Literature DB >> 32142124

3D mapping for the identification of the fossa ovalis in left atrial ablation procedures: a pilot study of a first step towards an electroanatomic-guided transseptal puncture.

Martin Eichenlaub1, Reinhold Weber1, Jan Minners1, Hans-Jürgen Allgeier1, Amir Jadidi1, Björn Müller-Edenborn1, Franz-Josef Neumann1, Thomas Arentz1, Heiko Lehrmann1.   

Abstract

AIMS: Transseptal puncture (TP) for left atrial (LA) catheter ablation procedures is routinely performed under fluoroscopic guidance. To decrease radiation exposure and increase safety alternative techniques are desirable. The aim of this study was to assess whether right atrial (RA) electroanatomic 3D mapping can reliably identify the fossa ovalis (FO) in preparation of TP. METHODS AND
RESULTS: Between May 2019 and August 2019, electroanatomic RA mapping was performed before TP in 61 patients with paroxysmal or persistent atrial fibrillation. Three electroanatomic methods for FO identification, mapping catheter-induced FO protrusion, electroanatomic-guided analysis, and voltage mapping, were evaluated and compared with transoesophageal echocardiography (TOE). Mapping catheter-induced FO protrusion was feasible in 60 patients (98%) with a mean displacement of 6.8 ± 2.5 mm, confirmed by TOE, and proofed to be the most valuable and easiest marker for FO identification. Electroanatomic-guided analysis localized the FO midpoint consistently in the lower half (43 ± 7%) and posterior (18.2 ± 4.4 mm) to a line between coronary sinus and vena cava superior. Analysis of RA voltage maps during sinus rhythm (n = 40, low-voltage cut-off value 1.0 and 1.5 mV) allowed secure FO recognition in 33% and 18%, only. A step-by-step approach, combining FO protrusion (first step) with anatomy criteria in case of protrusion failure (second step) would have allowed for the correct localization of a TP site within the FO in all patients.
CONCLUSION: Right atrial electroanatomic 3D mapping prior to TP proofed to be a simple tool for FO identification and may potentially be of use in the safe and radiation-free performance of TP prior to LA ablation procedures. Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author(s) 2020. For permissions, please email: journals.permissions@oup.com.

Entities:  

Keywords:  3D mapping; Fossa ovalis; Radiation; Safety; Transseptal puncture;  Left atrial ablation

Mesh:

Year:  2020        PMID: 32142124     DOI: 10.1093/europace/euaa034

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  1 in total

1.  Study on the Curative Effect and Safety of Radiofrequency Catheter Ablation of Paroxysmal Atrial Fibrillation via Zero-Fluoroscopy Transseptal Puncture under the Dual Guidance of Electroanatomical Mapping and Intracardiac Echocardiography.

Authors:  Fei Hang; Liting Cheng; Zhuo Liang; Ruiqing Dong; Xinlu Wang; Ziyu Wang; Zefeng Wang; Yongquan Wu
Journal:  Cardiol Res Pract       Date:  2021-05-24       Impact factor: 1.866

  1 in total