Literature DB >> 32219150

Measurement of the x-ray effective focal spot size with edge response analysis using digital detectors.

Masayuki Nishiki1, Satoshi Yanagita2, Noriko Nishikawa3, Hitomi Sugimura4.   

Abstract

Purpose: For the focal spot measurement of x-ray tubes, we propose a practical method in which only a metal edge and a digital detector are used, together with a process of removing detector blur inherently associated. Approach: The evaluation was made through the optical transfer function (OTF) measurements using the edge response of a 1-mm-thick tungsten plate. First, we made the acquisition of a geometrically magnified edge response, which consists of focal spot penumbra and detector blur, followed by the acquisition of nonmagnified edge response, which includes only detector blur. Then the detector blur was removed by taking the ratio of the two OTFs. Finally, the focal spot profile was obtained by the inverse Fourier transform of this ratio.
Results: Resultant full widths at the half-maximum of a small focus profile were 0.529 ± 0.005    mm for the proposed method and 0.527 ± 0.020    mm for the conventional slit method with film, indicating excellent agreement between both methods. Comparing between results obtained using two flat panel detectors with different pixel pitches (0.143 and 0.175 mm) confirmed no differences with these variations.
Conclusion: Through the whole study, the accuracy and the practicality of the proposed method were demonstrated, indicating a possibility of the method to be widely used to evaluate the effective focal spot size and profile of x-ray systems.
© 2020 Society of Photo-Optical Instrumentation Engineers (SPIE).

Entities:  

Keywords:  detector blur; edge response; flat panel; focal spot; slit camera; x-ray tube

Year:  2020        PMID: 32219150      PMCID: PMC7086143          DOI: 10.1117/1.JMI.7.2.023502

Source DB:  PubMed          Journal:  J Med Imaging (Bellingham)        ISSN: 2329-4302


  10 in total

1.  Accuracy of a simple method for deriving the presampled modulation transfer function of a digital radiographic system from an edge image.

Authors:  Egbert Buhr; Susanne Günther-Kohfahl; Ulrich Neitzel
Journal:  Med Phys       Date:  2003-09       Impact factor: 4.071

2.  Measurement of focal spot size with slit camera using computed radiography and flat-panel based digital detectors.

Authors:  Xiujiang J Rong; Kerry T Krugh; S Jeff Shepard; William R Geiser
Journal:  Med Phys       Date:  2003-07       Impact factor: 4.071

3.  New device for accurate measurement of the x-ray intensity distribution of x-ray tube focal spots.

Authors:  K Doi; B Fromes; K Rossmann
Journal:  Med Phys       Date:  1975 Sep-Oct       Impact factor: 4.071

4.  X-ray focal spot reconstruction by circular penumbra analysis-Application to digital radiography systems.

Authors:  Giovanni Di Domenico; Paolo Cardarelli; Adriano Contillo; Angelo Taibi; Mauro Gambaccini
Journal:  Med Phys       Date:  2016-01       Impact factor: 4.071

5.  Comparison of edge analysis techniques for the determination of the MTF of digital radiographic systems.

Authors:  Ehsan Samei; Egbert Buhr; Paul Granfors; Dirk Vandenbroucke; Xiaohui Wang
Journal:  Phys Med Biol       Date:  2005-07-19       Impact factor: 3.609

6.  Intercomparison of methods for image quality characterization. I. Modulation transfer function.

Authors:  Ehsan Samei; Nicole T Ranger; James T Dobbins; Ying Chen
Journal:  Med Phys       Date:  2006-05       Impact factor: 4.071

7.  Method for measuring the focal spot size of an x-ray tube using a coded aperture mask and a digital detector.

Authors:  Paolo Russo; Giovanni Mettivier
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

8.  A method for measuring the presampled MTF of digital radiographic systems using an edge test device.

Authors:  E Samei; M J Flynn; D A Reimann
Journal:  Med Phys       Date:  1998-01       Impact factor: 4.071

9.  Focal-spot measurement: comparison of slit, pinhole, and star resolution pattern techniques.

Authors:  J D Everson; J E Gray
Journal:  Radiology       Date:  1987-10       Impact factor: 11.105

10.  Focal spot measurements using a digital flat panel detector.

Authors:  Amit Jain; A Panse; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19
  10 in total
  1 in total

1.  Characterization of x-ray focal spots using a rotating edge.

Authors:  Linxi Shi; N Robert Bennett; Adam S Wang
Journal:  J Med Imaging (Bellingham)       Date:  2021-03-25
  1 in total