Alberto Prud'homme1, Frederic Nabki1. 1. Department of Electrical Engineering, École de Technologie Supérieure, Montreal, QC H3C 1K3, Canada.
Abstract
The comparison between potential coatings for the measurement of CO2 concentration through the frequency shift in micro-resonators is presented. The polymers evaluated are linear polyethylenimine, branched polyethylenimine and reduced graphene oxide (rGO) by microwave reduction with polyethylenimine. The characterization of the coatings was made by using 6 MHz gold-plated quartz crystals, and a proof-of-concept sensor is shown with a diaphragm electrostatic microelectromechanical systems (MEMS) resonator. The methods of producing the solutions of the polymers deposited onto the quartz crystals are presented. A CO2 concentration range from 0.05 % to 1 % was dissolved in air and humidity level were controlled and evaluated. Linear polyethylenimine showed superior performance with a reaction time obtained for stabilization after the concentration increase of 345 s, while the time for recovery was of 126 s, with a maximum frequency deviation of 33.6 Hz for an in-air CO2 concentration of 0.1%.
The comparison between potential n class="Chemical">coatings for the measurement of CO2concentration through the frequency shift in micro-resonators is presented. The polymers evaluated are linear polyethylenimine, branched polyethylenimine and reduced graphene oxide (rGO) by microwave reduction with polyethylenimine. The characterization of the coatings was made by using 6 MHz gold-plated quartz crystals, and a proof-of-concept sensor is shown with a diaphragm electrostatic microelectromechanical systems (MEMS) resonator. The methods of producing the solutions of the polymers deposited onto the quartz crystals are presented. A CO2concentration range from 0.05 % to 1 % was dissolved in air and humidity level were controlled and evaluated. Linear polyethylenimine showed superior performance with a reaction time obtained for stabilization after the concentration increase of 345 s, while the time for recovery was of 126 s, with a maximum frequency deviation of 33.6 Hz for an in-air CO2concentration of 0.1%.
Entities:
Keywords:
CO2 sensor; coatings; gas sensor; humidity sensor; mass sensor; micro-resonator; polyethylenimine; reduced graphene oxide