Developing and Validating a Fast and Accurate Method to Quantify 18 Antidepressants in Oral Fluid Samples Using SPE and LC-MS-MS.

Sanghee Sarah Shin1, Damon Borg2, Richard Stripp2.   

Abstract

Antidepressant drugs are one of the most widely used medicines for treating major depressive disorders for long time periods. Oral fluid (OF) testing offers an easy and non-invasive sample collection. Detection of antidepressants in OF is important in clinical and forensic settings, such as therapeutic drug monitoring and roadside testing for driving under influence. We developed and validated a comprehensive liquid chromatography-tandem mass spectrometry method for 18 antidepressants (amitriptyline, bupropion, citalopram, clomipramine, cyclobenzaprine, desipramine, desvenlafaxine, doxepin, duloxetine, fluoxetine, imipramine, mirtazapine, nortriptyline, paroxetine, sertraline, trazodone, trimipramine, venlafaxine) in oral fluid collected by Quantisal® oral collection devices. One-half milliliter of Quantisal® OF (125 μL of neat OF) was submitted to solid-phase extraction. The chromatographic separation was performed employing a biphenyl column in gradient mode with a total run time of 5 min. The MS detection was achieved by multiple-reaction monitoring with two transitions per compound. The range for linearity of all analytes was from 10 to 1,000 ng/mL, with a limit of detection of 10 ng/mL. Intra and inter-day accuracy and precision (n = 15) were all within acceptable limits, ±20% error and ±15% relative standard deviation. Analyte recovery at 400 ng/mL concentration (n = 15) ranged from 91 to 129%. Matrix effect ranged from 73.7 to 157%. The internal proficiency test detected all antidepressants with accuracy ranging from 83.1 to 112.1%. The authentic patient sample showed a percentage difference compared to the previously calculated concentration of 86.3-111%. This method provides for the rapid detection of 18 antidepressants and metabolites in OF, which is readily applicable to a routine laboratory.
© The Author(s) 2020. Published by Oxford University Press. All rights reserved. For permissions, please email: journals.permissions@oup.com.

Entities:  

Keywords:  Antidepressant; LC–MS-MS; Oral fluid

Year:  2020        PMID: 32115632     DOI: 10.1093/jat/bkz117

Source DB:  PubMed          Journal:  J Anal Toxicol        ISSN: 0146-4760            Impact factor:   3.367


  3 in total

1.  Stir bar sorptive-dispersive microextraction by a poly(methacrylic acid-co-ethylene glycol dimethacrylate)-based magnetic sorbent for the determination of tricyclic antidepressants and their main active metabolites in human urine.

Authors:  Víctor Vállez-Gomis; Sara Exojo-Trujillo; Juan L Benedé; Alberto Chisvert; Amparo Salvador
Journal:  Mikrochim Acta       Date:  2022-01-08       Impact factor: 5.833

2.  Polydopamine coated hypodermic needles as a microextraction device for the determination of tricyclic antidepressants in oral fluid by direct infusion MS/MS.

Authors:  Carmina Vejar-Vivar; María Teresa García-Valverde; Claudia Mardones; Rafael Lucena; Soledad Cárdenas
Journal:  RSC Adv       Date:  2021-06-28       Impact factor: 3.361

3.  New Method for the Monitoring of Antidepressants in Oral Fluid Using Dried Spot Sampling.

Authors:  Sofia Soares; Tiago Rosado; Mário Barroso; Eugenia Gallardo
Journal:  Pharmaceuticals (Basel)       Date:  2021-12-08
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.