; Christopher S D Lee; Woon-Hong Yeo
Journal:  Sci Adv       Date:  2021-12-22       Impact factor: 14.136

3.  Night-to-night variability of sleep apnea detected by cyclic variation of heart rate during long-term continuous ECG monitoring.

Authors:  Junichiro Hayano; Emi Yuda
Journal:  Ann Noninvasive Electrocardiol       Date:  2021-10-18       Impact factor: 1.468

4.  Comparison of Hospital-Based and Home-Based Obstructive Sleep Apnoea Severity Measurements with a Single-Lead Electrocardiogram Patch.

Authors:  Wen-Te Liu; Shang-Yang Lin; Cheng-Yu Tsai; Yi-Shin Liu; Wen-Hua Hsu; Arnab Majumdar; Chia-Mo Lin; Kang-Yun Lee; Dean Wu; Yi-Chun Kuan; Hsin-Chien Lee; Cheng-Jung Wu; Wun-Hao Cheng; Ying-Shuo Hsu
Journal:  Sensors (Basel)       Date:  2021-12-03       Impact factor: 3.576

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.