Estimation of intraocular pressure in normal canine eyes utilizing the newly introduced TonoVet Plus and TonoPen Avia, and their comparison to the established TonoVet.

Gil Ben-Shlomo1,2, Sarah F Muirhead2.   

Abstract

PURPOSE: To evaluate the IOP values obtained from normal canine eyes, by means of the TonoVet Plus (TVP) and TonoPen Avia (TPA), and compare them to an established tonometer, the TonoVet (TV).
METHODS: Tonometry was performed bilaterally in 50 healthy dogs, age 1-11 years. The rebound tonometers (TV and TVP) were used first (in a random order), and then, one drop of tetracaine hydrochloride was applied to each eye, and approximately one minute later the IOP was estimated using the TPA. The mean ± standard deviation (SD) was calculated for each tonometer and a paired Student's t test was used to compare the IOP values between tonometers. A P value ≤ .05 was considered significant. The agreement between tonometers was assessed by a Bland-Altman plot.
RESULTS: The mean ± SD (range) IOP values were 15.0 ± 3.2 mm Hg (7-22), 19.2 ± 3.1 mm Hg (11-25), and 12.8 ± 2.9 mm Hg (6-19), for the TV, TVP, and TPA, respectively. The IOP values obtained by each tonometer were significantly different compared to the other two (P < .0001). There was a good agreement with fixed bias between all tonometers.
CONCLUSIONS: The average IOP values of the TVP were significantly higher than those of the TV, which were significantly higher than those of the TPA, in normal canine eyes. Knowing the normal IOP values for these tonometers and the bias between them will help the clinician with the interpretation of IOP values obtained by these devices and compare between them.
© 2020 American College of Veterinary Ophthalmologists.

Entities:  

Keywords:  TonoPen Avia; TonoVet; TonoVet Plus; dog; intraocular pressure; tonometry

Year:  2020        PMID: 32154987     DOI: 10.1111/vop.12747

Source DB:  PubMed          Journal:  Vet Ophthalmol        ISSN: 1463-5216            Impact factor:   1.644


  4 in total

1.  Canine endotheliitis: Clinical characteristics, advanced imaging features, and treatment.

Authors:  Melaney A Mayes; Maria Isabel Casanova; Sangwan Park; Kirsten Steele; Lana Linton; Soohyun Kim; Kathryn L Good; Bret A Moore; Georgina M Newbold; Brian C Leonard; Jennifer Y Li; Sara M Thomasy
Journal:  Vet Ophthalmol       Date:  2021-12-31       Impact factor: 1.444

2.  Comparison of two rebound tonometers in healthy horses.

Authors:  Minna P Mustikka; Elina M Pietilä; Anna K Mykkänen; Thomas S C Grönthal
Journal:  Vet Ophthalmol       Date:  2020-09-04       Impact factor: 1.644

3.  Comparison among TonoVet, TonoVet Plus, Tono-Pen Avia Vet, and Kowa HA-2 portable tonometers for measuring intraocular pressure in dogs.

Authors:  João Victor Goulart Consoni Passareli; Felipe Franco Nascimento; Giovana José Garcia Estanho; Claudia Lizandra Ricci; Glaucia Prada Kanashiro; Rogério Giuffrida; Silvia Franco Andrade
Journal:  Vet World       Date:  2021-09-21

4.  Intraocular pressure measurements using the TONOVET® rebound tonometer: Influence of the probe-cornea distance.

Authors:  Blanche D Rodrigues; Fabiano Montiani-Ferreira; Mariza Bortolini; André T Somma; András M Komáromy; Peterson Triches Dornbusch
Journal:  Vet Ophthalmol       Date:  2020-10-17       Impact factor: 1.644

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.