Literature DB >> 3224775

Three studies comparing performance of the ER-3A tubephone with the TDH-50P earphone.

J L Clark1, R J Roeser.   

Abstract

Three studies compared the performance of the ER-3A Tubephone insert earphone and the TDH-50P-MX41/AR supra-aural earphone. The three factors addressed were: threshold differences in children 7 to 10 yr old compared to adults, differences in real ear attenuation, and threshold differences in the presence of high background noise levels. The influence of insertion depth of the ER-3A Tubephone was also investigated. Findings showed no significant threshold differences between children and adults, significantly better real ear attenuation for the ER-3A Tubephone, and significantly better thresholds for the ER-3A in the presence of high background noise levels. Most critically, there was a significant change in attenuation characteristics of the ER-3A Tubephone, which was dependent on the insertion depth of the ear-tip.

Entities:  

Mesh:

Year:  1988        PMID: 3224775     DOI: 10.1097/00003446-198810000-00007

Source DB:  PubMed          Journal:  Ear Hear        ISSN: 0196-0202            Impact factor:   3.570


  1 in total

1.  Going wireless and booth-less for hearing testing in industry.

Authors:  Deanna K Meinke; Jesse A Norris; Brendan P Flynn; Odile H Clavier
Journal:  Int J Audiol       Date:  2016-12-15       Impact factor: 2.117