| Literature DB >> 32203456 |
Claudia Cea1, George D Spyropoulos1, Patricia Jastrzebska-Perfect1, José J Ferrero2, Jennifer N Gelinas3,4, Dion Khodagholy5.
Abstract
Bioelectronic devices must be fast and sensitive to interact with the rapid, low-amplitude signals generated by neural tissue. They should also be biocompatible and soft, and should exhibit long-term stability in physiologic environments. Here, we develop an enhancement-mode, internal ion-gated organic electrochemical transistor (e-IGT) based on a reversible redox reaction and hydrated ion reservoirs within the conducting polymer channel, which enable long-term stable operation and shortened ion transit time. E-IGT transient responses depend on hole rather than ion mobility, and combine with high transconductance to result in a gain-bandwidth product that is several orders of magnitude above that of other ion-based transistors. We used these transistors to acquire a wide range of electrophysiological signals, including in vivo recording of neural action potentials, and to create soft, biocompatible, long-term implantable neural processing units for the real-time detection of epileptic discharges. E-IGTs offer a safe, reliable and high-performance building block for chronically implanted bioelectronics, with a spatiotemporal resolution at the scale of individual neurons.Entities:
Mesh:
Year: 2020 PMID: 32203456 DOI: 10.1038/s41563-020-0638-3
Source DB: PubMed Journal: Nat Mater ISSN: 1476-1122 Impact factor: 43.841