Immediate and 1-Year Outcomes with a Slim Modiolar Cochlear Implant Electrode Array.

Nedim Durakovic1, Dorina Kallogjeri1, Cameron C Wick1, Jonathan L McJunkin1, Craig A Buchman1, Jacques Herzog1.   

Abstract

OBJECTIVE: To explore the immediate and 1-year outcomes of patients who underwent implantation with the slim modiolar electrode (SME). STUDY
DESIGN: Consecutive case series with chart review.
SETTING: Tertiary referral academic center. SUBJECT AND METHODS: Between May 2016 and August 2018, a total of 326 cochlear implantations (CIs) were performed. Intraoperative x-rays were performed in all cases to identify tip rollovers. Scalar location was identified for 76 CIs that had postoperative computed tomography reconstructions. Speech outcomes were measured at 3, 6, and 12 months with consonant-nucleus-consonant word and AzBio sentences in quiet and noise (+10-dB signal-to-noise ratio). Preservation of hearing was defined as maintaining a low-frequency pure tone average ≤80 dB at 250 and 500 Hz.
RESULTS: Among 326 CIs, 23 (7%) had tip rollovers. Postoperative reconstructions revealed 5 of 76 (6.6%) scalar translocations. A subset of 177 cases met criteria for evaluation of speech perception scores. The marginal mean differences between presurgery and 12 months for speech tests were as follows: consonant-nucleus-consonant, 43.7 (95% CI, 39.8-47.6); AzBio in quiet, 49.7 (95% CI, 44.9-54.4); and AzBio in noise, 29.9 (95% CI, 25.2-34.7). Sixty-one patients were identified with preservable hearing (low-frequency pure tone average ≤80 dB), and 12 of 61 (20%) preserved hearing at 1 year.
CONCLUSION: CI with SME provides reliable scala tympani insertion in a consistent perimodiolar position. An initially increased tip rollover rate improved with case volume and sheath design improvement. For long-term outcomes, speech performance was comparable to that of other cochlear implants. While hearing preservation for the SME may be better than prior perimodiolar electrodes, consistent outcomes are unlikely.

Entities:  

Keywords:  cochlear implantation; electrode array placement; hearing preservation; low-frequency hearing; perimodiolar electrodes; speech recognition scores; tip fold-over

Mesh:

Year:  2020        PMID: 32180504     DOI: 10.1177/0194599820907336

Source DB:  PubMed          Journal:  Otolaryngol Head Neck Surg        ISSN: 0194-5998            Impact factor:   3.497


  5 in total

1.  Early Hearing Preservation Outcomes Following Cochlear Implantation With New Slim Lateral Wall Electrode Using Electrocochleography.

Authors:  Amit Walia; Matthew A Shew; Abhinav Ettyreddy; Shannon M Lefler; Pawina Jiramongkolchai; Cameron C Wick; Nedim Durakovic; Craig A Buchman; Jacques A Herzog
Journal:  Otol Neurotol       Date:  2022-04-01       Impact factor: 2.311

2.  Simpler and effective radiological evaluations for modiolar proximity of a slim modiolar cochlear implant electrode.

Authors:  Sang-Yeon Lee; Jin Hee Han; Marge Carandang; Yun Jung Bae; Byung Yoon Choi
Journal:  Sci Rep       Date:  2020-10-19       Impact factor: 4.379

3.  Review of Speech Outcomes in Cochlear Implant Recipients at a Nascent Cochlear Implant Program.

Authors:  Nathan Aminpour; Laura Levin; Mary Finkbone; Michael Morikawa; Melissa Blumgart; H Jeffrey Kim; Michael Hoa
Journal:  Cureus       Date:  2022-02-23

4.  Suitable Electrode Choice for Robotic-Assisted Cochlear Implant Surgery: A Systematic Literature Review of Manual Electrode Insertion Adverse Events.

Authors:  Paul Van de Heyning; Peter Roland; Luis Lassaletta; Sumit Agrawal; Marcus Atlas; Wolf-Dieter Baumgartner; Kevin Brown; Marco Caversaccio; Stefan Dazert; Wolfgang Gstoettner; Rudolf Hagen; Abdulrahman Hagr; Greg Eigner Jablonski; Mohan Kameswaran; Vladislav Kuzovkov; Martin Leinung; Yongxin Li; Andreas Loth; Astrid Magele; Robert Mlynski; Joachim Mueller; Lorne Parnes; Andreas Radeloff; Chris Raine; Gunesh Rajan; Joachim Schmutzhard; Henryk Skarzynski; Piotr H Skarzynski; Georg Sprinzl; Hinrich Staecker; Timo Stöver; Dayse Tavora-Viera; Vedat Topsakal; Shin-Ichi Usami; Vincent Van Rompaey; Nora M Weiss; Wilhelm Wimmer; Mario Zernotti; Javier Gavilan
Journal:  Front Surg       Date:  2022-03-24

5.  Intracochlear electrode array position and cochlear implant outcomes using the nucleus slim modiolar electrode and the extended round window approach: a follow-up study.

Authors:  Tim M Klabbers; Floris Heutink; Wendy J Huinck; Willem-Jan van der Woude; Berit M Verbist; Emmanuel A M Mylanus
Journal:  Eur Arch Otorhinolaryngol       Date:  2022-01-18       Impact factor: 3.236

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.