| Literature DB >> 32121721 |
Jiandong Sun, Yifan Zhu, Wei Feng, Qingfeng Ding, Hua Qin, Yunfei Sun, Zhipeng Zhang, Xiang Li, Jinfeng Zhang, Xinxing Li, Yang Shangguan, Lin Jin.
Abstract
Aiming at the requirement of passive terahertz imaging, we report a high-sensitivity terahertz detector based on an antenna-coupled AlGaN/GaN high-electron-mobility transistor (HEMT) at 77 K without using low-noise terahertz amplifier. The measured optical noise-equivalent power and the noise-equivalent temperature difference of the detector were about 0.3p W/H z and 370 mK in a 200 ms integration time over a bandwidth of 0.7 - 0.9 THz, respectively. By using this detector, we demonstrated passive terahertz imaging of room-temperature objects with signal-to-noise ratio up to 13 dB. Further improvement in the sensitivity may allow passive terahertz imaging using AlGaN/GaN-HEMT at room temperature.Entities:
Year: 2020 PMID: 32121721 DOI: 10.1364/OE.385042
Source DB: PubMed Journal: Opt Express ISSN: 1094-4087 Impact factor: 3.894