Peak deflection index as a predictor of a free-wall implantation of contemporary leadless pacemakers.

Takatsugu Kajiyama1, Yusuke Kondo2, Masahiro Nakano2, Kazuo Miyazawa3, Miyo Nakano3, Tomohiko Hayashi3, Ryo Ito3, Haruhiro Takahira3, Mari Kitagawa3, Yoshio Kobayashi3.   

Abstract

BACKGROUND: Leadless pacemakers are an effective treatment for bradycardia. However, some cases exhibit pericardial effusions, presumably associated with device implantations on the right ventricular free-wall. The present study was carried out to find the ECG features during ventricular pacing with a Micra, which enabled distinguishing free-wall implantations from septal implantations without using imaging modalities.
METHODS: Thirty-one consecutive patients who received Micra implantations in our facility were enrolled. The location of the device in the right ventricle was evaluated using echocardiography or computed tomography in order to determine whether the device was implanted on the septum (Sep group), apex (Apex group), or free-wall (FW group). The differences in the 12-lead ECG during ventricular pacing by the Micra were analyzed between the Sep and FW groups.
RESULTS: The body of the Micra was clearly identifiable in 22 patients. The location of the device was classified into Sep in 12 patients, Apex in 4, and FW in 6. The mean age was highest in the FW and lowest in the Sep group (82.7 ± 6.6 vs. 72.8 ± 8.7 years, p = 0.027). The peak deflection index (PDI) was significantly larger in the FW group than Sep/Apex group in lead V1 (Sep: 0.505 ± 0.010, Apex: 0.402 ± 0.052, FW: 0.617 ± 0.043, p = 0.004) and lead V2 (Sep: 0.450 ± 0.066, Apex: 0.409 ± 0.037, FW: 0.521 ± 0.030, p = 0.011), whereas there was no difference in the QRS duration, transitional zone, and QRS notching.
CONCLUSION: The PDI in V1 could be useful for predicting implantations of Micra devices on the free-wall and may potentially stratify the risk of postprocedural pericardial effusions.

Entities:  

Keywords:  Electrocardiogram; Leadless pacemaker; Peak deflection index; Ventricular pacing

Year:  2020        PMID: 32242303     DOI: 10.1007/s10840-020-00724-y

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  1 in total

1.  Individualized left anterior oblique projection based on pigtail catheter visualization facilitates leadless pacemaker implantation.

Authors:  Kazuya Yamao; Hitoshi Hachiya; Shigeki Kusa; Naoyuki Miwa; Yoshikazu Sato; Satoshi Hara; Hiroaki Ohya; Yoshito Iesaka; Tetsuo Sasano
Journal:  J Arrhythm       Date:  2021-05-05
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.