Impact of interruptions in radiofrequency energy delivery on lesion characteristics.

Daniel J Friedman1, Jed A Overmann2, Jeffrey M Fish2, Stephen A Gaeta3, John H Tranter2, Riki Thao2, Jonathan P Piccini4.   

Abstract

BACKGROUND: During catheter ablation, delivery of radiofrequency (RF) energy to a target site is sometimes interrupted by catheter instability and clinical factors. The impact of interruption of RF delivery on lesion characteristics has not been characterized.
OBJECTIVE: The purpose of this study was to determine the impact of interruption of RF application on lesion size.
METHODS: Forty-two RF ablation lesions (21 left ventricle, 21 right ventricle) were created in the ventricles of 6 swine using power control mode (30 W; target contact force 15g) with 1 of 3 conditions: 15-second ablation (15s), 30-second ablation (30s), or two 15-second ablations (15s×2) at the same site separated by a 2-minute pause.
RESULTS: Lesion volume was significantly larger for 30s lesions (501 ± 146 mm3) compared to both 15s×2 (314 ± 98 mm3) and 15s (242 ± 104 mm3) lesions (P <.001 for both pairwise comparisons). Compared to 15s lesions, lesion volume was numerically greater for 15s×2 lesions, but this did not reach statistical significance (P = .087). Differences in lesion volume between 30s and 15s×2 lesions were driven mainly by differences in lesion width (10.7 ± 1.1 mm vs 9.1 ± 1.7 mm; P = .04) rather than depth (9 ± 1.2 mm vs 8.4 ± 1.2 mm; P = .29). There were no differences in mean contact force by group. There was no difference in total force-time integral for the 30s and 15s×2 lesion groups [median 444 (interquartile range 312) g∙s vs 380 (164) g∙s; P = 1].
CONCLUSION: Compared to lesions resulting from continuous RF ablation, lesions resulting from interrupted ablation have a smaller overall lesion volume, predominantly due to smaller lesion width. These data suggest that if disruption in energy delivery occurs, lesions may need closer spacing to avoid gaps.
Copyright © 2020 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Catheter ablation; Electrophysiology; Lesion formation; Radiofrequency ablation; Ventricular ablation

Year:  2020        PMID: 32200047     DOI: 10.1016/j.hrthm.2020.03.008

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  3 in total

1.  Predictors of Lesions Contiguity and Transmurality in Canine Ventricular Models After Catheter Ablation.

Authors:  Abdel Hadi El Hajjar; Mario Mekhael; Chao Huang; Charbel Noujaim; Yichi Zhang; Eugene Kholmovski; Tarek Ayoub; Chan Ho Lim; Nassir Marrouche
Journal:  Front Cardiovasc Med       Date:  2022-06-23

2.  Continuous and discontinuous radiofrequency energy delivery on the atrial free wall: Lesion transmurality, width, and biophysical characteristics.

Authors:  Daniel J Friedman; Jed A Overmann; Jeffrey M Fish; Stephen A Gaeta; John H Tranter; Riki Thao; Jonathan P Piccini
Journal:  Heart Rhythm O2       Date:  2021-11-06

3.  Effect of Baseline Impedance in Radiofrequency Delivery on Lesion Characteristics and the Relationship Between Impedance and Steam Pops.

Authors:  Lijuan Qu; Min Guo; Meng Sun; Rui Wang; Nan Zhang; Xin Li
Journal:  Front Cardiovasc Med       Date:  2022-04-27
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.