Thermoresponsive polymers reversibly react to changes in temperature and water content of their environment (i.e., relative humidity, RH). In the present contribution, the thermoresponsiveness of poly(N-vinylcaprolactam) thin films cross-linked by di(ethylene glycol) divinyl ether deposited by initiated chemical vapor deposition are investigated to assess their applicability to sensor and actuator setups. A lower critical solution temperature (LCST) is observed at around 16 °C in aqueous environment, associated with a dramatic change in film thickness (e.g., 200% increase at low temperatures) and refractive index, while only thermal expansion of the polymeric system is found, when ramping the temperature in dry atmosphere. In humid environment, we observed a significant response occurring in low RH (already below 5% RH), with the moisture swelling the thin film (up to 4%), but mainly replacing air in the polymeric structure up to ∼40% RH. Non-temperature-dependent swelling is observed up to 80% RH. Above that, thermoresponsive behavior is also demonstrated to be present in humid environment for the first time, whereas toward 100% RH, film thickness and index appear to approach the values obtained in water at the respective temperatures. The response times are similar in a large range of RH and are faster than the ones of the reference humidity sensor used (i.e., seconds). A sensor/actuator hygromorphic device was built by coating a thin flower-shaped poly(dimethylsiloxane) (PDMS) substrate with the thermoresponsive polymer. The large swelling due to water uptake upon exposure to humid environment at temperatures below the LCST caused the petals to bend, mimicking the capability of plants to respond to environmental stimuli via reversible mechanical motion.
Thermoresponsive polymers reversibly react to changes in temperature and water content of their environment (i.e., relative humidity, RH). In the present contribution, the thermoresponsiveness of poly(N-vinylcaprolactam) thin films cross-linked by di(ethylene glycol) divinyl ether deposited by initiated chemical vapor deposition are investigated to assess their applicability to sensor and actuator setups. A lower critical solution temperature (LCST) is observed at around 16 °C in aqueous environment, associated with a dramatic change in film thickness (e.g., 200% increase at low temperatures) and refractive index, while only thermal expansion of the polymeric system is found, when ramping the temperature in dry atmosphere. In humid environment, we observed a significant response occurring in low RH (already below 5% RH), with the moisture swelling the thin film (up to 4%), but mainly replacing air in the polymeric structure up to ∼40% RH. Non-temperature-dependent swelling is observed up to 80% RH. Above that, thermoresponsive behavior is also demonstrated to be present in humid environment for the first time, whereas toward 100% RH, film thickness and index appear to approach the values obtained in water at the respective temperatures. The response times are similar in a large range of RH and are faster than the ones of the reference humidity sensor used (i.e., seconds). A sensor/actuator hygromorphic device was built by coating a thin flower-shaped poly(dimethylsiloxane) (PDMS) substrate with the thermoresponsive polymer. The large swelling due to water uptake upon exposure to humid environment at temperatures below the LCST caused the petals to bend, mimicking the capability of plants to respond to environmental stimuli via reversible mechanical motion.
Stimuli-responsive
swelling is a property of smart hydrogel materials,
which makes them suitable for a plethora of sensor[1] and actuator[2] setups. A variety
of materials have been developed exhibiting a specific water uptake
behavior depending on external stimuli including temperature, pH,
magnetic/electric fields, or concentrations of specific chemical species
(e.g., glucose).[3] Such hydrogels, as networks
of hydrophilic polymer chains, can swell to up to a multiple of their
dry size; they react reversibly by taking up or repelling out water
upon changing the environment (i.e., the magnitude of the external
stimulus).This water exchange occurs diffusion-based and is,
hence, time-limited
by water diffusivity.[4] To achieve fast
response times and, as a result, optimal device performance, thin
polymeric films are employed. Furthermore, device setups often require
the coating of delicate surfaces (e.g., drugs,[5] flexible substrates), exhibiting sophisticated nanostructure. In
this contribution, initiated chemical vapor deposition (iCVD) was
adopted to meet these specific requirements. This solvent-free technique
enables the conformal thin film deposition of a plethora of chemical
species and compositions from the vapor-phase similar to free radical
polymerization processes.[6] The mild processing
conditions employed allow for full retention of delicate functional
groups, such as the adopted thermoresponsive units, upon deposition.
A variety of process parameters (e.g., flow rates, pressure, filament
temperature) can be utilized to tune the properties of the deposited
thin films in a large range of directions (e.g., composition, morphology,
molecular weight).During swelling, the resulting smart hydrogel
thin films undergo
rearrangements of polymer chains that impose stress on the thin films,
possibly eventuating in poor adhesion or mechanical failure. To increase
the thin films’ mechanical stability, a second monomer, working
as a cross-linker, can be copolymerized. As the cross-linking agent
binds two separate polymer chains, the cross-linked hydrogel can be
viewed as a polymer network (with a certain mesh size), able to take
up a specific maximum amount of water.[7]The responsiveness to water and to the other external stimuli
can
be tailored via the chemical nature of monomer and cross-linker. In
case of thermoresponsive hydrogels, e.g., the most prominent example
poly(N-isopropylacrylamide) (pNIPAAm) exhibits hydrophilic
groups (i.e., amide) that are able to bind water molecules via hydrogen-bonds;
antagonistically, the polymeric mesh shows the potential for attractive
intrachain interactions leading to polymer collapse at higher temperatures.[8] The combination of such specific properties result
in the material exhibiting a lower critical solution temperature (LCST).[8] In aqueous environment, below the LCST, the polymer
exhibits a hydrated swollen state and a transition to a dehydrated
shrunken state, when the temperature is increased. Recently, another
temperature-responsive polymer, namely poly(N-vinylcaprolactam)
(pNVCL), was synthesized by iCVD for the first time.[9] Its nontoxicity and biocompatibility make it promising
for biomedical applications.[10] Its specific
chemical structure facilitates a typical Flory–Huggins demixing
behavior with water, which is, for example, different from the miscibility
of pNIPAAm.[11] Our group provided further
evidence of such behavior by showcasing the tunability of the thermoresponsiveness
of pNVCL-based copolymer thin films via molecular
weight (i.e., chain length) by varying the filament temperature during
iCVD.[12] Mechanical stabilization and further
tailoring of the LCST were demonstrated by copolymerizing
a cross-linking agent, namely di(ethylene glycol) divinyl ether (DEGDVE),
together with pNVCL, as described elsewhere.[12] DEGDVE has been previously shown to allow for tuning of the temperature-responsiveness
in pNIPAAm- and pNVCL-based copolymers by our group.[12−14]In this contribution, besides studying the swelling behavior
of
the p(NVCL-co-DEGDVE) thin films in water in situ, we aim at shedding light onto the thermoresponsive
behavior of the polymer in humid environment. Kinetic swelling data
is collected to gain information about the applicability of these
systems as fast-responding thin films in sensor and actuator setups
in humid environment as well as when water-immersed. The swelling
behavior of simple polymeric systems in humidity has been investigated
within few studies previously. For instance, Secrist and Nolte measured
the response of single-component polyelectrolyte thin films in controlled
humid environment by reflectometry.[15] They
report on swelling curves of these thin films similar to results of
our group published in previous contributions.[14,16] However, the literature is lacking clear, systematic studies. Furthermore,
the swelling behavior of more complex systems like temperature-responsive
hydrogels is not very well documented. Thijs et al., for instance,
measured the weight change of two macroscopic temperature-responsive
hydrogel powders (pNIPAAm, poly((dimethylamino)ethyl methacrylate)—pDMAEMA),
while introducing water vapor into these systems up to 90% RH.[17] They report on the weight gain being a function
of temperature in the whole range of investigated humidity values
(up to 90% RH). Preliminary measurements on cross-linked pNIPAAm thin
films contradict this observation. For such thin films, our group
reported on the temperature-responsive transition not being present
as a change in film thickness in nitrogen environment, but being tunable
by cross-linking, when immersing the samples in liquid water.[14] In humid environment now, we employed cross-linked
pNVCL thin films due to their great tunability in thermoresponsiveness,
as demonstrated by us recently.[12] Instead
of weight gain, we aim at measuring the change in film thickness in situ by spectroscopic ellipsometry (SE), upon introducing
water vapor into the system, while it is held at certain temperatures
below and above the LCST. The sharp temperature-dependent transition
observed in pNVCL-based systems in water facilitates the described
experimental determination of the thermoresponsiveness in RH-dependent
swelling. Furthermore, it yields great potential for the applicability
of the investigated systems in device setups.To provide a proof-of-concept
sensor/actuator device based on the
fast and large swelling response and thermoresponsiveness of the pNVCL
thin films, we fabricated a bilayer hygromorphic system, inspired
by Taccola et al.[18] A pNVCL-based polymer
layer was deposited onto a thin polydimethylsiloxane (PDMS) substrate
and subsequently laser-cut into flower shapes. In such double-layered
structure, the PDMS acted as an elastic, mechanically passive layer,
not being swollen by humidity. The swelling of the pNVCL layer caused
a macroscopic, fast and reversible bending of the petals upon changes
in the environment (i.e., temperature, RH). While previous studies
focused on actuation of responsive hydrogels immersed in water,[19−21] we provide a first demonstration of macroscopic humidity-driven actuation with temperature-responsive tuning/switching.
Experimental Section
Thin Film Synthesis
Poly(N-vinylcaprolactam-co-di(ethylene
glycol) divinyl ether) thin films were synthesized
by initiated chemical vapor deposition. The depositions were run in
a custom-built iCVD reactor, as described elsewhere.[14]N-Vinylcaprolactam (NVCL, 98%; Aldrich,
Germany) is used as monomer and di(ethylene glycol) divinyl ether
(DEGDVE, 99%; Aldrich, Germany) as cross-linker. NVCL and DEGDVE are
kept in their individual glass jars at 78 and 70 °C, respectively.
The monomer and cross-linker flow rates are both set to 0.2 sccm.
Nitrogen is used as a carrier gas at a flow rate of 1.8 sccm. The
corresponding values of the ratio between the monomer partial pressure
and the saturation pressure at the substrate temperature (pM/psat) are 0.063
and 0.018, for monomer and cross-linker, respectively. In this range
of pM/psat values, a linear relation to the surface concentration of the chemical
species has been reported.[22] The applied
flow rates yield polymer layers of p(NVCL-co-DEGDVE)
with nominal cross-linking of 20%. To monitor the deposited thickness, in situ laser interferometry with a He–Ne laser (λ
= 633 nm; Thorlabs, USA) is performed through a removable quartz glass
lid. Several samples exhibiting film thickness values of (50 ±
5) nm have been synthesized and investigated in terms of their swelling
behavior in different environmental conditions. Directly after deposition,
the thin film samples were rinsed for 30 s with deionized water for
equilibration reasons, as reported earlier.[14,23]
Characterization and Investigation
Spectroscopic ellipsometry
(SE) in a wavelength range of 370–1000 nm (M-2000S, J. A. Woollam,
USA) was applied to determine film thickness and optical properties
of the thin films in the different environmental conditions. Swelling
experiments in deionized water were performed in a temperature controlled
liquid stage (J. A. Woollam, USA). The recorded data was evaluated
with an optical model consisting of a c-Si semi-infinite layer on
the bottom (temperature dependent), a 1.6 nm thick native SiO2 layer in the middle and the polymer film on top, modeled
as a Cauchy function with an Urbach tail accounting for adsorption
in the low wavelength region. H2O with temperature-dependent
optical properties was set as the surrounding medium. For the temperature-dependent
swelling experiments, the liquid stage and the mounted sample (already
exposed to deionized water) were precooled to 10 °C. The respective
signal was then recorded while applying a temperature ramp from 10
to 50 °C (also the limitations of the temperature stage) at a
heating rate of 0.5 °C/min. Similar heating rates have previously
been shown to allow the temperature-responsive material to equilibrate
during deswelling and hence yield an “equilibrium swelling
behavior”.[14] For thermal equilibration
reasons reported earlier,[14,23] the temperature-dependent
swelling behavior and the LCST of the iCVD thin films in water were
evaluated from the third heating experiment. All subsequent measurements
yielded similar curves and consistent results.The spectroscopic
ellipsometry measurements in controlled humid environment were performed
in a THMS600 temperature stage (Linkam, U.K.) at different substrate
temperatures (10–24 °C). Closing the lid constrains the
volume within the cell to about 70 cm3. A custom-built
setup has been used to provide controlled relative humidity (RH) with
mixing N2 bubbled through water (100% RH) and pure N2 via needle valves. The relative humidity has been monitored in situ inside the temperature stage (2 cm from the substrate)
with an SHT15 humidity sensor (Sensirion, Switzerland). To measure
the response of the film as a function of RH at a constant substrate
temperature, the humidity was increased from 0% RH up to the point
where condensation occurs on the sample surface. With the temperature
and RH value measured at the humidity sensor, the saturation vapor
pressure values (p) at the temperature measured by the humidity sensor (T) and at the substrate temperature
(T) were calculated
with the help of the Arden–Buck equation (eq ):[24]The
relative humidity at the substrate level (RH), that the sample is effectively exposed to, was then calculated
as the product of the relative humidity measured at the sensor (RH) and the ratio of the saturation vapor
pressure values at the sensor temperature and at the substrate temperature
(Equation ):The point of condensation
could be taken as a reference for the
validity of this calculation and could successfully be estimated by
±2% RH, which is within the uncertainty of the RH sensor used.
To evaluate the film thickness and refractive index of the sample,
the same optical model as in the liquid case has been used to fit
the recorded data, but with the ambient material being set to air
(n ≈ 1). Thickness and refractive index were
monitored at each humidity step and read out after several minutes
of equilibration, so that the observed film thickness would not vary
for more than 1% in 2 min. The transition temperature of the deposited
systems has been evaluated from the optical thin film properties recorded
by spectroscopic ellipsometry during swelling experiments in water
while increasing the temperature from 10 to 50 °C.Additionally,
kinetic measurements were performed in situ while
applying water or a certain amount of humidity onto the thin
film samples. The same optical models as in the temperature-dependent
measurements (liquid and humid environment, respectively) were employed
for fitting the data.
Fabrication and Testing of Hygromorphic Devices
To
fabricate hygromorphic actuator/sensor devices that respond to changes
in the environment (i.e., temperature, RH), thin polydimethylsiloxane
(PDMS) substrates were prepared. PDMS (10:1 ratio of base elastomer
to curing agent; Sylgard 184 silicone elastomer base and curing agent,
Dow Corning Corp.) was spin coated onto polystyrene (PS) disks for
60 s at a speed of 650 rpm and then cured at T =
80 °C for 3 h in an oven. The PDMS film had a thickness of tPDMS = (55 ± 4) μm, as determined
by stylus pofilometry (Alpha Step D-500 Profiler, KLA Tencor, USA).
PDMS substrates, still supported by PS, were then coated with 300
nm of the 20% cross-linked p(NVCL-co-DEGDVE) thin
film with the iCVD process as described above. After iCVD, the bilayer
system was laser-cut into a flower shape (see design and schematics
of bilayered actuators in Figure S1, Supporting
Information), with a direct-write CO2 laser (VLS
2.30, Universal Laser Systems, Inc., USA) equipped with a 30 W source.
Power, speed, and resolution were tuned to optimize sharp cutting
of PDMS with negligible damage to the underlying PS support.For investigating the response to RH, a flower-shaped device was
peeled off from the PS support and placed on a paper rod support,
with the thermoresponsive layer pointing upward. After placing it
in a cuboid (8 × 8 × 8 cm3) to constrain the
gas exchange with the environment, the different levels of humidity
were supplied at 25 °C by a gas stream flowing in from top, using
the same mixing setup as described above. To investigate the response
to temperature, the device was placed onto the temperature stage with
the thermoresponsive layer being in direct contact with the sample
table (i.e., pointing downward). The RH of the environment was measured
to be 35% at 25 °C.
Results and Discussion
Thermoresponsiveness in
Water and Dry Atmosphere
Effective
sensors are characterized by a fast and large amplitude response.
In a previous contribution, we demonstrated that the cross-linker
amount largely affects the amplitude of the response in thermoresponsive
films deposited by iCVD; together with employing distinct amounts
of cross-linking, utilizing the knob of the filament temperature makes
it possible to tune the LCST in the range 16–40 °C.[12] For the present study, we synthesized p(NVCL-co-DEGDVE) thin films by initiated chemical vapor deposition
with a minimal nominal amount of cross-linking of 20% to ensure the
films’ stability in water. We use a model system with an LCST
of about 16 °C (cf. Figure ). From ellipsometric measurements, the thickness and
refractive index of the thin films in water have been evaluated via
fitting the measurement data; they are plotted in Figure a as a function of temperature.
Figure 1
(a) Film
thickness normalized by the dry thickness (d/ddry) and refractive index (n) measured during swelling of a nominally 20% cross-linked
50 nm p(NVCL-co-DEGDVE) thin film in water while
heating from 10 to 50 °C, with dry thickness measured in N2 atmosphere at 10 °C and transition temperature between
swollen and collapsed state (as mean value of points of inflection
of thickness and index curves) indicated with a vertical line at 16.1
°C. (b) d/ddry and n measured during temperature ramps between 10 and 50 °C
(10 min ramp; 5 min wait at max/min T) of the thin
film in nitrogen atmosphere
(a) Film
thickness normalized by the dry thickness (d/ddry) and refractive index (n) measured during swelling of a nominally 20% cross-linked
50 nm p(NVCL-co-DEGDVE) thin film in water while
heating from 10 to 50 °C, with dry thickness measured in N2 atmosphere at 10 °C and transition temperature between
swollen and collapsed state (as mean value of points of inflection
of thickness and index curves) indicated with a vertical line at 16.1
°C. (b) d/ddry and n measured during temperature ramps between 10 and 50 °C
(10 min ramp; 5 min wait at max/min T) of the thin
film in nitrogen atmosphereFrom the measurement data represented in Figure a, the transition temperature could be determined
as the mean value of the points of inflection of the temperature-dependent
thickness and refractive index curves, as already described elsewhere.[14] This gives an LCST value of 16.1 °C. The
temperature-dependent swelling curves show an increase of the film
thickness compared to the dry one of more than 250% at temperatures
below the LCST. A strong transition between a swollen state with large
film thickness and low refractive index below the LCST, and a collapsed
state with smaller film thickness and higher refractive index above
the LCST is observed. At temperatures above the LCST, the film thickness
reaches values of ∼1.5 times the dry film thickness, hinting
toward a significant amount of water being retained in the polymeric
structure also above the transition temperature, as already reported
before.[12,14] Furthermore, at temperatures above ∼40
°C, the refractive index appears to decrease, while the film
thickness increases again. From 35 to 50 °C, the hydrogel film
thickness increases by ∼8%, while the refractive index decreases
by ∼0.012. Hypothetically, this can be attributed to thermal
expansion at temperatures above the LCST, at which the transition
is completely finished and the polymeric system appears in the collapsed
state. However, the increase in film thickness and the decrease in
refractive index seem unusually large.To investigate this effect
further, the thermal expansion of the
hydrogel thin films in a dry environment (i.e., N2 atmosphere)
has been investigated by ramping the temperature from 10 to 50 °C
and vice versa; ramping times of 10 min and wait
times of 5 min were used, while recording the optical response of
the material by spectroscopic ellipsometry. The film thickness normalized
by the thickness obtained at 10 °C and the refractive index gained
from the respective experiment are plotted in Figure b as a function of temperature. Interestingly,
the hydrogel thin film does not exhibit a temperature-responsive behavior
without the presence of water (in N2 atmosphere). Hence,
the (linear) thermal expansion of the polymeric material can be estimated.
For comparison, from 35 to 50 °C, the film thickness is observed
to increase by ∼0.5% and the refractive index is observed to
decrease by ∼0.005. At the respective temperatures (above the
LCST), the increase in film thickness in water appears to be larger
than the one observed in N2 atmosphere by 1 order of magnitude.
Hypothetically, the water retained in the polymeric system in the
collapsed state decreases the attractive interaction of polymeric
structures. This promotes the entire polymeric system to expand more
easily and would explain the higher thermal expansion in water.
Swelling in Humid Environment
Overall, the magnitude
of the observed swelling and temperature-dependent response in water
make the material promising for sensor and actuator systems. To investigate
the applicability of the synthesized hydrogel thin films also in humid
environment, we probed the thermoresponsive thin films in terms of
swelling in RH below and above the transition temperature of 16.1
°C by SE. For that, we monitored the film thickness and the refractive
index, while introducing controlled amounts of water vapor into the
nitrogen filled system. As described before, these experiments were
performed at various substrate temperatures between 10 and 24 °C,
measuring the response of the thin films at temperatures below and
above the LCST, respectively (Figure ). As shown later, the response of the thin films investigated
is in the order of seconds. To ensure measuring a reliable thickness
value, the values were read out after several minutes of equilibration,
so that the observed film thickness would not vary for more than 1%
in 2 min.
Figure 2
Film thickness normalized by the dry film thickness (d/ddry) of 50 nm p(NVCL-co-DEGDVE) thin films evaluated from spectroscopic ellipsometry measurements
at various substrate temperatures between 10 and 24 °C measured
in pure nitrogen environment and while introducing controlled amounts
of water vapor (i.e., relative humidity, RH) into a nitrogen-filled
system up to condensation; full symbols represent the described measurement
points; empty symbols (at 100% RH) are taken from measurements in
water (cf. Figure ); the inset shows a zoom-in of the region below 80% RH.
Film thickness normalized by the dry film thickness (d/ddry) of 50 nm p(NVCL-co-DEGDVE) thin films evaluated from spectroscopic ellipsometry measurements
at various substrate temperatures between 10 and 24 °C measured
in pure nitrogen environment and while introducing controlled amounts
of water vapor (i.e., relative humidity, RH) into a nitrogen-filled
system up to condensation; full symbols represent the described measurement
points; empty symbols (at 100% RH) are taken from measurements in
water (cf. Figure ); the inset shows a zoom-in of the region below 80% RH.The SE measurements of p(NVCL-co-DEGDVE)
thin
films in humid environment at different temperatures (see Figure ) show swelling of
the thin films already occurring at humidity levels of below 5% RH
for all the investigated substrate temperatures. The measurements
performed at different substrate temperatures yield similar swelling-vs-RH
curves up to a humidity level of approximately 80% RH (see inset of Figure ). Above 80% RH,
the thermoresponsiveness of the thin film samples, apparent when studied
immersed in water, can also be observed in humid environment. In this
region, the degree of swelling is a function of the substrate temperature
and approaches the values observed in water at humidity levels close
to 100% RH. Even though the uncertainty of fitting the optical measurement
data increases at humidity levels above 90% RH, the difference between
the recorded measurement points at the different temperatures is clear
and distinct.Secrist and Nolte’s measurements[15] of the response of non-thermoresponsive single-component
polyelectrolyte
thin films in controlled humid environment by reflectometry yield
comparable results to the swelling-vs-RH-curves presented in Figure . Thijs et al.[17] also report on the weight change of macroscopic
hydrogel powders, exposing non-thermoresponsive hydrogels (pHEMA)
to up to 90% RH; they observed similar swelling behavior as well.
However, for thermoresponsive hydrogels investigated in the mentioned
study,[17] only the trends of water uptake
reported for temperatures below the LCST appear to be comparable to
the ones reported in the present contribution. As described earlier,
the cited publication reports on the weight gain being a function
of temperature in the whole range of investigated humidity levels
(measured up to 90% RH). Counterintuitively, even a negative weight
change was reported for the measurements performed at elevated temperatures
(above the LCST) as a function of RH.The different results
either arise from the different sample morphologies
(macroscopic pNIPAAm powder vs cross-linked pNVCL-based thin film)
or stem from the different measurement setups. In either way, temperature
gradients are nearly unavoidable to measure thermoresponsive properties,
but can be problematic in studying and interpreting the response of
such a material in humid environment. In a more recent case, further
pNIPAAm-based hydrogels were investigated in terms of moisture uptake
from a humid environment.[25] Again, the
weight change was reported to be a function of temperature in the
whole range of investigated humidity levels; however, the resulting
swelling-curves are different from the ones reported by Thijs et al.[17] gained from similar measurements. The results
at temperatures below the LCST, again, appear to be comparable to
the trends reported in Figure . However, the data suggest that the systems have not been
investigated at temperatures above the LCST at the same humidity levels
as below the LCST. For example, the amount of adsorbed moisture reported
for a temperature of 20 °C at 30% RH compares well to the one
reported for 90% RH at 40 °C;[25] as
a matter of fact, generating a humidity of 90% RH at 20 °C translates
to a humidity of ∼30% at 40 °C (cf. eqs and 1 and Figure ). Figure should serve as an illustration
for the relative humidity changing with temperature.
Figure 3
Exemplary illustration
of the relative humidity (RH) as a function
of temperature (T), if an RH of 90% is generated
at 20 °C (and kept constant) and T is varied.
Condensation occurs below ∼18.3 °C, as indicated by a
dashed line and a bar at 100% RH; calculations according to eqs and 1.
Exemplary illustration
of the relative humidity (RH) as a function
of temperature (T), if an RH of 90% is generated
at 20 °C (and kept constant) and T is varied.
Condensation occurs below ∼18.3 °C, as indicated by a
dashed line and a bar at 100% RH; calculations according to eqs and 1.Overall, several studies report
on data at a specific relative
humidity, varying the temperature. However, they appear to disregard
that changing the temperature also changes the relative humidity drastically;
for example, a humidity of 90% RH generated at 20 °C increases
by ∼5.6% for an object that is exposed to this very humidity,
but is held at 1 °C lower (i.e., 19 °C) (cf. eqs and 1).
To avoid such problems in the experiments performed for the present
contribution, the humidity was varied while keeping the temperature
constant.To give further insight into the experimental setup,
the humidity
(up to 100% RH) is generated by bubbling nitrogen through water (at
70 °C), air filtering and mixing it with pure nitrogen to be
able to control the humidity level. This mixture is then flown into
the measurement chamber via a 1-m tube to avoid fluctuations in temperature.
In the chamber, a humidity sensor measures the RH approximately at
room temperature. Only the substrate is cooled to temperatures below
room temperature to prevent condensation happening anywhere else before
100% RH are reached at the substrate level. From the RH measured at
room temperature right next to the substrate, as described earlier,
the real RH that the sample is exposed to at the substrate temperature
can be calculated via eqs and 1. As stated above, temperature gradients
need to be addressed with care. However, we believe this procedure
leads to reliable RH values, as supported by the fact that the point
of condensation could be estimated by ±2% RH, so within the error
of the humidity sensor used in all the investigated temperatures.A more systematic and detailed view on the swelling behavior of
the hydrogel thin films investigated in humid environment is given
in Figure ; there,
the normalized film thickness (d/ddry in part a)) and the refractive index (n in part b)) are plotted as a function of the relative humidity as
obtained from SE measurements, while increasing the RH from 0 to almost
100% at 20 °C substrate temperature. From this data, the response
behavior of the hydrogel thin film to relative humidity can be divided
into three different regions (boxes in Figure ): First, after starting off in dry atmosphere
(0% RH), up to ∼40% RH, the film thickness increases only slightly
(by ∼4%), whereas the refractive index also increases with
respect to the dry state. This phenomenon occurred in all the temperatures
investigated. The peak in refractive index appears at very similar
values within the ∼5% resolution arising from the RH increase
steps (data not shown). This can be viewed, primarily, as filling
of voids in the polymeric structure with water, as the refractive
index of water (∼1.33) is higher than that of air (∼1.0).
For comparison, the investigated polymer exhibits a refractive index
of ∼1.52 in the dry state (see Figure ). Second, from 40% up to ∼80% RH,
the film thickness increases fairly linearly to ∼15% when compared
to the dry state. In this region, the thin film starts to incorporate
water into its polymeric structure, which can be observed as a decrease
of the system’s refractive index. However, the response is
still small compared to the swelling in water at the specific temperature.
Third, above 80% RH, the film thickness increases in a highly nonlinear
manner, approaching the film thickness of the system observed in water
at the corresponding temperature. The observed refractive index trend
in the mentioned region above 80% RH exhibits the corresponding behavior.
Figure 4
(a) Film
thickness normalized by the dry thickness measured in
pure nitrogen atmosphere (d/ddry) and (b) refractive index (n) as obtained
from spectroscopic ellipsometry measurements of 50 nm p(NVCL-co-DEGDVE) thin film samples while ramping the relative
humidity from 0 to ∼100% at 20 °C substrate temperature;
data in part a, also part of Figure .
(a) Film
thickness normalized by the dry thickness measured in
pure nitrogen atmosphere (d/ddry) and (b) refractive index (n) as obtained
from spectroscopic ellipsometry measurements of 50 nm p(NVCL-co-DEGDVE) thin film samples while ramping the relative
humidity from 0 to ∼100% at 20 °C substrate temperature;
data in part a, also part of Figure .The monotonous proportionality
of the film thickness to the relative
humidity at a certain temperature can directly be used in sensor and
actuator setups (as shown below). The refractive index cannot be utilized
in the whole range of RH due to its non-monotonous behavior. However,
two regions can be identified with respect to n:
Up to 40% RH, the response is small, but monotone; above 40%, the
response is larger and also monotone. For example, in optical systems
for RH determination, the refractive index can, thus, be employed
in applications, where the environment remains within one of these
mentioned ranges.
Kinetics in Water
As, especially,
the response characteristics
of the film thickness of the presented system to various environmental
conditions are promising, to test its applicability to sensors and
actuators, also the kinetic response behavior was experimentally probed.
First of all, the kinetic swelling of the synthesized layers when
flooded with water after being in dry atmosphere has been investigated
by in situ SE measurements (see Figure ). The fast response of the
film was shown, as the film thickness did not change for more than
1% already ∼30s after the water exposure had been started.
Anyhow, the behavior cannot be further resolved by SE because it takes
time to fill the entire measurement chamber (5 mL) with water and
align the sample, during which it is not possible to record meaningful
data due to multiple gas–liquid interfaces interfering with
the measurement. However, these measurements appear to yield
stable thickness values already within seconds of water exposure,
perfectly in agreement with the values obtained from the temperature-dependent
SE measurements in water (cf. Figure a).
Figure 5
Thickness normalized by the dry thickness measured in
nitrogen
atmosphere (d/ddry) obtained
from spectroscopic ellipsometry measurements as a function of time
while exposing a p(NVCL-co-DEGDVE) sample (50 nm
dry thickness) to water after starting off the measurement in dry
atmosphere; the dashed lines give interpolated data during flooding
and aligning of the liquid cell of the ellipsometer.
Thickness normalized by the dry thickness measured in
nitrogen
atmosphere (d/ddry) obtained
from spectroscopic ellipsometry measurements as a function of time
while exposing a p(NVCL-co-DEGDVE) sample (50 nm
dry thickness) to water after starting off the measurement in dry
atmosphere; the dashed lines give interpolated data during flooding
and aligning of the liquid cell of the ellipsometer.
Kinetics in Humid Environment
As the response to water
appears to be promisingly fast, also the kinetic behavior of the hydrogel
thin film in humid environment was investigated. At a constant temperature
(24 °C), the humid environment in the measurement chamber of
the ellipsometer was changed by controlling the flow rates of humid
and dry nitrogen vapors; the RH was measured in situ in the chamber. Relative humidity levels of below 20%, around 50%
and above 80% were introduced to investigate the applicability in
differently humid environments. The evident response in thickness
was utilized to calculate directly back to an RH value via the d/ddry-vs-RH plots measured
for the respective temperature (cf. Figure ). For that, a triple-exponential function
was used to interpolate the data. The respective responses of the
investigated layers in RH together with the measurements of the reference
sensor can be found in Figure .
Figure 6
Relative humidity (RH) as evaluated from the response of the sensor
layers (cf. Figure ) and from the reference sensor during RH exposure at different levels
over time at 25 °C.
Relative humidity (RH) as evaluated from the response of the sensor
layers (cf. Figure ) and from the reference sensor during RH exposure at different levels
over time at 25 °C.The film thickness and,
hence, the calculated RH follow the signal
of the relative humidity as measured by the reference sensor and also
stabilize, when a constant humidity is reached. The data suggest a
faster response of the investigated layer compared to the signal of
the reference sensor for all humidity levels employed. From the response
curves over time, the respective time constants (τ) could be
extracted from fitting double-exponential functions to the measured
data. The time constants of the p(NVCL-co-DEGDVE)
were evaluated to be τ = (5.3 ± 0.6) s, τ = (5.1 ± 0.3) s, and τ = (4.9 ± 0.2) s. It
should be noted that the measurement time of the ellipsometer was
set to 2 s. The time constants determined from the response of the
reference sensor are τ = (14.4 ± 0.4), τ = (8.8 ± 0.1) s, and τ = (7.6 ± 0.1) s. A secondary
time constant of about 1 min was evaluated to be present for all the
data sets for both the responses of the investigated layer and the
reference sensor, suggesting that the corresponding “long-term”
response stems from the experimental setup. Hence, the primary time
constants evaluated from all the measurements yield information about
the kinetics of the response of the investigated systems. Also for
these primary time constants, the kinetics of the visible/extracted
responses of both the investigated layer and the reference sensor
(τ, as given above) are, most probably, a convolution of the
real response of the sensing system (i.e., water uptake behavior,
τ) and the time it takes to
fill the measurement chamber experimentally with a certain relative
humidity (τ); thus, τ
= τ+ τ.Overall, the response of
the 50 nm p(NVCL-co-DEGDVE)
layer determined from the increase in thickness upon humidity exposure
appears to be about twice as fast as the response of the reference
sensor. For decreasing the RH toward dry N2 atmosphere,
similar results were obtained; also when investigating the response
at 10 °C, the p(NVCL-co-DEGDVE) film was observed
to respond faster than the reference sensor, but the experimental
setup at lower temperatures appears to stabilize slower as temperature
gradients are involved (data not shown). In the data sheet of the
reference sensor, a response time of 8 s is given; the response of
the sensor to RH levels above 50% is comparable. At lower RH, the
apparent larger time constant determined from the response of the
sensor suggests a strong contribution from the experimental setup.
For the reference sensor and the investigated polymer layer, the time
constants appear to be lower (and, thus, the responses faster) for
higher RH. Intuitively, the opposite would be expected, as it takes
more time to experimentally reach a higher level of RH after starting
out in dry nitrogen environment. However, the different flow rates
of dry N2 and a certain level of humidity (as controlled
by the needle valves experimentally) could interfere with this consideration.
As a consequence, it is not certain how much the experimental filling
process (τ) contributes to
the evaluated time constants; in general, it seems to be plausible
to assume a filling time in the order of seconds. Regardless, the
developed sensing layer is at least as fast as evident from the derived
response times or, most probably, faster.The similar time constants
at different levels of RH are, especially,
promising for applications of the layers in humidity sensors, where
operation over a wide range of humidity is needed. The observed fast
responses make the layers appealing for applications, where changes
in the RH need to be resolved in the order of seconds or below. An
example of a fast sensor device based on such stimuli-responsive hydrogels
is presented in Buchberger et al.[26]
Testing
of the Sensor/Actuator Hygromorphic Device
The thermoresponsive
swelling at high RH could be interesting for
actuator setups, where two stimuli, T and RH, are used to obtain a
mechanical deformation of the system. Hygromorphs, as objects responding
to environmental humidity by changing their shape, show striking examples
in nature (especially in plants, such as in pine cones)[27] and they have recently gained much interest
for the development of biomimetic artificial structures for sensing
and soft robotics.[18,28] To the author’s knowledge,
this is one of the first examples of a fast, humidity-driven and temperature-controlled
actuator with a bilayered structure involving hydrogel thin films.To investigate this potential application and provide a first proof
of concept, a flower-shaped hygromorphic sensor/actuator system was
fabricated (the schematics/design are provided in Figure S1, Supporting Information) based on a bilayer of
PDMS/p(NVCL-co-DEGDVE). In such bilayer, the PDMS
substrate (55 μm thick) is acting as a mechanically passive
and elastic foundation, while the pVNCL coating (just 300 nm thick)
is providing the capability to be swollen upon water uptake and thermoresponsiveness.
Each petal of the flower structure can thus act as a simple unimorph
actuator (bending beam geometry) upon uptake/release of water from
the environment, as schematized in Figure S1. The double-layered device setup was tested upon changing the relative
humidity and the substrate temperature. Movies (see Supporting Information) and pictures (see Figure ) were taken during these tests.
The flower-shaped devices reacted to the stimuli by bending of their
petals due to swelling/deswelling of the p(NVCL-co-DEGDVE). As can be appreciated in the pictures and videos, the device
is able to act against gravity upon deswelling (RH-response, Figure a) and upon swelling
(in the experiment to probe its temperature-response, Figure b). The (unbent) reference
state of the flower was retrieved by lowering the RH (Figure a). Alternatively, the humidity-driven
swelling/deswelling could be triggered by temperature, given the thermoresponsiveness
of the active layer (Figure b). Again, the kinetics of the mechanical response appear
to be very fast (i.e., in the order of seconds) and mostly limited
by the kinetics of the external stimuli applied: The temperature ramp
was applied with cooling/heating rates of 60 °C/min. In case
of RH, with the flows supplied, it takes about 10 s to reach the respective
maximal/minimal values. The bending radius of the petals is unequivocally
set by a specific couple of temperature/humidity values; this fact
makes the system interesting for further investigation to obtain reliable
and fully reversible actuation. Remarkably, the bending motion is
obtained despite the low thickness of the pNVCL layer (i.e., 300 nm)
compared to the PDMS foundation (around 150 times thinner) and compared
to similar test devices (e.g., millimeters in thickness,[19,21] or hydrogel films of tens of micrometers thickness[20]). Such results support our hypothesis that the large and
fast response of the pNVCL layers can be used for fast and integrated
humidity–temperature actuators and sensors.
Figure 7
Series of pictures of
flower-shaped PDMS/pNVCL-based double-layer
hygromorphic devices taken during ramping the (a) RH in a chamber
between 10 and 70% at 25 °C (active pNVCL layer pointing upward)
and (b) plate temperature T between 40 and 10 °C
in a humid environment of 35% RH (measured at room temperature, i.e.,
25 °C) (active layer of pNVCL pointing downward). Bending of
petals due to differential swelling of the thermoresponsive polymer
film as a function T/RH and time t.
Series of pictures of
flower-shaped PDMS/pNVCL-based double-layer
hygromorphic devices taken during ramping the (a) RH in a chamber
between 10 and 70% at 25 °C (active pNVCL layer pointing upward)
and (b) plate temperature T between 40 and 10 °C
in a humid environment of 35% RH (measured at room temperature, i.e.,
25 °C) (active layer of pNVCL pointing downward). Bending of
petals due to differential swelling of the thermoresponsive polymer
film as a function T/RH and time t.
Conclusions
Thermoresponsive
p(NVCL-co-DEGDVE) thin films
were successfully synthesized via iCVD and probed in terms of response
behavior to temperature and different environments (i.e., water, humidity,
N2). This study revealed that in water, the layer swells
to up to 250% of its dry thickness below the transition temperature
and repels out water upon undergoing a phase transition toward a shrunken
state at elevated temperatures. In nitrogen atmosphere, only thermal
expansion (∼1.5% thickness increase) is observed upon heating
the samples in a similar temperature range as in water (10 to 50 °C).
In humid environment, a first detailed study on the swelling behavior
of hydrogel thin films shows that the layer monotonously swells with
increasing the RH. To the author’s knowledge, with the described
measurement setup, this is also the first contribution in which thermoresponsive
hydrogel thin films are investigated at humidity levels above 80%
RH below and above their LCST. This is why this is also the first
contribution in which the thermoresponsiveness of hydrogel thin films
is genuinely proved to occur as a change in film thickness and refractive
index at different temperatures (also) in humid environment. We show
that the thermoresponsive transition (in film thickness and refractive
index) only occurs at humidity levels above 80% RH for the investigated
system. Toward 100% RH, the film thickness approaches the values observed
in water at the respective temperatures. We also observed a significant
response in low RH (already below 5% RH), with the moisture swelling
the thin film (up to 4%), but mainly replacing air in the polymeric
structure up to ∼40% RH.Kinetically, the investigated
p(NVCL-co-DEGDVE)
thin films were found to respond quickly to changes in the environment
(i.e., water and different levels of humidity). With the help of calibrating
the system by measuring a thickness-vs-RH curve, in subsequent measurements,
the RH was directly evaluated from these curves after measuring the
thickness at a certain temperature. The response in differently humid
environments was about twice as fast as the one of the market-available
reference sensor used.The fast response together with the similar
time constants measured
for differently humid environments make the investigated system particularly
interesting for its application in sensor and actuator setups. Such
applicability was successfully demonstrated in flower-shaped double-layered
hygromorphic devices. The artificial petals bend due to a humidity-driven
and temperature-tunable actuation mechanism, showcasing also a fast
mechanical response against gravity in both swelling and deswelling.