Raffaele Liuzzi1, Consiglia Piccolo2, Vittoria D'Avino1,3, Stefania Clemente4, Caterina Oliviero4, Laura Cella1, Mariagabriella Pugliese3,5.
Abstract
PURPOSE: The aim of the study was to exploit the feasibility of thermoluminescent dosimeters (TLDs) in radiation therapy techniques in which high dose per fraction is involved.Entities:
Keywords: TLD; calibration; correction factor; hypofractionated radiotherapy
Year: 2020 PMID: 32110167 PMCID: PMC7019407 DOI: 10.1177/1559325819894081
Source DB: PubMed Journal: Dose Response ISSN: 1559-3258 Impact factor: 2.658
Figure 1.Schematic diagram showing the thermoluminescent dosimeter (TLD)-100 calibration setup using water-equivalent RW3 slab phantom for photon (A) and electron irradiations (B).
Figure 2.Distribution of element correction factors (ECFs) generated for thermoluminescent dosimeters (TLDs) exposed to 6-MeV photon beam (A and B) and 6-MeV electron beam (C).
Figure 3.Calibration curves of thermoluminescent dosimeters (TLDs) exposed to 6-MV photon beam in different dose range: 0-8 Gy (A), 0-10 Gy (B), and 0-30 Gy (C). The solid line represents the linear fit, and the dot line represents the quadratic fit.
Figure 4.Calibration curves of thermoluminescent dosimeters (TLDs) exposed to 6-MeV electron beam in different dose range: 0-8 Gy (A), 0-10 Gy (B), and 0-30 Gy (C). The solid line represents the linear fit, and the dot line represents the quadratic fit.
Best-Fit Regression Coefficients and Statistical Parameters for Linear and Quadratic Dose–Response Models for the TLD-Set Exposed to 6-MV Photon Beam.
| Dose Range, Gy | Equation Model | Equation Model | ||||
|---|---|---|---|---|---|---|
| Linear | Quadratic | |||||
|
| y = α + β0
| |||||
| Parameter | Value | SE | Parameter | Value | SE | |
| 0-8 | α | −0.6 | 0.9 | α | 0 | 1 |
| β0 | 5.5 | 0.2 | β0 | 4.5 | 0.7 | |
| β1 | 0.12 | 0.08 | ||||
| Adjusted | .967 | Adjusted | .968 | |||
| Aw | 0.51 | Aw | 0.49 | |||
| 0-10 | α | −1 | 1 | α | 0 | 1 |
| β0 | 5.7 | 0.2 | β0 | 4.5 | 0.6 | |
| β1 | 0.13 | 0.06 | ||||
| Adjusted | .969 | Adjusted | .973 | |||
| Aw | 0.20 | Aw | 0.79 | |||
| 0-30 | α | −8 | 2 | α | 1 | 1 |
| β0 | 7.07 | 0.15 | β0 | 4.4 | 0.3 | |
| β1 | 0.10 | 0.01 | ||||
| Adjusted | .977 | Adjusted | .992 | |||
| Aw | 0 | Aw | 1 | |||
Abbreviations: SE, standard error; TLD, thermoluminescent dosimeters.
Best-Fit Regression Coefficients and Statistical Parameters for Linear and Quadratic Dose–Response Models for the TLD-Sets Exposed to 6-MeV Electron Beam.
| Dose Range, Gy | Equation Model | Equation Model | ||||
|---|---|---|---|---|---|---|
| Linear | Quadratic | |||||
|
|
| |||||
| Parameter | Value | SE | Parameter | Value | SE | |
| 0-8 | α | −1 | 1 | α | 0.1 | 1.1 |
| β0 | 6.8 | 0.2 | β0 | 5.3 | 0.7 | |
| β1 | 0.2 | 0.1 | ||||
| Adjusted | .979 | Adjusted | .981 | |||
| Aw | 0.28 | Aw | 0.72 | |||
| 0-10 | α | −1 | 1 | α | 0 | 1 |
| β0 | 7.03 | 0.15 | β0 | 5.8 | 0.5 | |
| β1 | 0.12 | 0.05 | ||||
| Adjusted | .982 | Adjusted | .984 | |||
| Aw | 0.17 | Aw | 0.83 | |||
| 0-30 | α | −19 | 3 | α | 2 | 3 |
| β0 | 9.7 | 0.2 | β0 | 5.1 | 0.5 | |
| β1 | 0.147 | 0.015 | ||||
| Adjusted | .973 | Adjusted | .989 | |||
| Aw | 0 | Aw | 1 | |||
Abbreviations: SE, standard error; TLD, thermoluminescent dosimeters.