Literature DB >> 32132311

Validation of Intraoperative Catheter Phase Mapping Using a Simultaneous Optical Measurement System in Rabbit Ventricular Myocardium.

Naoki Tomii1, Keisuke Asano2, Hiroshi Seno2, Takashi Ashihara3, Ichiro Sakuma2, Masatoshi Yamazaki2.   

Abstract

BACKGROUND: Recently, an interoperative catheter electrode mapping system, termed ExTRa Mapping (EXT), was developed for precise diagnosis and effective treatment of non-paroxysmal atrial fibrillations (non-PAF). However, the mapping accuracy of EXT is still unclear.Methods and 
Results: In this study, the reliability of the EXT in comparison with that of high-resolution optical membrane potential mapping was compared. Spiral wave re-entries (SWRs) were induced in the excised rabbit hearts (n=8, 42 episodes). Electrical signals were measured by electrodes on a transparent silicone plate, with the same arrangement as in the clinical catheter, and fluorescence signals were recorded simultaneously across the plate. Based on the phase maps derived by EXT, activation patterns (one-directed propagations: 26, rotational activities: 16) were identified correctly with 95% accuracy (40/42), and the correlation coefficient of the ratio of the non-passive period was 0.95. In the rotational episodes (15), the mean position error of the centers of gravity of the SWR trajectory (2,000 ms) was 2.0 mm. For the one-directional episodes (25), the correlation coefficient of the directions of one-way propagation was 0.99.
CONCLUSIONS: The phase map sequence by EXT is consistent with that by the analyses of high-resolution optical mapping. EXT is reliable for analyzing the activation pattern in the region of interest.

Entities:  

Keywords:  Catheter ablation; Non-paroxysmal atrial fibrillation; Phase mapping; Phase singularity

Year:  2020        PMID: 32132311     DOI: 10.1253/circj.CJ-19-1020

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  2 in total

1.  The impact of the atrial wall thickness in normal/mild late-gadolinium enhancement areas on atrial fibrillation rotors in persistent atrial fibrillation patients.

Authors:  Toshihiro Nakamura; Kunihiko Kiuchi; Koji Fukuzawa; Mitsuru Takami; Yoshiaki Watanabe; Yu Izawa; Makoto Takemoto; Jun Sakai; Atsusuke Yatomi; Yusuke Sonoda; Hiroyuki Takahara; Kazutaka Nakasone; Kyoko Yamamoto; Yuya Suzuki; Ken-Ichi Tani; Noriyuki Negi; Atsushi Kono; Takashi Ashihara; Ken-Ichi Hirata
Journal:  J Arrhythm       Date:  2022-01-13

2.  A case report on the usefulness of combining online continuous wavelet transform analysis with a novel real-time phase mapping system during nonparoxysmal atrial fibrillation catheter ablation.

Authors:  Yasuhiro Yokoyama; Hiroaki Nakamura; Nobutaka Kato; Koji Higuchi
Journal:  HeartRhythm Case Rep       Date:  2022-01-19