The low acute effectiveness of a high-power short duration radiofrequency current application technique in pulmonary vein isolation for atrial fibrillation.

Ekrem Ücer1, Carsten Jungbauer2, Christian Hauck2, Manuel Kaufmann2, Florian Poschenrieder3, Lars Maier2, Sabine Fredersdorf2.   

Abstract

BACKGROUND: Application of high power radiofrequency (RF) energy for a short duration (HPSD) to isolate pulmonary vein (PV) is an emerging technique. But power and duration settings are very different across different centers. Moreover, despite encouraging preclinical and clinical data, studies measuring acute effectiveness of various HPSD settings are limited.
METHODS: Twenty-five consecutive patients with symptomatic atrial fibrillation (AF) were treated with pulmonary vein isolation (PVI) using HPSD. PVI was performed with a contact force catheter (Thermocool SF Smart-Touch) and Carto 3 System. The following parameters were used: energy output 50 W, target temperature 43°C, irrigation 15 mL/min, targeted contact force of > 10 g. RF energy was applied for 6 to 10 s. Required minimal interlesion distance was 4 mm. Twenty minutes after each successful PVI adenosine provocation test (APT) was performed by administrating 18 mg adenosine to unmask dormant PV conduction.
RESULTS: All PVs (100 PVs) were successfully isolated. RF lesions needed per patient were 131 ±  41, the average duration for each RF application was 8.1 ± 1.7 s. Procedure time was 138  ±  21 min and average of total RF energy duration was 16.3 ±  5.2 min and average amount of RF energy was 48209 ± 12808W s. APT application time after PVI was 31.1 ± 8.3 min for the left sided PVs and 22.2 ± 4.6 min (p = 0.005) for the right sided PVs. APT was transiently positive in 18 PVs (18%) in 8 (32%) patients.
CONCLUSIONS: Pulmonary vein isolation with high power for 6-10 s is feasible and shortens the procedure and ablation duration. However, acute effectiveness of the HPSD seems to be lower than expected. Further studies combining other ablation parameters are needed to improve this promising technique.

Entities:  

Keywords:  ablation; adenosine; atrial fibrillation; high power; reconnection

Year:  2020        PMID: 32207843     DOI: 10.5603/CJ.a2020.0033

Source DB:  PubMed          Journal:  Cardiol J        ISSN: 1898-018X            Impact factor:   2.737


  2 in total

1.  Meta-analysis comparing outcomes of high-power short-duration and low-power long-duration radiofrequency ablation for atrial fibrillation.

Authors:  Dibbendhu Khanra; Abdul Hamid; Saurabh Deshpande; Anindya Mukherjee; Sanjiv Petkar; Mohammad Saeed; Indranill Basu-Ray
Journal:  Anatol J Cardiol       Date:  2022-01       Impact factor: 1.596

Review 2.  Efficiency, Safety, and Efficacy of High-Power Short-Duration Radiofrequency Ablation in Patients with Atrial Fibrillation.

Authors:  Xuerong Sun; Jiang Lu; Jinxuan Lin; Tianjie Feng; Ni Suo; Lihui Zheng; Zhimin Liu; Gang Chen; Xiaohan Fan; Shu Zhang; Guodong Niu
Journal:  Cardiol Res Pract       Date:  2021-02-15       Impact factor: 1.866