Frederik A Hansen1, Henrik Jensen2, Stig Pedersen-Bjergaard1,2. 1. Department of Pharmacy, University of Oslo, P.O. Box 1068 Blindern, 0316 Oslo, Norway. 2. Department of Pharmacy, Faculty of Health and Medical Sciences, University of Copenhagen, Universitetsparken 2, 2100 Copenhagen, Denmark.
Abstract
In this paper, we report the first example of employing a sacrificial electrode in the acceptor solution during electromembrane extraction (EME). The electrode was based on a silver wire with a layer of silver chloride electroplated onto the surface. During EME, the electrode effectively inhibited electrolysis of water in the acceptor compartment, by accepting the charge transfer across the SLM, which enabled the application of 500 μA current without suffering gas formation or pH changes from electrolysis of water. The electroplating strategy was optimized with a design-of-experiments (DOE) methodology that provided optimal conditions of electroplating. With an optimized electrode, 1 cm of the electrode in contact with the acceptor solution inhibited electrolysis of water for approximately 30 min at 500 μA current (redox capacity). Further, the redox capacity of the electrode was found to increase through multiple uses. The advantage of the electrode was demonstrated by extracting polar analytes at high-current conditions in a standard EME system comprising 2-nitrophenyl octyl ether (NPOE) as SLM and 10 mM HCl as sample/acceptor solutions. Application of high current enabled significantly higher recoveries than could otherwise be obtained at 100 μA. Sacrificial electrodes were also tested in μ-EME and were found beneficial by eliminating detrimental bubble formation. Thus, the sacrificial electrodes improved the stability of μ-EME systems. The findings of this paper are important for development of stable and robust systems for EME operated at high voltage/current and for EME performed in narrow channels/tubing where bubble formation is critical.
In this paper, we report the first example of employing a sacrificial electrode in the acceptor solution during electromembrane extraction (EME). The electrode was based on a silver wire with a layer of silver chloride electroplated onto the surface. During EME, the electrode effectively inhibited electrolysis of water in the acceptor compartment, by accepting the charge transfer across the SLM, which enabled the application of 500 μA current without suffering gas formation or pH changes from electrolysis of water. The electroplating strategy was optimized with a design-of-experiments (DOE) methodology that provided optimal conditions of electroplating. With an optimized electrode, 1 cm of the electrode in contact with the acceptor solution inhibited electrolysis of water for approximately 30 min at 500 μA current (redox capacity). Further, the redox capacity of the electrode was found to increase through multiple uses. The advantage of the electrode was demonstrated by extracting polar analytes at high-current conditions in a standard EME system comprising 2-nitrophenyl octyl ether (NPOE) as SLM and 10 mM HCl as sample/acceptor solutions. Application of high current enabled significantly higher recoveries than could otherwise be obtained at 100 μA. Sacrificial electrodes were also tested in μ-EME and were found beneficial by eliminating detrimental bubble formation. Thus, the sacrificial electrodes improved the stability of μ-EME systems. The findings of this paper are important for development of stable and robust systems for EME operated at high voltage/current and for EME performed in narrow channels/tubing where bubble formation is critical.
For more
than two decades, miniaturization
of analytical techniques for sample preparation has been an active
area of research, due to several potential advantages over conventional
approaches such as reduced cost and sample consumption, and environmental
benefits of lower consumption of organic solvents. One such technique
is electromembrane extraction (EME), which falls within the liquid-phase
microextraction category. In EME, which was originally introduced
in 2006,[1] two aqueous compartments holding
sample and acceptor solutions, respectively, are separated by a water-immiscible
organic solvent. The latter is immobilized in the pores of a polymeric
membrane, to yield a supported liquid membrane (SLM) or is a free
liquid membrane (FLM) squeezed between the aqueous compartments in
a narrow channel. The pH-value of the aqueous solutions is adjusted
to facilitate ionization of the analytes of interest. By applying
an electric field across the membrane, the charged analytes are transported
by electrokinetic migration into the acceptor solution. The transport
of ions, both analyte and background ions, across the membrane result
in generation of current, the magnitude of which is dependent on the
applied voltage and the properties of the membrane. These properties
may be altered in several ways by, for example, addition of ionic
carriers[2−4] or employing different membrane solvents.[5−8] Such alterations may be necessary for successful extraction of polar
ions, as they do not naturally partition into a hydrophobic membrane.[9,10]However, increasing the permeability of the membrane toward
polar
ions will generally increase the current. The current results in electrolysis
of water, as water becomes both donor and acceptor of electrons. Electrolysis
results in gas formation and pH changes in the sample and acceptor
compartments. Both of these effects may be devastating to a successful
extraction, and consequently it is a general rule-of-thumb to keep
the current below 50 to 100 μA.[11] Several reports have described how the extraction recovery typically
increases with extraction time, until a certain point where it starts
to decrease,[1,11−15] due to changes in the acceptor solution’s
pH-value. This causes loss of analyte ionization and may result in
back-extraction by passive diffusion. Even minor changes in the pH
of the acceptor solution may have a negative impact on the extraction
efficiency. Previous reports have described the formation of a boundary
layer at the membrane/acceptor interface[16−21] during EME, where pH is higher than in bulk acceptor solution. As
Restan et al.[21] described, this means in
practice that the acceptor solution’s pH should be at least
3–4 units lower than the pKa-value
of the analyte of interest for basic analytes, in order not to affect
the extraction yield negatively. pH may be stabilized by adding a
high concentration of pH modifier[22] (e.g.,
hydrochloric or formic acid) or by using a buffer.[23] This strategy is, however, like treating the symptoms rather
than the underlying cause, and even this approach may not be sufficient
at high-current conditions, particularly not in μ-EME where
the volumes are very small.[11] Furthermore,
extreme pH is not always an option due to analyte instability or the
simple fact that increased concentration of background electrolyte
also increases the extraction current.In the present contribution,
we introduce for the first time the
concept of sacrificial electrodes to the acceptor solution, to enable
stable and robust EME under high-current EME conditions. Sacrificial
metals/electrodes are frequent outside analytical chemistry to prevent
oxidation of, for example, iron; however, utilization of the concept
in analytical chemistry for sample preparation is very scarce. Mamat
and See[24] reported the use a bubbleless
electrode in EME, and although this was not a sacrificial electrode,
it solved many of the problems of electrolysis. The electrode was
essentially a salt-bridge that allowed the electrolytic processes
to be moved away from the extraction compartments. The sacrificial
electrode presented in the present work consisted of a silver wire
with a sacrificial silver chloride (AgCl) layer deposited on the surface.
This eliminated electrolysis of water entirely in the acceptor solution,
by accepting electrons from the current flow across the SLM instead
of water. The purpose of the current paper was to investigate the
feasibility of sacrificial electrodes to eliminate electrolysis of
water in the acceptor compartment. In addition, the fabrication of
the AgCl electrode was optimized to gain the highest capacity possible
for inhibition of electrolysis, and to investigate the reusability.
This is the first report on EME with sacrificial electrodes, and we
foresee great potential in the concept for future high-current applications
and for EME in microfluidic systems.
Experimental Section
Chemicals
and Reagents
Milli-Q (MQ) water was prepared
by a purification system (Molsheim, France). 2-Nitrophenyl octyl ether
(NPOE), bis(2-ethylhexyl) phosphate (DEHP), hydrochloric acid 37%
(HCl), formic acid, ammonium acetate, trichloroacetic acid (TCA),
potassium chloride, phenolphthalein, papaverine hydrochloride, mianserine
hydrochloride, prochlorperazine dimaleate, haloperidol, pethdine hydrochloride,
cocaine hydrochloride, methadone hydrochloride, loperamide hydrochloride,
nortriptyline hydrochloride, thiamine hydrochloride, pyridoxine hydrochloride,
norepinephrine bitartrate, epinephrine hydrochloride, and metformin
hydrochloride were all obtained from Sigma-Aldrich (St. Louis, MO).
Acetonitrile and methanol were purchased from Merck (Darmstadt, Germany).
Plasma sample were obtained from Oslo University Hospital (Oslo, Norway)
and stored at −32°C.
Preparation of Solutions
Stock solutions of each analyte
were prepared at a concentration of 1–3 mg/mL in methanol and
stored at −32°C for analytes with a positive log P value, while analytes with a negative log P value were dissolved in 30% methanol/MQwater and stored at 4 °C.
Standard solutions were prepared in 10 mM hydrochloric acid by adequate
dilution from stock solutions.
Preparation of AgCl Electrode
A piece of silver wire
(∼8 cm, ⌀ 0.5 mm, 99.9%, K.A. Rasmussen, Hamar, Norway)
was cut from a roll, and the surface was activated by a light polishing
with a fine-grade sand paper. The wire was subsequently cleaned with
a wipe soaked with 96% ethanol, after which approximately 5 cm of
it was shaped into a coil. After the cleaning step, the electrode
was only touched with gloves to avoid leaving fatty fingerprints.
Five centimeters of the coiled wire was submerged into a 100 mL solution
of potassium chloride (varying concentration), along with a platinum
wire. The silver wire was then connected to the positive terminal
(anode) of a power supply (model ES 0300e0.45, Delta Elektronika BV,
Zierikzee, Netherlands), while the platinum wire (⌀ 0.5 mm,
99.9%, K.A. Rasmussen, Hamar, Norway) was connected to the negative
terminal (cathode). A certain voltage/current was then applied for
a given duration of time to electroplate a layer of AgCl onto the
surface of the silver wire. After end plating, the AgCl electrode was flushed under Milli-Q water to
remove any residual potassium chloride.
Test of AgCl Electrode’s
Redox Capacity
To evaluate
the redox capacity of the AgCl electrode, the plated part of the electrode
was submerged fully into a 100 mL solution consisting of 1 M potassium
chloride for conductivity, and phenolphthalein as an indicator of
rising pH. The AgCl electrode was connected to the negative terminal
(cathode), while a platinum wire was used as the positive terminal
(anode). A constant current of 2 mA cm–1 electrode
was applied, and the AgCl electrode was monitored closely. As long
as the electrode had capacity for inhibiting reduction of water, there
was no visible change in the surroundings of the electrode. At the
point where no redox capacity was left, a purple zone formed rapidly
around the electrode due to increasing pH, and a few seconds after,
a significant amount of gas bubbles was also observed. The time point
of the onset of the purple zone’s formation was recorded as
the loss of redox capacity.
EME Procedure
Extractions were performed
in a 96-well
system, comprising a laboratory-built stainless steel plate with wells,
and a commercially available 96-well MultiScreen-IP filter plate with
polyvinylidene fluoride (PVDF) filter membranes with 0.45 μm
pore size (Merck Millipore Ltd., Carrigtwohill, Ireland). Prior to
use, 3 μL of SLM solvent (typically NPOE) was immobilized into
the pores of the membrane. A 100 μL amount of sample solution
was loaded into a well of the steel plate, while 100 μL of acceptor
solution was loaded into the MultiScreen filter well. The filter plate
and steel plate were then clamped together, and the coiled electrode
was inserted into the acceptor solution though a rubber stopper. See Supporting Information 1 for pictures. A model
ES 0300e0.45 (Delta Elektronika BV, Zierikzee, Netherlands) was used
as power supply, by connecting the anode to the steel plate holding
the sample, and the cathode to the electrode in contact with the acceptor
solution. A Vibramax 100 Heidolph shaking board (Kellheim, Germany)
was used to agitate the extraction system at 900 rpm. Extractions
were performed for 15 min at a constant current.
μ-EME
Procedure
The basic principle and instrumentation
of μ-EME has previously been described in detail,[25] and we refer to this for an elaborate description.
Briefly, 2 cm of PFA tubing (1/16 in. × 0.75 mm ID, Vici-Jour,
Schenkon, Switzerland) was cut, and 0.5 μL of 1-pentanol was
sandwiched between two 1.5 μL segments of aqueous solution (anolyte
and catholyte) in the tubing. Electrodes (0.25 mm in diameter) were
inserted from either end of the tubing into the aqueous solution and
connected to a power supply (model ES 0300e0.45, Delta Elektronika
BV, Zierikzee, Netherlands). The tubing was placed under a USB-microscope
(QS.20200-P, Q-scope, Euromex Microscopen bv, Arnhem, The Netherlands),
and photos were taken using associated software (Q-focus, v. 1.2.1.2,
Euromex Microscopen bv, Arnhem, The Netherlands). The extraction current
was recorded with a Fluke 287 multimeter (Everett, WA) at an acquisition
rate of 8 Hz.
pH Measurements
For pH measurements
of the sample and
acceptor solution, the entirety of the volumes was transferred to
a 0.5 mL Eppendorf tube for each, into which a pH microelectrode (Biotrode
6.0224.100, Metrohm, Switzerland) was inserted and the pH-value recorded.
Protein-Precipitation Procedure
Protein-precipitation
of plasma was performed by addition of 100% w/v trichloroacetic acid
(TCA) to a plasma sample at a 0.075:1 volume ratio. The mixture was
then vortexed at 1800 rpm for 3 min, before being centrifuged at 10 000
rpm for 5 min. The supernatant was then collected and diluted 10-fold
with Milli-Q water to produce a final pH-value of approximately 1.5.
UHPLC Conditions
The postextraction analysis of the
acceptor solutions was performed by a reversed-phase (RP-UV) method
for nonpolar analytes (positive log P value), and
a hydrophilic interaction liquid chromatographic (HILIC-UV) method
for polar analytes (negative log P value). See Table for list of analytes.
Both methods were run on a Dionex UltiMate 3000 RS UHPLC system comprising
a pump, autosampler, column compartment, and UV-detector.
Table 1
List of Model Analytes with pKa and log P values calculated
by MarvinSketch 18.22 (ChemAxon, 2018)
compound
pKa, pKa2
log P
Nonpolar Analytes
papaverine
6.03
3.08
mianserine
6.92
3.83
prochlorperazine
7.99, 2.20
4.38
haloperidol
8.05
3.66
pethidine
8.16
2.46
cocaine
8.85
2.28
methadone
9.12
5.01
loperamide
9.41
4.77
nortriptyline
10.47
4.43
Polar Analytes
thiamine
5.54
–3.10
pyridoxine
5.58
–0.95
norepinephrine
8.85
–0.68
epinephrine
8.91
–0.43
metformin
12.33
–0.92
For the reversed-phase method, an Acquity UPLC HSS
T3 column (100
mm × 2.1 mm I.D, 1.8 μm particle size) from Waters (Wexford,
Ireland) maintained at 50 °C was used. The analysis was performed
as a gradient elution, with mobile phase A consisting of 0.1% formic
acid in 95:5 v/v MQ:ACN, and mobile phase B consisting of 0.1% formic
acid in 5:95 v/v MQ:ACN. The gradient, operated at a flow of 0.5 mL/min,
consisted of a linear ramp from 0 to 40% B for 0 to 15 min, where
it was maintained for 2 min, before mobile phase B was increased to
80% within 0.1 min. Here it was kept for 2 min before being returned
to 0% for a final 6 min re-equilibration for a total run time of 25
min. Injection volume was 5 μL, and UV-detection was performed
at 210, 230, and 254 nm.For the HILIC-UV method, an Acquity
UPLC BEH amide column (150
mm × 2.1 mm I.D, 1.7 μm particle size) from Waters (Wexford,
Ireland) maintained at 30 °C was used. The analysis was performed
as a gradient elution, with mobile phase A consisting of 10 mM ammonium
acetate in 80:20 v/v MQ:ACN, and mobile phase B consisting of 10 mM
ammonium acetate in 90:10 v/v ACN:MQ. The gradient, operated at a
flow of 0.5 mL/min, consisted of 2 min at 100% B, after which a linear
ramp to 50% B during 5 min was made. After this, the %B was returned
to 100% within 0.5 min, where it was maintained for 8.5 min for a
total run time of 16 min. Injection volume was 5 μL, and UV-detection
was performed at 254 and 280 nm. Prior to injection, the aqueous acceptor
solution was diluted 10-fold in acetonitrile to match the initial
composition of the mobile phase.
Calculations
The
extraction recovery (%R) was calculated by the following
equation for each analyte:where na,final and ns,initial are
the number of moles
of analyte finally collected in the acceptor solution and the number
of moles of analyte originally present in the sample, respectively. Ca,final is the final concentration of analyte
in the acceptor solution, Cs,initial is
the initial analyte concentration within the sample, Va is the acceptor volume, and Vs is the sample volume.
Results and Discussion
Proof-of-Concept
To test the potential for AgCl-electrodes
to limit effects of electrolysis of water during EME, a model system
was used comprising NPOE as SLM and 100 μL of 10 mM HCl as both
sample and acceptor solution. The cathode was placed in the acceptor
compartment for extraction of cationic analytes. This was chosen,
as it represents one of the most commonly used systems for EME[2,12,26−30] and therefore is a good benchmark to test the concept
of sacrificial electrodes against. In a normal system with platinum
(Pt) electrodes, the amount of electrolysis is directly dependent
on the current. Therefore, to control the amount of electrolysis,
EME was performed in constant current mode, rather than the conventional
approach of applying a constant voltage. In principle, the 96-well
format used in the present work can generate several milliamperes
of current if the maximum voltage (300 V) is applied. This is due
to a relatively thin SLM (∼100 μm) and a high degree
of convection. However, for proof-of-concept, 500 μA for 15
min was used, as this is the highest current previously reported in
any EME system (to the best of our knowledge), though a current this
high was reported to produce somewhat unstable extraction systems.[9,31] Higher currents in another system of up to 2600 μA have also
been reported, but this resulted in a breakdown of the extraction
system within 2 min.[32] The power supply
enabled a constant current of 500 μA (±20–30 μA),
with voltages typically ranging from 100 to 250 V. The effects of
electrolysis of water were evaluated by the change in pH of the acceptor
solution (ΔpH) during the extraction, from the measured starting
pH-value of 2.04 for 10 mM HCl. Besides the AgCl electrode, platinum
(Pt) and silver (Ag) electrodes were tested for comparison as cathodes
without any redox capacity. Extractions at 100 μA (75–150
V) for 15 min were also performed to compare system performance at
two different current levels. All electrodes were prepared from 5
cm of 0.5 mm thick wire shaped into a coil that allowed the wire to
be inserted into the acceptor well-plate. For proof-of-concept studies,
the AgCl electrode was prepared by electroplating the coiled silver
wire in 1 M KCl, by connecting it to the anode of a power supply and
with an inert platinum wire as the cathode. Two volts was applied
until a stable current was reached after approximately 10 min.As seen from Figure , pH in the acceptor solution increased 5–6 units with 500
μA for both Ag and Pt electrodes due to electrolysis, while
only a minor increase was observed with 100 μA. The increase
in pH was accompanied by substantial formation of gas bubbles in the
acceptor and on the electrode. During the extraction, H+ ions were reduced to H2(g) in the acceptor at a constant
rate depending on the current, and pH increased with time like a titration
curve. Contrary to Pt and Ag, with the AgCl electrode, pH decreased
very slightly (for both 100 μA and 500 μA) and no gas
bubbles were observed. This demonstrated that the AgCl electrode inhibited
electrolysis of water in the acceptor compartment. We emphasize that
the sacrificial electrode was only used in the acceptor compartment,
and that electrolysis took place as normally in the sample. As seen
from eq , electrolysis
should result in acidification of the sample solution. Figure however shows the opposite
trend. The small pH decrease in acceptor and the increase in sample
was explained by comigration of H+ and OH– ions across the SLM, from sample and acceptor, respectively. Though
a constant pH is generally desirable, the minor acidification of the
acceptor favored extraction of basic substances. Likewise, basic substances
in the sample are not sensitive to minor pH changes in the bulk of
the solution due to formation of a sample/SLM boundary rich in H+ ions.[21]
Figure 1
pH change (ΔpH)
in the acceptor solution and sample solution
after 15 min of extraction at 100 μA (n = 3)
and 500 μA (n = 6) with Pt, Ag, and AgCl electrodes.
Error bars represent standard deviation.
pH change (ΔpH)
in the acceptor solution and sample solution
after 15 min of extraction at 100 μA (n = 3)
and 500 μA (n = 6) with Pt, Ag, and AgCl electrodes.
Error bars represent standard deviation.A similar set of experiments was conducted with 10 and 50 mM phosphate
buffer (Figure S3). With a Pt electrode
(no redox capacity), the pH of the 50 mM acceptor increased by 0.45
± 0.035 units from initially 1.98 after 15 min EME at 500 μA.
With the Pt electrode, gas bubbles were observed. With the AgCl electrode
and 50 mM phosphate, pH decreased by 0.28 ± 0.022 under similar
conditions. The latter was comparable with 10 mM HCl, and this indicated
that electrode redox capacity was more important than acceptor buffer
capacity in this case. The pH of a sufficiently buffered solution
was as such slightly more stable with a sacrificial electrode. In
practice, the difference is not expected to cause any change in extraction
efficiency. For some applications, good buffer capacity may thus be
sufficient to stabilize the pH with a platinum electrode. However,
gas formation would still pose a problem, as this previously has been
reported to produce an overpressure in the compartment, leading to
loss of the solution.[9] Sacrificial electrodes
can eliminate this problem, in addition to stabilizing the pH-value.
Theory of Sacrificial AgCl Electrodes
In a normal EME
system, electrolysis of water proceeds by the following equations,
where the standard reduction potentials for the half-cell reactions[33] also are indicated for the anode (eq 1) and cathode (eq 2).The reduction reaction for AgCl at
the cathode can be written:During extractions where
the AgCl electrode
is employed, eqs and 3 are essentially competing with each other. However,
due to the higher standard reduction potential of AgCl, the reduction
of this will be thermodynamically favorable over that of H+, and thus the AgCl layer on the electrode has to be consumed before
electrolysis of water can take place.The reduction of AgCl
is directly proportional to the transfer
of charges across the SLM, as expressed by eq :where nAgCl used is the moles of AgCl that have been reduced, t is
time measured in seconds, Q is charges measured in
Coulomb [C], F is Faraday’s constant (96485
C mol–1), and I is the current
[A or C s–1]. Equation assumes a constant current throughout the extraction,
but in experiments with varying current the total transfer of charges
(Q) can be calculated by eq .From eq , it is apparent that chloride
ions are generated
in the acceptor solution during extraction. Because these ions are
not expected to move into the SLM, an accumulation in the acceptor
solution is expected. From eq , the rate of formation can be calculated to 0.62 μmol
mA–1 min–1. In our model system,
comprising 500 μA for 15 min with a 100 μL acceptor solution,
this corresponds to 46 mM Cl– formed after 15 min.
We do not expect this to affect the extraction, as hydrochloric acid
up to 100 mM routinely is used with successful extraction. The formation
of chloride may be of concern for subsequent analysis, such as capillary
electrophoresis, where the conductivity of the solution may have an
impact. For HPLC analysis, formation of chloride is of no concern.In EME setups with conventional electrodes (e.g., platinum), eq may also be used to estimate
the theoretical pH change of the acceptor solution for a given time,
current, and acceptor volume. Figure S4 shows the theoretical pH curve for the extraction conditions given
in Proof-of-Concept above. This curve shows
the rapid increase in pH to happen after approximately 200 s. In practice,
it will however take longer time, as the observed extraction current
not only is a product of the electrolytic process but also the transfer
of all charged species across the SLM, including the H+ and OH– ions formed,[11] and other electrolytic processes.For the sacrificial AgCl
electrode, the capacity for inhibiting
electrolysis of water is naturally dependent on the electrode’s
surface area, and hence wire length, as well as current and time.
The redox capacity is therefore here characterized by the unit, mA·min·cm–1.
Design and Preparation of Sacrificial AgCl
electrode
To gain the highest redox capacity possible, the
surface area of
the electrode in contact with the acceptor solution should be large.
We therefore decided to shape the electrode into a coil. For practical
reasons we made the coil of 5 cm wire. Preparation of the AgCl layer
was performed by electroplating the silver wire in a potassium chloride
solution, with the silver wire as anode and a platinum wire as cathode.
When current was passed through the system, eq was essentially reversed and Ag(s) was oxidized
to Ag+, which in a chloride-rich environment formed an
insoluble deposit of AgCl(s) on the electrode surface.To assess
the importance of operational parameters of the electroplating process,
a design of experiments (DOE) methodology was utilized for optimization
of redox capacity. The design was a central composite orthogonal (CCO)
design, which included 20 experiments distributed across factorial
point, star points, and 6 center points. The factors chosen for optimization
were the potassium chloride (KCl) concentration of the plating solution
(X1), time (X2), and current (X3), while the redox capacity of the
electrode was used as the response. These factors were chosen for
optimization because they were the primary parameters that affected
the redox capacity. The levels of time and current were decided based
on initial experiments. Here a silver wire was plated with AgCl at
2 mA cm–1 and at a voltage no higher than 3 V, for
15 min in 1 M KCl. The limits in terms of voltage and current were
set to avoid oxygen formation during the electrode plating. Figure shows a typical
current profile during this process. Initially, 2 mA cm–1 could be applied with a voltage of less than 1 V. Within a few minutes,
the voltage increased to the limit of 3 V, resulting in a decreasing
current that ultimately formed a steady state which signified completed
plating. Based on these initial experiments, the factor levels given
in Table S1 were chosen and the experiments
were performed. The resulting model provided a good fit of the data
with a high coefficient of determination (R2 = 0.975). All three main factors, as well as a square term of the
current and interaction term of time and current (Table S2), were significant (p < 0.05).
Further details about the quality of the model are given in Supporting Information 3.
Figure 2
Typical current profile
during electroplating of silver wire at
2 mA cm–1 or maximum 3 V for 15 min in 1 M KCl.
Typical current profile
during electroplating of silver wire at
2 mA cm–1 or maximum 3 V for 15 min in 1 M KCl.Figure visualizes
the effect of the three factors by contour plots. As expected, increasing
plating current and time both resulted in increased redox capacity.
Interestingly, however, higher concentration of KCl in the plating
solution had a small negative impact on the redox capacity, at least
within the concentration range evaluated in this design. A possible
explanation for this may be that during the oxidation of the silver
wire, Cl– ions can also be oxidized to Cl2(g), which at a higher concentration creates more competition in
the oxidation process. At the optimal level of the three factors,
the estimated redox capacity was 14.5 mA·min·cm–1.
Figure 3
Contour plot of the electrode’s redox capacity depending
on KCl concentration, plating time, and plating current.
Contour plot of the electrode’s redox capacity depending
on KCl concentration, plating time, and plating current.Following completed plating of the electrode, the coated
surface
gained a strong purple-blackish color, as seen in Figure A,C. Mechanically, the coating
was relatively robust and was resistant to the mechanical forces due
to handling. However, when bending the wire to the preferred shape
of the electrode, parts of the coating could brake off, and therefore
the wire was shaped prior to electroplating. When the AgCl layer was
removed completely, the electrode had a greyish appearance as seen
in Figure B.
Figure 4
(A) Photo of
the coiled electrode after completed electroplating
of the AgCl layer. (B) Photo of the coiled electrode after completed
electrochemical removal of the AgCl layer. (C) Micrograph of the coated
part (left half) and the uncoated part (right half) of the electrode.
(A) Photo of
the coiled electrode after completed electroplating
of the AgCl layer. (B) Photo of the coiled electrode after completed
electrochemical removal of the AgCl layer. (C) Micrograph of the coated
part (left half) and the uncoated part (right half) of the electrode.
Regeneration of Sacrificial Electrode
The potential
for regeneration of electrodes was evaluated by 10 cycles of electroplating
and redox capacity testing. Testing the redox capacity implied a full
removal of the AgCl coating. For further insight into the stability
through multiple uses, we weighed electrodes before and after electroplating
and after removal of AgCl coating. Figure shows the development in redox capacity.
The first test yielded 14.6 mA·min·cm–1, which was almost exactly what the DOE model predicted. However,
through the next few cycles, the redox capacity increased considerably,
and by cycle 10 the capacity approached 30 mA·min·cm–1. Thus, the redox capacity of the AgCl electrodes
could easily be regenerated, and we improved capacity after multiple
uses. The increase in weight during electroplating correlated (R2 = 0.9849) with the experimental redox capacity
as illustrated in Figure A and corresponded to the theoretical chloride uptake as calculated
from eq (blue line).
During removal of the AgCl coating according to eq , silver (Ag(s)) formed by reduction can (1)
transfer into solution as colloidal silver, or (2) redeposit on the
surface of the electrode. Figure B shows the loss of weight for the electrode during
removal of the AgCl coating, with linear correlation (R2 = 0.9779) between weight loss and redox capacity. The
plot also includes the estimated weight loss for Cl– only (red) and for AgCl (blue). As seen, the observed weight loss
gives a strong indication that only chloride was lost during the removal
of the AgCl layer, thereby leading to the conclusion that Ag(s) was
redeposited on the surface of the electrode. This may explain the
increase in redox capacity through multiple regenerations, as Ag redeposited
with increasing porosity on the electrode surface.
Figure 5
Development in redox
capacity through multiple use cycles of an
electrode. One use cycle consisted of electroplating the electrode
followed by full removal of the AgCl to test the redox capacity.
Figure 6
Weight change of electrodes with different redox capacities.
(A)
Increase in weight during electroplating. (B) Decrease in weight during
removal of the AgCl coating. The stippled lines indicate the theoretical
decrease in weight if one assumes only Cl– is lost
(red), and if also the Ag atom is lost (blue). The data points are
a compilation of experiments from DOE and evaluation of reuse of electrodes.
Development in redox
capacity through multiple use cycles of an
electrode. One use cycle consisted of electroplating the electrode
followed by full removal of the AgCl to test the redox capacity.Weight change of electrodes with different redox capacities.
(A)
Increase in weight during electroplating. (B) Decrease in weight during
removal of the AgCl coating. The stippled lines indicate the theoretical
decrease in weight if one assumes only Cl– is lost
(red), and if also the Ag atom is lost (blue). The data points are
a compilation of experiments from DOE and evaluation of reuse of electrodes.
Analyte Extraction at High-Current Conditions
with Sacrificial
Electrodes
The concept of sacrificial AgCl electrodes was
then tested for real extractions under high current conditions. First,
high current EME was tested for extraction of five polar bases selected
as model analytes (Table ). As discussed previously, polar analytes are difficult to
extract due to their poor partitioning into the organic phase. To
solve this, increased extraction voltage/current or addition of an
ionic carrier such as DEHP to the SLM is required. Figure shows the extraction recoveries
obtained for both strategies individually and combined. Extractions
were performed at 100 or 500 μA with 10 mM HCl as sample and
acceptor phase, and AgCl-coated electrodes in the acceptor phase.
At 100 μA and with NPOE as SLM, all analytes were extracted
with recoveries of less than 11%. Addition of DEHP to the SLM improved
this considerably for pyridoxine, metformin, and thiamine, while a
minor improvement was observed for epinephrine and norepinephrine.
This observation is consistent with previous reports.[10] Extractions at 500 μA with NPOE, enabled by the AgCl
electrode, were likewise considerably more efficient for pyridoxine,
metformin, and thiamine than at 100 μA. The improvement was
less for epinephrine and norepinephrine. This highlights that increasing
the voltage/current is no guarantee of improving the extraction efficiency.
With the DEHP/NPOE SLM, extraction efficiency also improved at high
current. Despite the high current, no evidence of electrochemical
degradation of analytes was found using the sacrificial electrode.
Comparative extractions were performed at 500 μA with platinum
electrodes. With these, pH in the acceptor increased from 2 to 8 during
extraction, serious bubble formation occurred, partial or total loss
of acceptor was observed, and the SLM was often damaged (Figure S8). Quantitation of the acceptor solution
could therefore not be performed. The observed difference in system
stability with/without sacrificial electrodes and the improved extraction
efficiency of polar basic compounds at high-current conditions thus
clearly demonstrated the benefits of performing EME with sacrificial
electrodes.
Figure 7
Extraction recoveries from 50 μg mL–1 of
five polar bases in 10 mM HCl. The extractions were all performed
with a AgCl electrode for the acceptor compartment at 100 μA
or 500 μA and with either NPOE or 10% w/w DEHP/NPOE as the SLM,
for 15 min. Error bars represent the standard deviation (n = 6).
Extraction recoveries from 50 μg mL–1 of
five polar bases in 10 mM HCl. The extractions were all performed
with a AgCl electrode for the acceptor compartment at 100 μA
or 500 μA and with either NPOE or 10% w/w DEHP/NPOE as the SLM,
for 15 min. Error bars represent the standard deviation (n = 6).Experiments were complemented
with extraction of nine nonpolar
bases (Table ). The
results are shown in Figure S9. As expected,
these compounds were extracted equally efficient at 100 μA with
platinum electrodes and 500 μA with sacrificial electrode. This
was due to the nonpolar nature of the compounds, which enabled them
to pass through the SLM under low-current conditions. We therefore
emphasize that sacrificial electrodes are mainly beneficial for analytes
extracted under high-current conditions. Further, to demonstrate compatibility
of AgCl electrodes with biological matrixes, extractions were also
performed from spiked protein-precipitated plasma samples (Figure S9). These data were in agreement with
existing literature.[12,34,35] The sacrificial electrodes were thus compatible with a complex biological
matrix. Between extractions, the sacrificial electrode was cleaned
with ethanol to avoid carryover, similar to the typical procedure
for conventional electrodes.
Application in Microelectromembrane Extraction
Kubáň
and Boček performed microelectromembrane extraction (μ-EME)
for the first time in 2014.[36] μ-EME
differs from conventional EME by using a free liquid membrane (FLM)
rather than a conventional SLM, where the organic solvent is immobilized
in the pores of a supporting membrane. μ-EME is conducted in
small-diameter tubing or capillary, and the volumes of solvents are
just a few microliters. Due to the smaller volumes, the general tolerance
to high current is therefore less in these systems. This does not
only apply to pH changes but particularly also bubble formation that
in the narrow tubing may disrupt the electric circuit. Application
of the sacrificial electrode was therefore evaluated for μ-EME
in a final set of experiments. The experimental conditions used were
based on previous work.[11] As seen in Figure , when using a normal
silver electrode without redox capacity, bubble formation (H2) due to electrolysis at the cathode was clearly observed. After
2 min operation with tens of microampere (equal to 5.0 mC), the size
of the bubble disrupted the electric circuit and extraction was no
longer possible. However, with the sacrificial electrode, no bubble
formation occurred. Even after testing the system equivalent to 25
mC, no bubbles formed at the cathode. This experiment clearly demonstrated
the benefit of sacrificial electrodes in μ-EME. Interestingly,
no bubbles were observed at the silver wire anode. Silver was oxidized
under the selected μ-EME conditions and therefore essentially
also functioned as a sacrificial anode. Stable pH conditions with
sacrificial electrodes in μ-EME were also demonstrated, as described
in Supporting Information 5.
Figure 8
(Top) Illustration
of μ-EME configuration. (Bottom) Photographs
showing bubble formation at the cathode with a normal silver electrode
and the complete lack of bubble formation with a sacrificial AgCl
electrode. Experimental conditions: anolyte and catholyte were both
1.5 μL of 0.1 M formic acid, and 0.5 μL of 1-pentanol
was used as FLM. Anode electrode was a 0.25 mm silver wire. Voltage:
100 V.
(Top) Illustration
of μ-EME configuration. (Bottom) Photographs
showing bubble formation at the cathode with a normal silver electrode
and the complete lack of bubble formation with a sacrificial AgCl
electrode. Experimental conditions: anolyte and catholyte were both
1.5 μL of 0.1 M formic acid, and 0.5 μL of 1-pentanol
was used as FLM. Anode electrode was a 0.25 mm silver wire. Voltage:
100 V.As a miniaturized EME system,
μ-EME is operated under stagnant
conditions. Other miniaturized systems, such as the chip-based, are
however often operated in continuous flow mode.[37−39] In this mode,
pH changes of solutions are expected to be minor as fresh solution
is continuously delivered. However, bubbles can still form on normal
electrodes. These bubbles are expected to grow until they break off
and disrupt the electric circuit. Depending on the flow rate, the
bubble is expected to be flushed out of the channels, and the disruption
is therefore temporary. Employing sacrificial electrodes may however
still be useful to ensure the best conditions for repeatable extractions.
Comparison with Other High-Current EME Systems
Table lists previous EME
systems operated at high current. Generally, current levels above
300 μA are reported to result in system instability when electrolysis
effects have not been eliminated. Besides the present work, only the
approach of Mamat and See[24] with a bubbleless
electrode have eliminated electrolysis in the extraction compartments.
In their approach, an in-capillary salt bridge was used to electrically
connect the extraction compartment to an external compartment where
the extraction voltage was applied. The electrolysis hence took place
in this external compartment. Because the distance from the actual
electrode to the SLM was longer, the authors applied voltages of up
to 3000 V, corresponding to 250 μA, to gain efficient extraction.
Table 2
Comparison of System Stability for
Different EME Systems Operating at High Current
EME format
current (μA)
inhibition
of electrolysis?
system instability
ref
96-well
500
yes
stable
this work
96-well
500
no
pH increase, loss of acceptor solution
this work
96-well
500
no
overpressure and volume loss in acceptor compartment
(9)
96-well
300
no
stable below 300 μA
(9)
hollow-fiber
250
yes
stable below 250 μA
(24)
hollow-fiber
300
no
pH increase and bubble formation
(31)
96-well
400
no
overpressure
and volume loss in acceptor compartment
(40)
flat-membrane
>100
no
decrease of recovery
(7)
Conclusion
The
concept of sacrificial electrodes in electromembrane extraction
was demonstrated for the first time. With such an electrode in the
acceptor solution, electrolysis was avoided, pH was stabilized, and
bubble formation was eliminated. The redox capacity of the electrode
was more than adequate to stabilize the system with 500 μA current,
which in a conventional system with 10 mM HCl and platinum electrodes
is not feasible. This allowed several fold increase in extraction
recovery of basic polar model analytes, compared to operation at lower
current. The specific sacrificial electrode (AgCl) presented here
only provided inhibition of electrolysis as the cathode. For acidic
analytes, the electrode can thus not inhibit electrolysis in the acceptor
compartment, because the anode must be placed here. The sacrificial
electrode was additionally tested in a μ-EME system, providing
considerably higher tolerance to high current, particularly related
to bubble formation. The latter may also be useful for chip-based
EME performed in narrow channels. We consider the present work to
be an important step expanding the EME toolbox. Particularly, for
extraction of analytes suffering from poor SLM partitioning, and for
applications such as μ-EME or chip-based EME vulnerable to gas
formation at the electrodes, sacrificial electrodes should be considered.
Extraction of multiply charged biomolecules, such as peptides, may
require a stable pH, and represents another area where sacrificial
electrodes may be favorable. We expect EME to be explored for biomedical
applications of increasing complexity in the future, and for such
applications, the use of buffered solutions combined with sacrificial
electrodes may play a key role.
Authors: Frøydis Sved Skottvoll; Frederik André Hansen; Sean Harrison; Ida Sneis Boger; Ago Mrsa; Magnus Saed Restan; Matthias Stein; Elsa Lundanes; Stig Pedersen-Bjergaard; Aleksandra Aizenshtadt; Stefan Krauss; Gareth Sullivan; Inger Lise Bogen; Steven Ray Wilson Journal: Anal Chem Date: 2021-02-03 Impact factor: 6.986