A total of 284 patients with posterior chamber lenses (Sinskey-Kratz type) implanted in the capsular bag were re-examined on average 14 months after implantation. In 87% optical centration was excellent; 12% showed a minimal displacement exceeding 0.4 mm. In four patients (1%) gross displacement of the lens within the capsular bag was seen, caused by proliferation of residual lens cortex, asymmetric shrinkage of the capsular bag, and perforation of the bag by a lens loop.

Entities:  

Mesh:

Year:  1988        PMID: 3221655     DOI: 10.1055/s-2008-1050301

Source DB:  PubMed          Journal:  Klin Monbl Augenheilkd        ISSN: 0023-2165            Impact factor:   0.700


  1 in total

1.  Forces on intraocular lens haptics induced by capsular fibrosis. An experimental study.

Authors:  R Guthoff; F Abramo; J Draeger; L C Chumbley; G K Lang; W Neumann
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1990       Impact factor: 3.117