Literature DB >> 3213452

Fixation of unstable trochanteric hip fractures. A cadaver study comparing three different devices.

S Larsson1, M Elloy, L I Hansson.   

Abstract

A four-fragment trochanteric fracture was produced in 24 human cadaver femora. After reduction to anatomic position, the fractures were stabilized by either a Jewett 135 degree nail-plate, a 135 degree Nolok sliding screw-plate without a key, or a 140 degree Hansson pin-plate with an additional trochanteric plate. Using a hip-force simulator, the preparations were subjected to 10,000 loading cycles simulating partial weight bearing, followed by another 10,000 cycles with full weight bearing, while measuring the elastic and permanent fracture displacements. Failure occurred in 3/8 of the Jewett stabilized fractures and in 1/8 each of the Nolok and Hansson stabilized specimens. Each failure was preceded by an increase in elastic tilt and rotation followed by a permanent rotation after which failure occurred as varus angulation of the fracture. Further, two Jewett nails, one Nolok screw, and three Hansson pins showed bending. Due to the controlled telescopic action in the Nolok and Hansson dynamic devices, the resistance to fatigue when subjected to repetitive loadings was increased and thereby a better overall result was achieved when compared with the rigid Jewett nail-plate. To maintain the dynamic function, the strength of the dynamic screw-pin must be sufficient to withstand physiologic loadings without bending; otherwise, the telescopic action will be obstructed.

Entities:  

Mesh:

Year:  1988        PMID: 3213452     DOI: 10.3109/17453678809149419

Source DB:  PubMed          Journal:  Acta Orthop Scand        ISSN: 0001-6470


  3 in total

Review 1.  Is rotation the mode of failure in pertrochanteric fractures fixed with nails? Theoretical approach and illustrative cases.

Authors:  C Kokoroghiannis; D Vasilakos; K Zisis; G Dimitriou; E Pappa; D Evangelopoulos
Journal:  Eur J Orthop Surg Traumatol       Date:  2019-09-20

2.  [The biofeedback sole tested is a suitable device for avoiding overload under partial load].

Authors:  R Stangl; R Krug; F F Hennig; J Gusinde
Journal:  Unfallchirurg       Date:  2004-12       Impact factor: 1.000

3.  Outcomes of Internal Fixation with Compression Hip Screws in Lateral Decubitus Position for Treatment of Femoral Intertrochanteric Fractures.

Authors:  Cheon-Gon Park; Taek-Rim Yoon; Kyung-Soon Park
Journal:  Hip Pelvis       Date:  2018-12-06
  3 in total