Cardiac anatomical axes by CT scan and confirmation of the accuracy of fluoroscopic individualized left anterior oblique projection for right ventricular lead implantation.

Fabien Squara1, Etienne Fourrier2, Yann Diascorn3, Alexandre Poulard2, Didier Scarlatti2, Sok-Sithikun Bun2, Pamela Moceri2, Emile Ferrari2.   

Abstract

BACKGROUND: The fluoroscopic individualized LAO (i-LAO) projection has demonstrated high accuracy for identifying right ventricular (RV) lead positioning, likely by approximating a view along the septal or RV long axes. However, RV and septal anatomical axes have not been studied, and their relation with i-LAO is unknown. We sought to determine RV, septal, and left ventricular (LV) long-axis orientations by CT scan and to compare them to the i-LAO angle, to confirm the anatomical relevance of i-LAO.
METHODS: We prospectively included patients (pts) for whom i-LAO angle was determined during pacemaker or defibrillator implant. Then, RV, septal, and LV long-axis orientations were determined by CT scan by a physician blinded to i-LAO data. The horizontal components of the cardiac axes were compared with those of the i-LAO angle.
RESULTS: We included 26 pts. Median values were 57.5° for i-LAO angle (range 47.5-70), 64.5° for RV axis (range 48-90), 51.5° for septal axis (range 39-74), and 37° for LV axis (range 25-67). i-LAO angle best correlated with septal axis (r = 0.91 and ρc = 0.71). Up to an angle of 70° (maximal measurable i-LAO value; 23/26 pts), the i-LAO angle was comprised between the septal and the RV axes (21/23 pts, 91.3%), or within 2° of this interval (2/23 pts, 8.7%).
CONCLUSIONS: RV and septal anatomical axes present major interindividual variations, prompting the use of individualized fluoroscopy criteria for lead implantation. i-LAO angle demonstrated to be almost constantly between the septal and RV long axes, thus confirming its anatomical relevance for RV lead implantation.

Entities:  

Keywords:  Fluoroscopy; Implantable cardiac defibrillator; Pacemaker

Year:  2020        PMID: 32219588     DOI: 10.1007/s10840-020-00729-7

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900