BACKGROUND: Cardiovascular issues (especially arrhythmia and sudden cardiac death) are one of the most common causes of mortality in patients with chronic kidney disease (CKD). To minimize cardiac mortality, these patients frequently require various cardiac devices, such as pacemakers, loop recorders, and defibrillators which can compromise their vascular access. In this study, we aim to determine the prevalence of CKD in patients undergoing cardiac device placement and their progression of CKD. METHODS: Institutional review board approval was obtained for this study. A total of 688 patients undergoing cardiac device placement were included in this study over a 3-year period at Jersey Shore University Medical Center. Demographic characteristics, comorbidities, base-line renal functions during the procedure, types of cardiac devices, sites of vascular access and follow-up renal function when available were assessed retrospectively. Patients were categorized into CKD stages 1 - 5 based on the National Kidney Foundation-Kidney Disease Outcomes Quality Initiative (NKF-KDOQI) guidelines. The patients who were already on hemodialysis were excluded in this study. RESULTS: The average age of the patient were 73.9 years with male predominance (60%). A total of 227 patients (33%) had estimated glomerular filtration rate (eGFR) < 60 mL/min consistent with the evidence of advanced-stage CKD (stages 3 - 5) at the time of cardiac device placement. The most common types of device placements were new insertion/replacement of atrial and ventricular leads (39.5%), loop recorder implantation (21.1%) and generator changes on an already implanted device (11%). Only 4% (28/688) had a leadless cardiac device placement. The most common access sites were subclavian (47.1%), axillary (32.3%) and femoral (12.2%). CONCLUSIONS: The present study demonstrated that nearly one-third of the patient undergoing cardiac device placement had an advanced degree of renal failure. Because CKD is a progressive disease, many of these patients might require renal replacement therapy in the future. Transvenous devices is not a good choice in this group of patients as they will ultimately require an arteriovenous fistula. Subcutaneous leadless cardiac device insertion might be a better option in patients with advanced CKD. Copyright 2020, Hossain et al.
BACKGROUND: Cardiovascular issues (especially arrhythmia and sudden cardiac death) are one of the most common causes of mortality in patients with chronic kidney disease (CKD). To minimize cardiac mortality, these patients frequently require various cardiac devices, such as pacemakers, loop recorders, and defibrillators which can compromise their vascular access. In this study, we aim to determine the prevalence of CKD in patients undergoing cardiac device placement and their progression of CKD. METHODS: Institutional review board approval was obtained for this study. A total of 688 patients undergoing cardiac device placement were included in this study over a 3-year period at Jersey Shore University Medical Center. Demographic characteristics, comorbidities, base-line renal functions during the procedure, types of cardiac devices, sites of vascular access and follow-up renal function when available were assessed retrospectively. Patients were categorized into CKD stages 1 - 5 based on the National Kidney Foundation-Kidney Disease Outcomes Quality Initiative (NKF-KDOQI) guidelines. The patients who were already on hemodialysis were excluded in this study. RESULTS: The average age of the patient were 73.9 years with male predominance (60%). A total of 227 patients (33%) had estimated glomerular filtration rate (eGFR) < 60 mL/min consistent with the evidence of advanced-stage CKD (stages 3 - 5) at the time of cardiac device placement. The most common types of device placements were new insertion/replacement of atrial and ventricular leads (39.5%), loop recorder implantation (21.1%) and generator changes on an already implanted device (11%). Only 4% (28/688) had a leadless cardiac device placement. The most common access sites were subclavian (47.1%), axillary (32.3%) and femoral (12.2%). CONCLUSIONS: The present study demonstrated that nearly one-third of the patient undergoing cardiac device placement had an advanced degree of renal failure. Because CKD is a progressive disease, many of these patients might require renal replacement therapy in the future. Transvenous devices is not a good choice in this group of patients as they will ultimately require an arteriovenous fistula. Subcutaneous leadless cardiac device insertion might be a better option in patients with advanced CKD. Copyright 2020, Hossain et al.
Chronic kidney disease (CKD) affects 30 million people or 15% of the US population [1]. Cardiovascular issues (especially arrhythmia and sudden cardiac death) are one of the most common causes of mortality in patients with CKD [2, 3]. To mitigate cardiac complications, these patients often require various cardiac devices, such as pacemakers, loop recorders and defibrillators. The placement of these devices poses multiple predicaments in renal failure patients. Many of these implantable cardiac devices have leads that can cause central venous stenosis and thrombosis which is very critical for renal patients as compared to the general population [4-8]. Because CKD is a progressive disease, these patients depend on patent central venous access for the creation of an arteriovenous (AV) fistula in the near future [4]. Loss of this venous access can be an important factor for their ultimate survival. In addition to that, abnormal renal function is an independent risk factor for cardiac implantable electronic device (CIED) infection [9-11]. In this study, we aim to determine the prevalence of CKD in patients undergoing cardiac device placement and their progression of CKD.
Materials and Methods
A total of 688 patients who underwent cardiac device placement at Jersey Shore University Medical Center over a 3-year period (June 2015 to June 2018) were included in this study. Demographic characteristics, comorbidities, base-line renal functions at the time procedure, types of cardiac device, and insertion sites were assessed retrospectively. Estimated glomerular filtration rate (eGFR) was calculated by the Chronic Kidney Disease Epidemiology Collaboration (CKD-EPI) creatinine equation for all of the patients to ascertain the degree of advanced CKD. Advanced CKD was defined as CKD stage greater than 3 or more. Patients were categorized into CKD stages 1 - 5 based on the National Kidney Foundation Dialysis Outcomes Quality Initiative (NKF-KDOQI) guidelines [12].Institutional review board approval was obtained for this study. All study procedures were carried out in accordance with the Declaration of Helsinki regarding research involving human subjects. The summary statistics of continuous variables were reported as mean ± standard deviation.
Results
A total of 688 patients underwent cardiac device placement at Jersey Shore University Medical Center over a 3-year period. Demographic characteristics revealed that the average age of the patient was 73.9 years with male predominance (60%) (Table 1). Major comorbidities included hypertension (76.2%), hyperlipidemia (66.4%), cardiovascular disease (44.4%) and congestive heart failure (44.9%) (Table 1). Most common types of device placements were new insertion or replacement of atrial and ventricular leads (39.5%), loop recorder implantation (21.1%) and generator changes on an already implanted devices (11%), and only 4% (28/688) had leadless cardiac device placement (Table 2). The most common insertion sites were subclavian (47.1%), axillary (32.3%) and femoral (12.2%). A total of 227 patients (33%) had evidence of stage 3 - 5 CKD (stage 3A GFR = 45 - 59 mL/min (n = 116); stage 3B GFR = 30 - 44 mL/min (n = 69); stage 4 GFR = 15 - 29 mL/min (n = 23); stage 5 GFR < 15 mL/min (n = 19)) at the time of cardiac device placement (Table 3).
Table 1
Demographics of Patient Undergoing Cardiac Device Placement (n = 688)
Age
73.9 years
Gender
Men
414 (60%)
Women
274 (40%)
Race
White
636 (92.4%)
Black
40 (5.8%)
Asian
4 (0.6%)
Hispanic
3 (0.4%)
Other
5 (0.7%)
Congestive heart failure
309 (45%)
Diabetes
165 (24%)
Renal disease
227 (33%)
Dialysis
20 (3%)
Cardiovascular disease
399 (58%)
Cerebrovascular accident
114 (17%)
Peripheral vascular disease
46 (7%)
Lung disease
135 (20%)
Hyperlipidemia
457 (66%)
Hypertension
524 (76%)
Table 2
Types of Cardiac Device and Sites of Access (n = 688)
Types of Access
Radial
5 (0.7%)
Femoral
84 (12%)
Axillary
222 (32%)
Subclavian
324 (47%)
Subcutaneous
50 (7%)
Other
3 (0.4%)
Procedure Types
Insertion or replacement of atrial and ventricular leads
272 (39.5%)
Loop recorder implant
145 (21.1%)
Loop recorder explant
23 (3.3%)
Leadless permanent pacemaker
28 (4.1%)
Watchman
43 (6.3%)
Lead repositioning
9 (1.3%)
Generator changes
76 (11.0%)
New bi-ventricular system
49 (7.1%)
Upgrade to bi-ventricular system
24 (6.2%)
Other
19 (2.8%)
Table 3
Stages of CKD in Patients’ Who Underwent Cardiac Device Placement
Stages of CKD
Number of patients
eGFRa
Stage 3A
116 (17%)
45 - 59 mL/min
Stage 3B
69 (10%)
30 - 44 mL/min
Stage 4
23(3%)
15 - 29 mL/min
Stage 5
19 (3%)
< 15 mL/min
Stage 1 and 2 CKD are not included.CKD: chronic kidney disease; eGFR: estimated glomerular filtration rate. aeGFR reported by labs > 60 mL/min.
Stage 1 and 2 CKD are not included.CKD: chronic kidney disease; eGFR: estimated glomerular filtration rate. aeGFR reported by labs > 60 mL/min.
Discussion
This study finds that CKD stages 3 - 5 presented in nearly 33% of patients undergoing cardiac device implantation. Our results are consistent with the previous study that demonstrated a similar prevalence of CKD in patient undergoing CIED removal [13].Recognizing the high prevalence of advanced CKD in this population is important to minimize the risk of device infection. The most common sites of insertion for the cardiac device were subclavian (47.1%), axillary (32.3%) and femoral (12.2%). Almost 80% of the patient had their device placed through central veins. While placing cardiac devices through central veins is a common pathway, the leads can cause central vein stenosis (CVS) and carry the risk of tricuspid regurgitation by valve adhesion, perforation or entanglement [5-8, 14]. This point is critically important as many patients will progress to end-stage renal disease and require an AV fistula [10]. The placement of AV fistula requires the preservation of the vascular system in an upper extremity. In our study, a high prevalence of CKD patients who underwent a cardiac device procedure is at a higher likelihood of requiring hemodialysis. The development of central venous stenosis may prevent the creation of the life time (i.e. AV fistula) for a dialysis patient.To preserve the central vascular access in CKD patient who are planning to have cardiac devices, placement of the leadless cardiac device and the use of the epicardial route is an important alternative to avoid the transvenous pathway. The validity of this approach was provided by recent studies [15-18]. Data from epicardial studies emphasized that the effectiveness and survival of epicardial leads are essentially the same as the endocardial leads [15-18]. Recently, a subcutaneous approach has also been reported in hemodialysis patients [16-18]. In our study only 4% (28/688) patients had leadless implanted cardiac devices (ICDs) implanted. Leadless ICDs do not use central veins and are located in the subcutaneous tissue. In this way, they avoid the risk of CVS.
Conclusions
Our study showed that a significant amount of patients undergoing cardiac device placement had advanced CKD. Because CKD is a progressive disease, many of these patients might require renal replacement therapy in the future. Transvenous devices is not a good choice in this group of patients as they cause CVS. This complication can prevent the creation of an AV fistula. The subcutaneous leadless cardiac device may be a better option in a patient with advanced CKD.
Authors: Gust H Bardy; Warren M Smith; Margaret A Hood; Ian G Crozier; Iain C Melton; Luc Jordaens; Dominic Theuns; Robert E Park; David J Wright; Derek T Connelly; Simon P Fynn; Francis D Murgatroyd; Johannes Sperzel; Jörg Neuzner; Stefan G Spitzer; Andrey V Ardashev; Amo Oduro; Lucas Boersma; Alexander H Maass; Isabelle C Van Gelder; Arthur A Wilde; Pascal F van Dessel; Reinoud E Knops; Craig S Barr; Pierpaolo Lupo; Riccardo Cappato; Andrew A Grace Journal: N Engl J Med Date: 2010-05-12 Impact factor: 91.245
Authors: Christine Tompkins; Rhondalyn McLean; Alan Cheng; Jeffrey A Brinker; Joseph E Marine; Saman Nazarian; David D Spragg; Sunil Sinha; Henry Halperin; Gordon F Tomaselli; Ronald D Berger; Hugh Calkins; Charles A Henrikson Journal: J Cardiovasc Electrophysiol Date: 2011-04-13
Authors: Gorav Ailawadi; Damien J Lapar; Brian R Swenson; Cory D Maxwell; Micah E Girotti; James D Bergin; John A Kern; John P Dimarco; Srijoy Mahapatra Journal: Heart Rhythm Date: 2010-01-20 Impact factor: 6.343
Authors: Rajiv K Dhamija; Henry Tan; Edward Philbin; Roy O Mathew; Mandeep S Sidhu; Jeffrey Wang; Basil Saour; Syed S Haqqie; Gerald Beathard; Alexander S Yevzlin; Loay Salman; William E Boden; Gary Siskin; Arif Asif Journal: Am J Kidney Dis Date: 2015-04-22 Impact factor: 8.860
Authors: Heather Bloom; Brian Heeke; Angel Leon; Fernando Mera; David Delurgio; John Beshai; Jonathan Langberg Journal: Pacing Clin Electrophysiol Date: 2006-02 Impact factor: 1.976
Authors: J C Lekkerkerker; C van Nieuwkoop; S A Trines; J G van der Bom; A Bernards; E T van de Velde; M Bootsma; K Zeppenfeld; J W Jukema; J-W Borleffs; M J Schalij; L van Erven Journal: Heart Date: 2008-11-26 Impact factor: 5.994
Authors: Theodore H Teruya; Ahmed M Abou-Zamzam; Whitney Limm; Linda Wong; Livingston Wong Journal: Ann Vasc Surg Date: 2003-09-10 Impact factor: 1.466
Authors: Lesley A Inker; Brad C Astor; Chester H Fox; Tamara Isakova; James P Lash; Carmen A Peralta; Manjula Kurella Tamura; Harold I Feldman Journal: Am J Kidney Dis Date: 2014-03-16 Impact factor: 8.860