Xirui Chen1,2, Yuankui Leng1,2, Liangwen Hao1,2, Hong Duan1,2, Jing Yuan1,2, Wenjing Zhang1,2, Xiaolin Huang1,2, Yonghua Xiong1,2,3.
Abstract
Background: Traditional lateral flow immunoassay (LFIA) based on 20-40 nm gold nanoparticles (AuNPs) as signal reporter always suffers from relatively low detection sensitivity due to its insufficient brightness, severely restricting its wide-ranging application in the detection of target analytes with trace concentration.Entities:
Keywords: colloidal gold superparticle; gold nanoparticle; lateral flow immunoassay; self-assembly; sensitivity
Mesh:
Substances:
Year: 2020 PMID: 32206119 PMCID: PMC7069069 DOI: 10.7150/thno.42364
Source DB: PubMed Journal: Theranostics ISSN: 1838-7640 Impact factor: 11.556
Summary of the detection performance of AuNP- and GSP-LFIA strip in detecting HCG.
| Labels | Linear range | R2 | LOD | Hook-effect | Regression equation |
|---|---|---|---|---|---|
| AuNP40 | 19.5-1250 | 0.9978 | 19.5 | >5000 | y = 0.0142x0.6674 |
| AuNPs80 | 7.8-1000 | 0.9921 | 3.90 | 2000 | y = 0.0129x0.85 |
| AuNPs120 | 1.9-1000 | 0.9955 | 0.97 | 2000 | y = 0.0082x0.8577 |
| AuNPs180 | 3.9-1000 | 0.9912 | 1.95 | 2000 | y = 0.036x0.8912 |
| GSP100 | 15.6-2000 | 0.9914 | 7.80 | 4000 | y = 0.0223x0.6448 |
| GSP160 | 3.9-2000 | 0.9949 | 1.03 | 4000 | y = 0.0772x0.4695 |
| GSP200 | 1.9-1000 | 0.9912 | 0.90 | 2000 | y = 0.0563x0.5942 |
| GSP270 | 0.49-1000 | 0.9961 | 0.49 | 2000 | y = 0.0968x0.5538 |
| GSP400 | 1.95-2000 | 0.9989 | 1.80 | 8000 | y = 0.2156x0.3802 |
Precision analysis of the developed GSP270-LFIA test strip in testing HBsAg-positive serum with concentrations of 10, 50, 100, 200, and 500 ng/mL.
| Diluted | Intra-assay precisiona) | Inter-assay precisionb) | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Recovered | CV (%) | Recovery | Recovered concentration | CV | Recovery | |||||
| 500 | 491.8±9.89 | 2.01 | 98.36 | 502.89±65.32 | 12.99 | 100.58 | ||||
| 200 | 162.83±10.52 | 6.46 | 81.42 | 159.05±16.17 | 10.16 | 79.53 | ||||
| 100 | 105.08±8.27 | 7.87 | 105.08 | 110.17±6.98 | 6.34 | 110.17 | ||||
| 50 | 43.27±4.72 | 10.91 | 86.54 | 48.89±3.24 | 6.62 | 97.78 | ||||
| 10 | 9.02±1.03 | 11.41 | 90.20 | 9.17±1.23 | 13.41 | 91.70 | ||||
a) The assay was carried out in triplicates on the same day; b) The assay was performed on three consecutive days.