Literature DB >> 32187386

A robust automated surface-matching registration method for neuronavigation.

Yifeng Fan1, Xufeng Yao2, Xiufang Xu1.   

Abstract

PURPOSE: The surface-matching registration method in the current neuronavigation completes the coarse registration mainly by manually selecting anatomical landmarks, which increases the registration time, makes the automatic registration impossible and sometimes results in mismatch. It may be more practical to use a fast, accurate, and automatic spatial registration method for the patient-to-image registration.
METHODS: A coarse-to-fine spatial registration method to automatically register the patient space to the image space without placing any markers on the head of the patient was proposed. Three-dimensional (3D) keypoints were extracted by 3D Harris corner detector from the point clouds in the patient and image spaces, and used as input to the 4-points congruent sets (4PCS) algorithm which automatically registered the keypoints in the patient space with the keypoints in the image space without any assumptions about initial alignment. Coarsely aligned point clouds in the patient and image space were then fine-registered with a variant of the iterative closest point (ICP) algorithm. Two experiments were designed based on one phantom and five patients to validate the efficiency and effectiveness of the proposed method.
RESULTS: Keypoints were extracted within 7.0 s with a minimum threshold 0.001. In the phantom experiment, the mean target registration error (TRE) of 15 targets on the surface of the elastic phantom in the five experiments was 1.17 ± 0.04 mm, and the average registration time was 17.4 s. In the clinical experiments, the mean TRE of the targets on the first, second, third, fourth, and fifth patient's head surface were 1.70 ± 0.32 mm, 1.83 ± 0.38 mm, 1.64 ± 0.3 mm, 1.67 ± 0.35 mm, and 1.72 ± 0.31 mm, respectively, and the average registration time was 21.4 s. Compared with the method only based on the 4PCS and ICP algorithm and the current clinical method, the proposed method has obvious speed advantage while ensuring the registration accuracy.
CONCLUSIONS: The proposed method greatly improves the registration speed while guaranteeing the equivalent or higher registration accuracy, and avoids a tedious manual process for the coarse registration.
© 2020 American Association of Physicists in Medicine.

Entities:  

Keywords:  4-points congruent sets; keypoint; neuronavigation; spatial registration

Mesh:

Year:  2020        PMID: 32187386     DOI: 10.1002/mp.14145

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Calibrating 3D Scanner in the Coordinate System of Optical Tracker for Image-To-Patient Registration.

Authors:  Wenjie Li; Jingfan Fan; Shaowen Li; Zhaorui Tian; Zhao Zheng; Danni Ai; Hong Song; Jian Yang
Journal:  Front Neurorobot       Date:  2021-05-14       Impact factor: 2.650

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.