PURPOSE: The Italian Tailored Assessment of Lung Indeterminate Accidental Nodule (ITALIAN) trial is a trial drawn to determine the performance of 18F-FDG-PET/CT in patients with solitary pulmonary nodules (SPN), stratified for a different kind of risk. An additional end-point was to compare the diagnostic information and estimated dosimetry, provided by a segmental PET/CT (s-PET/CT) acquisition instead of a whole body PET/CT (wb-PET/CT), in order to evaluate if segmental thoracic PET/CT can be used in patients with SPN.Entities:
Keywords: FDG; Lung cancer; PET/CT; likelihood; single pulmonary nodules
Year: 2020 PMID: 32186274 PMCID: PMC8206189 DOI: 10.2174/1874471013666200318142210
Source DB: PubMed Journal: Curr Radiopharm ISSN: 1874-4710
Technical characteristics of PET/CT scanner.
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| 1 | S. G. Moscati Hospital, Avellino | Ge Discovery 710 | 60±10 min | TOF | No |
| 2 | Veneto Institute of Oncology, Padua | Biograph 16S updated with HD software, Siemens | 60±10 min | PSF | No |
| 3 | University of Naples Federico II, Naples | Discovery LS scanner | 60±10 min | PSF | No |
| 4 | Hospital of Bolzano, Bolzano | Philips Gemini TF 16 | 60±10 min | TOF | No |
| 5 | Rionero in Vulture Hospital, Rionero in Vulture, Potenza | GE VCT 64 Slices | 60±8 min | PSF | Ge-68 phantom-based FDG-PET site qualification (FIL) |
| 6 | Università Cattolica del S. Cuore, Rome | Biograph, Siemens | 60±10 min | TOF | No |
| 7 | Humanitas Hospital, Rozzano, Milan | 1) Biograph 6, Siemens | 60±10 min | Measured | EARL accreditation for both the scanners |
| 8 | Azienda Ospedaliera Universitaria Integrata di Verona, Verona | GEMINI TF BIG BORE software version 3.6, Philips | 60±10 min | TOF | No |
| 9 | Medicina Futura IOS, Acerra, | Siemens biograph TruePoint 6 slice | 60±10 min | PSF | No |
| 10 | SDN Foundation, Naples | 1) GE Discovery 710 | 60±10 min | PSF, TOF | No |
| 11 | San Gerardo Hospital, University of Milano Bicocca, Monza | 1) Discovery 600, GE Healthcare | 60±10 min | PSF | EARL accreditation |
| 12 | Policlinico S. Orsola Malpighi, University of Bologna, Bologna | E Discovery STE PET/CT system / GE Discovery D710 PET/CT System | 60±5 min | PSF, TOF | No |
| 13 | University Tor Vergata, Roma | GE Discovery VCT | 60±5 min | PFS | No |
Center, City PET/CT scanner Imaging reconstruction Ospedale S.G. Moscati, Avellino Discovery 710 64 slices, GE Istituto Oncologico Veneto IOV - IRCCS, Padova Biograph 16, Siemens Università di Napoli Federico II, Napoli Discovery LS, GE Ospedale di Bolzano, Bolzano Gemini TF 16, Philips Centro di Riferimento per il Cancro della Basilicata-IRCCS, Rionero in Vulture (Potenza) VCT 64, GE Università Cattolica del Sacro Cuore, Roma Biograph, Siemens Ospedale Humanitas, Milano Biograph 6, Siemens and Discovery ST 690, GE Azienda Ospedaliera Universitaria di Verona, Verona Gemini TF 3.6, Philips Medicina Futura IOS, Acerra (Napoli) Biograph, Siemens SDN-IRCCS, Napoli Discovery 710, GE and Gemini TF 64, Philips Università di Milano Bicocca, Milano Discovery 600 and Discovery IQ 5 Rings, GE Università degli Studi di Bologna, Bologna Discovery 710, GE Università Tor Vergata, Roma Discovery VCT, GE TOF PSF PSF TOF PSF TOF Measured TOF PSF PSF, TOF PSF PSF, TOF PSF TOF time of flight, PSF point spread function