Literature DB >> 32140438

Develop a high energy proton beam position monitor using linear contact image sensor.

Tung-Yuan Hsiao1, Huan Niu1, Tzung-Yuang Chen1, Chien-Hsu Chen1.   

Abstract

A compact beam-position monitor was constructed using a linear contact image sensor attached to a plastic scintillator and tested using a 230 MeV proton beam. The results indicate that the beam position can be obtained in real-time, and the beam position with a precision of up to 0.03 mm. The compactness and high precision of the device hold considerable potential for it to be used as a beam-position monitor and offline, daily quality assurance monitor in hadron therapy. •The method can provide a high precision and high resolution beam position for flash irradiation in particle therapy in real-time.•The method using contact image sensor with scintillator does not require a long focal length for camera and it is free of image distortion.•The method can be integrated into medical particle accelerator for feedback control and daily quality assurance.
© 2019 The Author(s).

Entities:  

Keywords:  Beam Position Monitor using contact image sensor with scintillator; Beam diagnostics; Beam position monitor; Beam profile monitor; Contact image sensor; Ionization chamber; Scintillator

Year:  2019        PMID: 32140438      PMCID: PMC7046811          DOI: 10.1016/j.mex.2019.100773

Source DB:  PubMed          Journal:  MethodsX        ISSN: 2215-0161


Specification Table

Method details

Introduction

High-energy ion beams (proton and carbon) hold great potential in the field of radiation oncology due to their special Bragg peak characteristics, permitting the deposition of most of the beam’s energy within a narrow depth range [2]. In radiation therapy, accurate and precise delivery of a specific radiation dose to the tumor site is crucial, and different beam monitors have been developed accordingly. These devices have been equipped with on beam-line for beam adjustment and also for day-to-day beam quality assurance. Ion chambers are the most commonly used devices to monitor the delivery of a dose from the therapy unit and comprise mainly two types: strip type and pixel matrix ionization chamber [1,[3], [4], [5]]. However, for more advanced treatments, such as hadron therapy, a faster and more precise beam-measurement device is needed. Therefore, plastic scintillators (PS) have been developed for verifying the proton range, to act as a quality assurance tool in clinical usage [6]. The fast response of a plastic scintillator is essential for its usefulness as a suitable sensor for monitoring beam direction and position during the scan of an object. Moreover, the light yield of a plastic scintillator is dose-dependent, potentially allowing it to be used as a dose monitor. However, until now, the plastic scintillator system has required a relatively large space for its installation because it was observed by a long focal lenses camera, and the long focal lenses may cause image distortion. It is necessary to position the camera off the beam axis to prevent damage to it caused by high-energy particle bombardment [6]. These requirements have limited the practical application of plastic scintillators. The use of scintillator fibers can obviate the installation-space challenge, but it unavoidably presents a complex and expensive front-end readout system problem [7]. The Linear Contact Image Sensor (CIS) module is the core device of a commercially available scanner [8,9]. It consists of an array of microlenses to focus light onto the corresponding photodiode detector array, which converts the light to an electronic signal. The slim shape and ultrashort focus length of the CIS make it suitable for attachment to a plastic scintillator. The proton and radiation therapy center of Chang Gung Memorial Hospital is the first proton center in Taiwan, which has a cyclotron and four treatment rooms with a rotating gantry for proton therapy. Recently, an experimental room was set up for high-energy proton irradiation for physics and biology research, as shown in Fig. 1. As a site for various types of experiment, a flexible, fast-response, and simply constructed beam monitoring system is required. We present the first-time use of a plastic scintillator attached to a slim CIS (PS + CIS) to monitor the position of a high-energy proton beam.
Fig. 1

Layout of proton irradiation experiment room in CGMH: inset shows schematic diagram of the PS + CIS test system.

Layout of proton irradiation experiment room in CGMH: inset shows schematic diagram of the PS + CIS test system.

Method and experiments

A high-energy proton beam was produced using a cyclotron (P235, SHI, Japan). The beam was passed through a transporting line, becoming incident to the experimental room, passing through an exit window and then to the air, to where the test PS + CIS was placed on a bench. The inset of Fig. 1 is a schematic diagram of the PS + CIS test system. Fig. 2 shows the picture of the test configuration of the BPM in the CGMH experimental room. The PS was covered by a black sheet in case of the influence of indoor lighting. A 60-mm-thick copper block with a 10-mm-diameter hole was placed in front of the PS + CIS test system to function as a beam-shape collimator, and used to decrease the beam jitter. The PS is BC-408. The dimensions of the PS were 100 × 100 × 3 mm3, and the density 1.03 g/cm3, which is similar to the density of tissue. The CIS used in this test was around 187 mm long with a resolution of 100 dpi (254 μm) at a scan speed <20 μs/line, and was attached to the top-edge side of the PS. The plastic scintillator converted proton deposition energy into light, which was collected by the CIS and converted to electric signals, which were then fed to the front-end readout electronics. The readout electronics comprised a Cortex-M3 RISC processor with 10-bit ADC (analog to digital converter), which sent data to a PC by a Raspberry Pi Model 3 B (signal board computer), equipped with a 1.2-GHz 64-bit quad-core ARM Cortex-A53 and 1 G-Byte of LPDDR2 RAM. A EBT3 film [10], was appended to the backside of the test PS + CIS, which recorded the beam information for comparison purposes.
Fig. 2

The picture of the test configuration of the BPM in the CGMH experimental room.

The picture of the test configuration of the BPM in the CGMH experimental room.

Method validation

Before the experiment, the background level of PS + CIS was determined in a dark box. Fig. 3 shows the background spectrum of the test PS + CIS, which primarily originated from the CIS dark current; the background level was 80 units. Thus, the threshold level of PS + CIS was set at 80 and the acquired data shifted to 80 units. Moreover, the ADC readout limitation has been customized as 944 due to the maximum number of 10 bits ADC being 1024.
Fig. 3

Background spectrum obtained in a dark box; the threshold level (red line) is set at a value of 80 units.

Background spectrum obtained in a dark box; the threshold level (red line) is set at a value of 80 units. Fig. 4(a) shows the measurement results for a proton-beam energy of 230 MeV and current of 1 nA, passing through the test PS + CIS for 21 s.
Fig. 4

Results of 230 MeV, 1 nA proton beam with a 1-cm collimator: (a) raw data of 21 measurements, (b) correspondent Gaussian fitting curves, (c) ETB3 film, (d) comparison between EBT3 and PS + CIS.

Results of 230 MeV, 1 nA proton beam with a 1-cm collimator: (a) raw data of 21 measurements, (b) correspondent Gaussian fitting curves, (c) ETB3 film, (d) comparison between EBT3 and PS + CIS. The response time of the PS and CIS are very short, and the measurement time is mainly limited by the followed analog-to-digital conversion time, the data transfer, and the data processing. For the sake of clarity, we designed the system so that one measurement took about 1 s; each measurement contained 726 data points and the total number of measurements was about 21. For human eye view, they were real-time, curve-fitted by a Gaussian equation, where the mean and standard deviation for the beam position and size were shown. The statistical information of the fitted curves is shown in Fig. 4(b), information obtained while the taking of a measurement was stopped or paused. Table 1 shows the fitting results and their FWHMs of the 21 measurements. The mean of peak position is 41.99 mm, and its standard deviation is 30 μm. These results indicate that the precision of the measured beam position is 30 μm. Fig. 4(c) shows the scanned result of the irradiated EBT3 film, the comparison of which with the PS + CIS results, in the same projection direction, is shown in Fig. 4(d). The coherent distributions between them indicate that the PS + CIS is a suitable beam position monitor. Fig. 5 shows the results of the PS + CIS at different beam intensities, without the beam-shape collimator.
Table 1

The peak positions and their FWHMs which were determined by Gaussian fit.

No.peak positionFHWM
142.0213.71
241.9813.81
342.0113.38
442.0313.73
541.9513.34
642.0213.78
741.9412.99
842.0113.13
942.0613.39
1042.0213.57
1142.0213.45
1241.9813.25
1341.9513.18
1442.0013.30
1541.9613.38
1642.0213.46
1741.9813.73
1842.0113.45
1941.9813.84
2041.9213.45
2142.0013.60



Mean41.9913.47
Std0.030.24
Unit: mm
Fig. 5

Results of 230 MeV proton beam, without the collimator, at different currents.

The peak positions and their FWHMs which were determined by Gaussian fit. Results of 230 MeV proton beam, without the collimator, at different currents. Note that the light yield of the plastic scintillator is associated with the energy deposited on it by the radiation. The signal height increases with beam intensity up to a signal unit of 800, which could be caused by the saturation level of the photodiode. The signal shape broadens at high beam current, perhaps due to some stray light passing through the edges of the focus lens. The integration of the signal intensity of each measurement vs the correspondent beam current is shown in Fig. 6. The linearity is more pronounced in the low-intensity region than in the high-intensity region, indicating that the dynamic range for 230 MeV protons in this system is around 0.5 nA–1 nA.
Fig. 6

The integration of the signal intensity vs correspondent beam current.

The integration of the signal intensity vs correspondent beam current.

Conclusions

In conclusion, we have presented a novel, high-resolution, and high-precision beam-position monitor for particle accelerators, tested with a 230 MeV proton beam. The resolution of BPM is 254 μm which defined by the 100 dpi CIS sensor we used. And the results shows that the 100 dpi CIS sensor equipped with PS can achieve precision of beam position measurement up to 30 μm by 21 measurements, and the precision is significant better than found in traditional ion chamber type detectors. Due to the physical properties of the CIS + PS and their compact size, this beam position monitor can be inserted into a beam line with limited space, and operated in real time as a beam-parameter monitor for high energy proton beam.

Declaration of Competing Interest

The authors declare no conflict of interest.
Subject Area:Physics and Astronomy
More specific subject area:Particle Accelerator
Beam Diagnostics
Method name:Beam Position Monitor using contact image sensor with scintillator
Name and reference of original method:Strip Ionization Chamber detector, Brusasco, C., et al., Strip ionization chambers as 3-D detector for hadron therapy. Nuclear Instruments and Methods in Physics Research Section A 389 (1997) 499 [1]. and Braccini, S., et al., Segmented ionization chambers for beam monitoring in hadrontherapy. Modern Physics Letters A, 2015. 30(17): p. 1540026 [3].
Resource availability:NA
  4 in total

1.  Fiber-optic Cerenkov radiation sensor for proton therapy dosimetry.

Authors:  Kyoung Won Jang; Wook Jae Yoo; Sang Hun Shin; Dongho Shin; Bongsoo Lee
Journal:  Opt Express       Date:  2012-06-18       Impact factor: 3.894

2.  Use of a two-dimensional ionization chamber array for proton therapy beam quality assurance.

Authors:  Bijan Arjomandy; Narayan Sahoo; Xiaoning Ding; Michael Gillin
Journal:  Med Phys       Date:  2008-09       Impact factor: 4.071

Review 3.  Proton therapy - Present and future.

Authors:  Radhe Mohan; David Grosshans
Journal:  Adv Drug Deliv Rev       Date:  2016-12-03       Impact factor: 15.470

4.  Quality assurance in proton beam therapy using a plastic scintillator and a commercially available digital camera.

Authors:  Mansour Almurayshid; Yusuf Helo; Andrzej Kacperek; Jennifer Griffiths; Jem Hebden; Adam Gibson
Journal:  J Appl Clin Med Phys       Date:  2017-07-29       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.