Elena Aloisio1, Erika Frusciante1, Sara Pasqualetti1, Ilenia Infusino1, Magdalena Krintus2, Grazyna Sypniewska2, Mauro Panteghini1.
Abstract
Background Laboratory professionals should independently verify the correct implementation of metrological traceability of commercial measuring systems and determine if their performance is fit for purpose. We evaluated the trueness, uncertainty of measurements, and transferability of six clinically important enzyme measurements (alanine aminotransferase [ALT], alkaline phosphatase [ALP], aspartate aminotransferase [AST], creatine kinase [CK], γ-glutamyltransferase [γGT], and lactate dehydrogenase [LDH]) performed on the Abbott Alinity c analytical system. Methods Target values and associated uncertainties were assigned to three pools for each enzyme by using the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference measurement procedures (RMPs) and the pools were then measured on the Alinity system. Bias estimation and regression studies were performed, and the uncertainty associated with Alinity measurements was also estimated, using analytical performance specifications (APS) derived from biological variability of measurands as goals. Finally, to validate the transferability of the obtained results, a comparison study between two Alinity systems located in Milan, Italy, and Bydgoszcz, Poland, was carried out. Results Correct implementation of traceability to the IFCC RMPs and acceptable measurement uncertainty fulfilling desirable (ALP, AST, LDH) or optimal APS (ALT, CK, γGT) was verified for all evaluated enzymes. An optimal alignment between the two Alinity systems located in Milan and Bydgoszcz was also found for all enzyme measurements. Conclusions We confirmed that measurements of ALT, ALP, AST, CK, γGT, and LDH performed on the Alinity c analytical system are correctly standardized to the IFCC reference measurement systems and the system alignment is consistent between different platforms.Entities:
Keywords: enzymes; standardization; traceability; trueness; uncertainty
Year: 2020 PMID: 32126012 DOI: 10.1515/cclm-2020-0015
Source DB: PubMed Journal: Clin Chem Lab Med ISSN: 1434-6621 Impact factor: 3.694