Mattias Kruskopf1, Albert F Rigosi2, Alireza R Panna2, Dinesh K Patel3, Hanbyul Jin1, Martina Marzano4, Michael Berilla2, David B Newell2, Randolph E Elmquist2.
Abstract
In this paper, we show that quantum Hall resistance measurements using two terminals may be as precise as four-terminal measurements when applying superconducting split contacts. The described sample designs eliminate resistance contributions of terminals and contacts such that the size and complexity of next-generation quantized Hall resistance devices can be significantly improved.Entities:
Keywords: Epitaxial graphene (EG); multi-series (MS) contacts; quantized Hall resistance (QHR) standards; quantum Hall effect (QHE); superconducting contacts
Year: 2019 PMID: 32116346 PMCID: PMC7047664 DOI: 10.1109/ted.2019.2926684
Source DB: PubMed Journal: IEEE Trans Electron Devices ISSN: 0018-9383 Impact factor: 2.917