Comparison of automated irrigation systems using an in vitro ureteroscopy model.

Donald Fedrigon1, Luay Alshara1, Manoj Monga1.   

Abstract

INTRODUCTION: Two automated irrigation systems have been released for use during endoscopic procedures such as ureteroscopy: the Cogentix RocaFlow® (CRF) and Thermedx FluidSmart® (TFS). Accurate pressure control using automated systems may help providers maintain irrigation pressures within a safe range while also providing clear visualization. Our objective was to directly compare these systems based on their pressure accuracy, pressure-flow relationships, and fluid heating capabilities in order to help providers better utilize the temperature and pressure settings of each system.
MATERIALS AND METHODS: An in vitro ureteroscopy model was used for testing, consisting of a short semirigid ureteroscope (6/7, 5F, 31cm Wolf 425612) connected to a continuous digital pressure transducer (Meriam m1550). Each system pressure output and flow-rate, via 100mL beaker filling time, was measured using multiple trials at pressure settings between 30 and 300mmHg. Output fluid temperature was monitored using a digital thermometer (Omega DP25-TH).
RESULTS: The pressure output of both systems exceeded the desired setting across the entire tested range, a difference of 15.7±2.4mmHg for the TFS compared to 5.2±1.5mmHg for the CRF (p < 0.0001). Related to this finding, the TFS also had slightly higher flow rates across all trials (7±2mL/min). Temperature testing revealed a similar maximum temperature of 34.0⁰C with both systems, however, the TFS peaked after only 8 minutes and started to plateau as early as 4-5 minutes into the test, while the CRF took over 18 minutes to reach a similar peak.
CONCLUSIONS: Our in vitro ureteroscopy testing found that the CRF system had better pressure accuracy than the TFS system but with noticeably slower fluid heating capabilities. Each system provided steady irrigation at safe pressures within their expected operating parameters with small differences in performance that should not limit their ability to provide steady irrigation at safe pressures. Copyright® by the International Brazilian Journal of Urology.

Entities:  

Keywords:  Research; Technology; Ureteroscopy

Year:  2020        PMID: 32167702     DOI: 10.1590/S1677-5538.IBJU.2019.0230

Source DB:  PubMed          Journal:  Int Braz J Urol        ISSN: 1677-5538            Impact factor:   1.541


  1 in total

1.  Editorial Comment: Continuous monitoring of intrapelvic pressure during flexible ureteroscopy using a sensor wire: a pilot study.

Authors:  Alexandre Danilovic
Journal:  Int Braz J Urol       Date:  2021 Jan-Feb       Impact factor: 1.541

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.