Wet-Side Breath-Enhanced Jet Nebulization: Controlling Drug Delivery During Mechanical Ventilation.

Ann D Cuccia1, Sunya Ashraf2, Michael McPeck2, Joshua Samuel2, Gerald C Smaldone2.   

Abstract

BACKGROUND: The present study tested a novel nebulizer and circuit that use breath enhancement and breath actuation to minimize ventilator influences. The unique circuit design incorporates "wet-side" jet nebulization (the nebulizer connected to the humidifier outlet port) to prevent unpredictable aerosol losses with active humidification. The system was studied using several ventilator brands over a wide range of settings, with and without humidification.
METHODS: During treatment, a 2-position valve directed all ventilator flow to the nebulizer, providing breath enhancement during inspiration. Aerosol was generated by air 50 psi 3.5 L/m triggered during inspiration by a pressure-sensitive circuit. Particles were captured on an inhaled mass filter. Testing was performed by using active humidification or bypassable valved heat and moisture exchanger (HME) over a range of breathing patterns, ventilator modes, and bias flows (0.5-5.0 L/m). The nebulizer was charged with 6 mL of radiolabeled saline solution. Mass balance was performed by using a gamma camera. Tidal volume was monitored by ventilator volume (exhaled VT) and test lung volume. The Mann-Whitney test was used.
RESULTS: A total of 6 mL was nebulized within 1 h. Inhaled mass (% neb charge): mean ± SD (all data) 31.1% ± 6.45; no. = 83. Small significant differences were seen with humidification for all modes (humidified 36.1% ± 5.60, no. = 26; bypassable valved HME 28.8% ± 5.51, no. = 57 [P < .001]), continuous mandatory ventilation modes [P < .001], and pressure support airway pressure release ventilation modes [P < .001]. Mass median aerodynamic diameter ranged from 1.04 to 1.34 μm. The VT was unaffected (exhaled VT -5.0 ± 12.9 mL; P = .75) and test lung (test lung volume 25 ± 14.5 mL; P = .13). Bias flow and PEEP had no effect.
CONCLUSIONS: Breath enhancement with breath actuation provided a predictable dose at any ventilator setting or type of humidification. Preservation of drug delivery during active humidification is a new finding, compared with previous studies. The use of wall gases and stand alone breath actuation standardizes conditions that drive the nebulizer independent of ventilator design. Wet-side nebulizer placement at the humidifier outlet allows delivery without introducing aerosol into the humidification chamber.
Copyright © 2020 by Daedalus Enterprises.

Entities:  

Keywords:  administration; aerosols; bias flow; drug delivery; humidifiers; inhalation; mechanical; nebulizers and vaporizers; ventilators

Mesh:

Substances:

Year:  2020        PMID: 32209704     DOI: 10.4187/respcare.07343

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  2 in total

Review 1.  The Unfulfilled Promise of Inhaled Therapy in Ventilator-Associated Infections: Where Do We Go from Here?

Authors:  Lucy B Palmer; Gerald C Smaldone
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2022-01-28       Impact factor: 3.440

2.  Real-Time In Vitro Assessment of Aerosol Delivery During Mechanical Ventilation.

Authors:  Michael McPeck; Janice A Lee; Ann D Cuccia; Gerald C Smaldone
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2021-07-06       Impact factor: 2.849

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.