Cardiogenic shock due to leaflet migration of On-X® aortic mechanical prosthesis.

Nora Göbel1, Sara Tanriverdi2, Ragi Nagib2, Ulrich F W Franke2.   

Abstract

Cardiogenic shock is a life-threatening condition requiring fast and efficient diagnostic and therapeutic measures. In this case, the history of several cardiac surgeries hindered finding the correct diagnosis initially. Only after indication of ultima ratio cardiac redo operation the underlying cause was found intraoperatively: a defective mechanical valve prosthesis with migration of one of the two leaflets. Strikingly, this happened in a contemporary On-X® prosthesis, only 4 years post-implant. Timely echocardiography is of utmost importance in patients with prosthetic heart valves presenting in heart failure.
Copyright © 2020. Published by Elsevier Inc.

Entities:  

Year:  2020        PMID: 32114040     DOI: 10.1016/j.athoracsur.2020.01.023

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  3 in total

1.  On-X mitral valve leaflet embolization presenting as acute pulmonary oedema.

Authors:  Lindiwe Sidali; Nandipha Ndibi; Manogran Moopanar; Rajhmun Madansein
Journal:  Egypt Heart J       Date:  2022-05-04

2.  Commentary: How can you mend a broken heart: Mechanical valve leaflet fracture.

Authors:  Parth M Patel; Edward P Chen
Journal:  JTCVS Tech       Date:  2020-06-30

3.  Leaflet fracture and embolization from an On-X mechanical mitral valve.

Authors:  Erik Helgeland; Kristoffer Andresen; Karl Andreas Dumont; Johannes Lagethon Bjørnstad
Journal:  JTCVS Tech       Date:  2020-06-24